CCCCCCCCCCCC CCCCCCCCCCCC CCCC CCCC ooooooo CCCC oo oo nnn CCCC oo oo n n t r i b u t e d Library for the Agilent ChemStation CCCC oo oo n n CCCC ooooooo CCCC CCCCCCCCCCCC CCCCCCCCCCCC ChemStation Revision A.10.01 Date: 28-July-03 LC Macro Contribution #15 Contribution 'Using Hypersequence without Well-Plate Handler' Revision History ---------------- 1. A.01 Original Version for ChemStation A.10.01 28-July-03 2. A.02 Version checked with ChemStation B.01.01 3. B.01 New macro version checked with ChemStation B.03.01 Contribution Purpose -------------------- This patch allows the use of the Hypersequence without using the Well-Plate Handler. NOTES: - It will not work if Security Pack is installed. - This is a user contributed macro. It is not officially supported by Agilent Technologies. - It is not fully tested so some unknown errors can occur. Contribution Structure ---------------------- This macro is a set of files, containing... - applics.mac (Macro File) README.TXT - this file Prerequisites ------------- - MS operating system as required by the revision of the ChemStation in use - The ChemStation software, Rev B.03.01 or higher Installation ------------ Copy content of Macro to the ChemStation's executable directory, 'C:\Chem32\CORE' by default. (Please save before the original version of that files, so you can restore them if something works wrong!) After the next restart of the Chemstation the Hypersequence should work without having a Well-Plate Handler installed. Use --- Important to know: - In the Hypersequence table you have not to specify a Plate Type. The Plate Types of the Configuration will be taken.