               BioPack Automatic Calibration for Solids
********************************************************************
        This procedure calibrates 15N and 13C cross polarization
        using a labeled peptide with C13 offset at 175 ppm and 
        15N offset at 115ppm.  Double CP is optimized at the 
        C13 (3Wr/2) and 15N (5Wr/2).  

        All calibration data are stored in the AutoTripRes
        directory and the probefile calibrations (_NCO) are updated.

     Experiments Performed Include:     (pulse sequence name)

      13C Spinlock to gB1=0 to 3Wr for Rotary
         Resonance Dips using aXlock               tancpxt1rho
      13C Spinlock to gB1=0 to 3Wr for Rotary
         Resonance Dips using aXlock               tancpxt1rho
      13C-15N Hartmann Hahn match at N15 (5Wr/2)
         and 13C (3Wr/2) using aXyx and aYyx       dcptan
      Fine 13C-15N Hartmann Hahn match at N15 
        (5Wr/2) and 13C (3Wr/2) using aXyx         dcptan
      13C-15N CP RAMP using bYX and dYX            dcptan
      13C-15N Contact Time using tHX               dcptan
      A Standard Double CP Spectrum (4scans)       tancpx


      