               BioPack Automatic Calibration for Solids
********************************************************************
        This procedure calibrates 1H and 13C cross polarization
        and decoupling using a labeled peptide with C13 offset
        at 60 ppm. CP is optimized for the 1H first upper sideband.

        All calibration data are stored in the AutoTripRes
        directory and the probefile calibrations (_HCa) are updated.

     Experiments Performed Include:     (pulse sequence name)

      13C-1H Hartmann-Hahn Match (1H +1)using aHhx     tancpx
      13C-1H CP Ramp using bHX and dHX                 tancpx
      13C-1H Contact Time using tHX                    tancpx
      1H SPINAL decoupling using pwHspinal, phHspinal  tancpx

      