               BioPack Automatic Calibration for Solids
********************************************************************
        This procedure calibrates 1H and 15N cross polarization
        and decoupling using a labeled peptide with N15 offset
        at 115 ppm. CP is optimized for the 1H first upper sideband.

        All calibration data are stored in the AutoTripRes
        directory and the probefile calibrations (_HNai) are updated.

     Experiments Performed Include:     (pulse sequence name)

      15N-1H Hartmann-Hahn Match (1H +1)using aHhx     tancpx
      15N-1H CP Ramp using bHX and dHX                 tancpx
      15N-1H Contact Time using tHX                    tancpx
      1H SPINAL decoupling using pwHspinal, phHspinal  tancpx

      