Product
Comparison - Power Sensors and Thermistor Mounts (Specials)
This selection guide comprises the common power sensor special options (including Thermistor Mount).
For further information
on power sensor specials, contact your local Agilent FE or sales office or call
1-800-452-4844 in the US.
Specials
Sensors Number
|
Description |
---|---|
Standard
Lab calibrations can be performed on 8480 series power sensors at specified
frequencies.
|
|
8480
Series
|
|
8481A-H42
|
Factory selected for the best SWR from batch |
8481D-H70
|
Frequency range 100 kHz to 4.2 GHz, with standard lab calibration, power range -70 to -20 dBm |
8481D-H71
|
Frequency range 100 kHz to 4.2 GHz, with advanced lab calibration, power range -70 to -20 dBm |
8481D-H72
|
Frequency range 100 kHz to 4.2 GHz, power range -70 to -20 dBm |
8481D-H06
|
Improved DC offset specification to allow use of long sensor cable (up to 1000 ft) |
8481D-H51
|
Guaranteed maximum SWR of 1.2 from 18 to 18.5 GHz (no other specifications to 18.5 GHz) |
8482B-H50
|
Allows an average input power of up to 50 Watts (+47 dBm), (25 Watts is standard power range) |
8483A-H03
|
Provides increased frequency range calibration to 3 GHz, (2 GHz is standard frequency range) |
V8486A-H01
|
Allows extended calibration to 78 GHz (standard upper frequency is 75 GHz) |
E-Series |
|
E4413A-H33
|
Specified performance for 0.05-33 GHz frequency range (standard frequency range is 26.5 GHz) |
E9300A-H24
|
Extended frequency range 10 MHz to 24 GHz, -60 to +20 dBm (APC 3.5(m) connector) |
E9300A-H25
|
Extended frequency range 10 MHz to 24 GHz, -50 to +30 dBm (APC 3.5(m) connector) |
E9304A-H18
|
Extended frequency range 9 kHz to 18 GHz, power range -60 to +20 dBm |
E9304A-H19
|
Extended frequency range 9 kHz to 18 GHz, power range -50 to +30 dBm |
Thermistor
Mounts
|
|
478A-H55
|
Frequency range 1 MHz to 1 GHz, with maximum SWR 1.3 |
478A-H63
|
Frequency range 100 kHz to 1 GHz, maximum SWR 1.8 to 300 kHz and 1.3 to 1 GHz |
478A-H72
|
Frequency range 1 MHz to 1 GHz, maximum SWR 1.2 |
478A-H73
|
Frequency range 1 MHz to 100 MHz, maximum SWR 1.1 except 1.05 at 50 MHz |
478A-H75
|
Frequency range 1 MHz to 1 GHz, with maximum SWR 1.05 at 50 MHz |
478A-H76
|
Frequency range 1 MHz to 1 GHz, maximum SWR 1.3 except 1.05 at 50 MHz. Standard lab calibration at 50 MHz |
8478B-H43
|
Factory selected for best SWR from batch. Standard lab calibration every 1 GHz to 18 GHz. |
8478B-H84
|
Standard lab calibration at specified frequencies |
© Agilent Technologies, Inc. 2003