HP 10020A Resistive Divider Probe Kit![]()
|
The HP 10020A Resistive Divider Probe Kit is a signal probing
system for measuring fast-transition-time signals in
high-impedance systems. It is designed for use with 50 ohm
input oscilloscopes, but may be used with other than 50 ohms
systems if a 50 ohm feedthrough termination such as the
HP 10100C, is used.
The HP 10240B Blocking Capacitor is a probing
accessessory that provides ac-coupling while
maintaining a 50 ohm system. This capacitor is designed
for use with the model 10020A Resistive Divider Kit,
but may be used with any probe that must be terminated
in 50 ohms.
The HP 10229A Hook Tip Adapter is an available accessory.
Output Connector: BNC
Length (overall): Approx. 1.2 m (4 ft)
Weight: Net 0.45 KG (1 lb)High-Frequency Resistive Dividers
(Supplied in the HP 10020A Resistive Divider Probe Kit)
Part
No. | Division Ratio | Input R ohms1 | Division Accuracy | Max V rms2 | Input C pF | Probe Rise Time ns3 | Bandwidth MHz |
10020-67701 | 1:1 | 50 | ----------- | 6 | ------ | <0.5 | 700 |
10020-67702 | 5:1 | 250 | +/-3% | 9 | <0.7 | <0.5 | 700 |
10020-67703 | 10:1 | 500 | +/-3% | 12 | <0.7 | <0.5 | 700 |
10020-67704 | 20:1 | 1000 | +/-3% | 15 | <0.7 | <0.5 | 700 |
10020-67705 | 50:1 | 2500 | +/-3% | 25 | <0.7 | <0.5 | 700 |
10020-67706 | 100:1 | 5000 | +/-3% | 35 | <0.7 | <0.5 | 700 |
1When terminated in 50ohms
2Limited by power dissipation of the resistive element
3Rise time with or without 10240B blocking capacitor