HP 10706A Plane Mirror Interferometer![]()
|
The HP 10706A is one member of a family of measurement optics that tailor interferometer systems for the physical layout and measurement requirements of each application. Configured with appropriate laser head and system electronics, the optics are part of a precise positioning system providing the highest positioning accuracy, resolution, and repeatability.
Typical Measurement Mirror Alignment Requirements (as a function of distance): 152 mm (6 in): +/- 6 arc-min from normal 305 mm (12 in): +/- 3 arc-min from normal 508 mm (20 in): +/- 1.5 arc-min from normal Measurement Optics Summary Application: Plane mirror measurement reflector Resolution: Optics: lambda/4 (158.25 nm) System: lambda/128 (5 nm) (When used with the HP 5527B Laser Positioning Transducer System, HP 10885A PC Axis Board or HP 10895A Laser Axis Board for VMEbus.) lambda/1024 (0.6 nm) (When used with the HP 18097B High Resolution Laser Axis Board for VMEbus.) Beam Separation: 12.7 mm (1/2 in) Recommended Reflector: HP 10724A Reflector Weight: 50 g (1.8 oz) Mount Used: HP 10711A Weight: 308 g (10.8 oz)
HP 10706A Plane Mirror Interferometer< Opt 080 Fiber-Optic Receiver AdapterThe HP 10711A Adjustable Mount is available for use with the HP 10706A. The HP 10706A is supported in combination with the HP 10724A Plane Mirror Reflector. The HP 10706A can be used in a variety of laser-interferometer positioning systems including the HP PC compatible and VMEbus compatible systems.
Related HP 10706 Plane Mirror Interferometer Family product datasheets: