[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]
The HP 10780F Remote Receivers with fiber-optics pickups provide maximum layout flexibility for HP laser interferometer positioning systems. Remote pickup capabilities allow mounting of the heat-dissipating receiver electronics away from measurement areas, resulting in high measurement stability, accuracy, and repeatability.

One receiver is used for each measurement axis (including wavelength tracker). Configured with HP electronics, laser heads, and optics, the HP 10780F receiver lets you tailor a precise positioning system for your exact application requirements.


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Easy Access

 
The fiber-optic cable used to attach the remote sensor to the receiver
electronics allows design flexibility and easy access to the receiver
gain adjustment.


Self-Aligning

 
The HP 10780F is self-aligning when mounted to selected HP sensors and
interferometers:  the HP 10715A, HP 10716A, HP 10717A, HP 10719A, HP
10721A, HP 10735A, HP 10736A, HP 10737L and the HP 10737R.
 


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP 10780F Remote Receiver

Typical Power Requirements: +15 volts at 136 mA

Maximum Sensitivity: 2.2 uW (with 2 m cable, becomes 5.0 uW with 
	a 10 m fiber cable)

Heat Dissipation:	0.0 W for remote sensor
			2.0 W typical for receiver

Output Signal: Differential square wave at Doppler-shifted split frequency
	(100kHz to 6.0 Mhz)

Fiber-optic Cable Length: 2 m standard, 10 m maximum recommended

Alignment Tolerances:	Roll: +/- 3 degrees
			Pitch: +/- 1 degree
			Yaw: +/- 1 degree

	HP 10780F is self-aligning when mounted to the HP 10715A, HP 10716A, HP 10717A, 
	HP 10719A, HP 10721A, HP 10735A, HP 10736A, HP 10737L, and HP 10737R


Weight:	126 g (4.5 oz) 
	26 g (0.9 oz) for remote sensor with 2m cable


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP 10780F 	Remote Receiver
   Opt 037 	Remove Lens Assembly
   Opt W50 	5-Year Return Repair Service

The HP 10780F can be used in a variety of laser-interferometer positioning systems, including the HP PC-compatible, VMEbus compatible, and HP 5527B systems.

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

To help you make the right instrument choices

    Your Field Sales Engineer is ready to answer your questions about specifications, applications and ordering. Find your Field Sales Engineer at your Local Sales Office.

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Related product families available online:



Contact HP Test & Measurement. (c) Copyright 1994, 1995, 1996, 1997 Hewlett-Packard Company.

Page Updated: Thursday November 13 08:02:58 UTC 1997