HP 1144A 800 MHz Active Probe

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]
The HP 1144A active probe provides an excellent combination of high bandwidth, high input resistance, and low input capacitance. This combination is superior to that of passive divider probes because it provides minimal circuit loading at high and low frequencies. As signal speeds increase, passive divider probes may not deliver the performance needed. This can result in waveshapes that do not represent the actual signal.

To help you minimize circuit loading, the HP 1144A boasts 800-MHz bandwidth, 1-M ohm input resistance, and 2-pF input capacitance. It can be connected to any instrument with 50-Ohm inputs. An external 50-Ohm termination can also be used with high impedance inputs.

The HP 1144A can access power directly from the HP 54542, HP 54540, the HP 54522, the HP 54520, the HP 54615B, the HP 54616B and the HP 54616C oscilloscopes. The HP 54500 family scopes scopes provide power for two channels of active probing. If four channels are needed, a special one-input, two-output adapter is available (HP 01144-61604). Two adapters are needed for four channels of probing. The HP 54615B, the HP 54616B, and the HP 54616C provide power for one channel of active probing. If two channels are needed the HP 01144-61604 can be used. If the HP 1144A is used with any oscilloscope not listed above, then the HP 1142A power supply is required. The HP 01144-61604 adapter can be used with this power supply to provide power for two channels of active probing.

Versatile accessories allow for easy connection to your device under test. For example, the dual lead adapter combines with the two mini-pincher tips to provide a solid connection to .050" pitch surface mount devices for hands-off probing. Also, the ground bayonet has a short ground lead to minimize measurement errors such as ringing, caused by parasitic inductances.

For hands-off probing of SMT components with .5mm (.020") to .8mm (.032") pitch, an additional accessory is available: the HP 10075A .5mm MicroGrabber Accessory Kit.


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Bandwidth:1                          Greater than or equal to 800 MHz
RiseTime:1,2                          Less than or equal to 440 ps
Attenuation3:                        10:1 +/-2%
Input Resistance:                    1 Mohm +/-5%
Maximum Input Voltage:               +/-40V (dc +peak ac)
1Above 35°C, bandwidth and rise time degrade approximately 1/2%/°C.

2Rise time figure calculated from tr=0.35/Bandwidth.

3When connected to an instrument input of 50 Ohm, +/-0.5%.


Characteristics

Input Capacitance: 2 pF (typical) Overshoot and Ringing:1 +/-10% for the first 6 ns +/-4% from 6 ns to 20 ms +/-1.5% thereafter Output Voltage offset: Less than +/- 15 mV Input Dynamic Range:2 +/-7.0V Output Load Requirement: 50 ohm Power Requirements: dc +/-12V to +/-15V +/-5% (at approximately 75 mA each supply) Weight: Net: Approximately 180g(-6.5oz) Probe body 35g(~1.25 oz) Shipping: Approximately 0.8kg (1.75 lb) 1When used with a 3-GHz oscilloscope (oscilloscope characteristics excluded). 2When connected to an instrument input of 50 ohm, +/-0.5%.


Environmental Characteristics

Temperature: Operating: 0°C to +55°C (32°F to +131°F) Non-Operating: -40°C to +70°C (-40°F to+158°F) Humidity: Operating: up to 95% relative humidity (non-condensing) at +40°C (+104°F) Non-Operating: up to 90% relative humidity at +65°C (+149°F) Altitude: Operating: up to 4,600 meters (15,000ft) Non-Operating: up to 15,300 meters (50,000ft) Vibration: Random vibration 5 to 500 Hz, 10 minutes per axis, 0.3 grms Random vibration 5 to 500 Hz, 10 minutes per axis, 2.41 grms Resonant search 5 to 500 Hz Swept sine, 1 octave/minute Sweep rate, (0.75g), 5 minute Resonant dwell at 4 resonances per axis


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP 1144A            800 MHz Active Probe

HP 10072A SMT Accessory Kit for HP 10070 Family Probes

HP 10075A 0.5 mm MicroGrabber Accessory Kit

HP 01144-61604 One-input, Two-output Adapter

HP 1142A Probe Control and Power Module

All orders include the following:

Available accessories for this product include:


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

No literature is available online at this time.

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]
Additional help is a click away!

Want to talk to a knowledgeable engineer about this product at our HP Test and Measurement Call Center? Do you have a question you want to send in by e-mail? A measurement set-up you need to fax in? Follow the link above and you will be given a choice to contact an engineer by telephone, e-mail or fax.

This service is available only in the United States. For assistance outside the U.S., contact your local HP DIRECT or HP Sales Office.

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Related product families available online:



Privacy Statement    Use of this site indicates that you accept the Terms of Use
Contact HP Test & Measurement. (c) Copyright 1994 - 1998 Hewlett-Packard Company.

Page Updated: Monday October 05 18:59:54 UTC 1998