The HP 1146A AC/DC Oscilloscope Current Probe expands oscilloscope
applications in industrial, automotive or power environments, and is
ideal for analysis and measurement of distorted current waveforms and
harmonics.
The probe permits accurate display and measurment of
currents from 100 mA to 100 A rms, DC to 100 kHz without breaking into
the circuit. The probe uses Hall effect technology to measure AC and
DC signals. The probe connects directly to an oscilloscope through a
coaxial cable with an insulated BNC.
The HP 1146A AC/DC Oscilloscope Current Probe is compatible with any
analog or digital oscilloscope or other voltage measuring instrument
which has the following features:
- BNC Input Connector
- Range Capable of Displaying 0.2 to 0.5 V per Division
- Minimum Input Impedance of 1 Mohm
Current Measurement
Connect the current probe to the proper input channel on the
oscilloscope. Begin with the least sensitive range on the current
probe (10 mV/A). Select the 0.5 V/Division range on your oscilloscope.
Clamp the probe on the conductor to be measured and read the current
flowing directly on your oscilloscope.
You may also use your oscilloscope to amplify the signal while using
the 100 mV/A probe range (which offers the best accuracy and least
phase shift).
![[ Summary ]](/tmo/datasheets/Graphics/tab-summary-reg.gif)
![[ Features ]](/tmo/datasheets/Graphics/tab-features-high.gif)
![[ Specifications ]](/tmo/datasheets/Graphics/tab-specs-reg.gif)
![[ Key Literature ]](/tmo/datasheets/Graphics/tab-keylit-reg.gif)
![[ Ordering ]](/tmo/datasheets/Graphics/tab-ordering-reg.gif)
![[ Assistance ]](/tmo/datasheets/Graphics/tab-assistance-reg.gif)
- Measures Current from dc-100 KHz
- Clamps onto Current Carrying Conductor
- Battery Powered
![[ Summary ]](/tmo/datasheets/Graphics/tab-summary-reg.gif)
![[ Features ]](/tmo/datasheets/Graphics/tab-features-reg.gif)
![[ Specifications ]](/tmo/datasheets/Graphics/tab-specs-high.gif)
![[ Key Literature ]](/tmo/datasheets/Graphics/tab-keylit-reg.gif)
![[ Ordering ]](/tmo/datasheets/Graphics/tab-ordering-reg.gif)
![[ Assistance ]](/tmo/datasheets/Graphics/tab-assistance-reg.gif)
Characteristics marked with asteriks are specific performaces,
all others are typical characteristics.
Electrical Characteristics
*Current Range: 100 mV/A: 100 mA to 10 A peak
10 mV/A: 1 to 100 A peak
Output Signal: 1000 mV peak max
*AC current accuracy: (After calibration and for one year)
(zero probe before making measurement)
Range
| Accuracy
|
100 mV/A (50 mA to 10 A peak)
| 3% of reading +/-50 mA
|
10 mV/A (500 mA to 40 A peak)
| 4% of reading +/-50 mA
|
10 mV/A (40 A to100 A peak)
| 15% max at 100A
|
Phase Shift:1 <1° from dc to 65 Hz on 10 mV/A
<1.5° from dc to 65 Hz on 100 mV/A
*Frequency Range: DC to 100 kHz (-3dB with current derating)
Noise: Range 10 mV/A: 480 µV
Range 100 mV/A: 3 mV
Slew Rate: Range 10 mV/A: 20 mV/µs
Range 100 mV/A: 0.3 V/µs
Load Impedance: >1 Mohm/ 100 pF
Insertion Impedance (50/60 Hz): 100 mV/A: 0.01 ohm
10 mV/A: 0.01 ohm
Rise or Fall time: Range 100 mV/A: 3 µs
Range 10 mV/A: 4 µs
Working Voltage: 600 Vrms max.
Common mode voltage: 600 Vrms max.
Influence of adjacent conductor: <0.2 mA/A AC
Influence of conductor position in jaw: 0.5% of reading at 1 kHz
Battery: 9 V alkaline (NEDA 1604A, IEC 6LR61)
Low battery: green LED off when greater than or equal
to 6.5V
Overload indication: red LED indicates input greater than the
selected range.
Typical Consumption: 8.6 mA
Battery Life: 55 hours typical
1Reference conditions
23°C +/-5°C, 20 to 75% relative humidity, DC
to 1kHz, probe zeroed, 1 minute warmup, batteries at 9V
+/- 0.1 V, external magnetic field <40 A/m, no DC
component, no external current carrying conductor,
1 M ohm / 100 pF load, conductor centered in jaw.
Environmental Characteristics
Operating Temperature: 0° to 50°C (32° to 122°F)
Storage Temperature: -30° to 80°C (-22° to 176°F)
Temperature Influence: <0.2% per °C
Operating Relative Humidity: 10° to 30°C: 85 +/- 5% RH (without condensation)
40° to 50°C: 45 +/- 5% RH (without condensation)
Altitude: Operating: 0 to 2000 m
Non operating:0 to 12,000 m
Mechanical Characteristics
Zero Adjustment: 20 turn potentiometer
Maximum cable diameter: 0.46"(11.8 mm)
Case Protection: IP20 per IEC 529
Drop Test: 1.0m on 38mm of oak on concrete; test according to
IEC 1010
Mechanical Shock: 100 G; test per IEC 68-2-27
Vibration: Test per IEC 68-2-6, frequency range 10 Hz to 55 Hz,
amplitude 0.15 mm
Handle: Lexan® 920A, UL 94 V2, color light gray
Dimensions: 9.09 x 1.42 x 2.64" ( 231 x 36 x 67mm)
Weight: 11.6 oz (330 g) with battery
Color: light gray
Output lead: insulated coaxial cable with insulated BNC connector
Cable length: 6.5 foot (2m)
Safety Characteristics
- Double insulation or reinforced insulation between primary or secondary
and outer case of the handle, per IEC 1010.
- 600 V category III, Pollution: 2.
- 300 V category IV, Pollution: 2.
- 5550 V 50/60 Hz between primary or secondary and the outer case of the
handle.
- 3250V 50/60 Hz between primary and secondary.
Electromagnetic compatibility CE
EN 50081-1 Class B ISM 1-A
EN 50082-2 Electrostatic discharge IEC 1000-4-2
Radiated Field IEC 1000-4-3
Fast transients IEC 1000-4-4
Magnetic Field at 50/60 Hz IEC 1000-4-8
![[ Summary ]](/tmo/datasheets/Graphics/tab-summary-reg.gif)
![[ Features ]](/tmo/datasheets/Graphics/tab-features-reg.gif)
![[ Specifications ]](/tmo/datasheets/Graphics/tab-specs-reg.gif)
![[ Key Literature ]](/tmo/datasheets/Graphics/tab-keylit-reg.gif)
![[ Ordering ]](/tmo/datasheets/Graphics/tab-ordering-high.gif)
![[ Assistance ]](/tmo/datasheets/Graphics/tab-assistance-reg.gif)
HP 1146A AC/DC Oscilloscope Current Probe
![[ Summary ]](/tmo/datasheets/Graphics/tab-summary-reg.gif)
![[ Features ]](/tmo/datasheets/Graphics/tab-features-reg.gif)
![[ Specifications ]](/tmo/datasheets/Graphics/tab-specs-reg.gif)
![[ Key Literature ]](/tmo/datasheets/Graphics/tab-keylit-high.gif)
![[ Ordering ]](/tmo/datasheets/Graphics/tab-ordering-reg.gif)
![[ Assistance ]](/tmo/datasheets/Graphics/tab-assistance-reg.gif)
No literature is available online at this time.
![[ Summary ]](/tmo/datasheets/Graphics/tab-summary-reg.gif)
![[ Features ]](/tmo/datasheets/Graphics/tab-features-reg.gif)
![[ Specifications ]](/tmo/datasheets/Graphics/tab-specs-reg.gif)
![[ Key Literature ]](/tmo/datasheets/Graphics/tab-keylit-reg.gif)
![[ Ordering ]](/tmo/datasheets/Graphics/tab-ordering-reg.gif)
![[ Assistance ]](/tmo/datasheets/Graphics/tab-assistance-high.gif)
Additional help is a click away!
Want to talk to a knowledgeable engineer about this product at our
HP Test and Measurement Call Center? Do you have a question you want
to send in by e-mail? A measurement set-up you need to fax in?
Follow the link above and you will be given a choice to
contact an engineer by telephone, e-mail or fax.
This service is available only in the United States.
For assistance outside the U.S., contact your
local HP DIRECT or
HP Sales Office.
![[ Summary ]](/tmo/datasheets/Graphics/tab-summary-reg.gif)
![[ Features ]](/tmo/datasheets/Graphics/tab-features-reg.gif)
![[ Specifications ]](/tmo/datasheets/Graphics/tab-specs-reg.gif)
![[ Key Literature ]](/tmo/datasheets/Graphics/tab-keylit-reg.gif)
![[ Ordering ]](/tmo/datasheets/Graphics/tab-ordering-reg.gif)
![[ Assistance ]](/tmo/datasheets/Graphics/tab-assistance-reg.gif)
Related product families available online:

Page Updated: Monday October 05 19:04:02 UTC 1998