HP 16535A MultiProbe Control Module |
The HP MultiProbe system makes it possible to connect scope probes to hundreds of nodes simultaneously, then select any two signals at a time for scope acquisition and viewing, without having to move scope probes. The HP MultiProbe system is part of HP's 16500C logic analysis system, so you have all the tools you need to solve difficutl problems in complex digital hardware.
An HP MultiProbe system consists of the following components:
One control module is required. Multiple control modules can be usedwhen more than four MultiProbe pods are needed, or when more than two scope channels are desired. When using the HP 16533A or 16534A oscilloscope, the control module and its associated scope module must be in the same HP 16500C mainframe or HP 16501 expander frame.
Specifications and characteristics apply only within an ambient temperature range of +10oC from the temperature at which self-calibration was performed.
The following specifications and characteristics apply when using an HP MultiProbe pod and an HP 16535A MultiProbe control module.
Specifications | HP E5320A | HP E5321A |
---|---|---|
HP E5322A | ||
Attenuation (Nominal) (from any input | 50:1 or 25:1 | 50:1 or 25:1 |
of the MultiProbe pod to the output of the | HP 16535A MultiProbe control module)1 | dc offset 2 | +/- 1mV | +/- 1mV |
Bandwidth | > 1 GHz | >750 MHz |
Notes | |
---|---|
1 | The actual probe attenuation is indicated in the user interface. The indicated probe attenuation is correct within +2% (specification). | 2 | At the temperature at which the self-calibration was performed. |
Characteristics | HP E5320A | HP E5321A |
---|---|---|
HP E5322A | ||
Input resistance | 10K Ohms +2% | 20K Ohms (Nominal) |
Input capacitance (typ) | 1 pF | 3.3 pF3 |
Mutual capacitance (typ)(adjacent pins) | N/A | 0.45 pF |
Dynamic range (referenced toinput) | ||
| -6V to +13V | -4V to + 10V |
| -3V to + 10V | -3V to + 9V |
| < 2% | < 4% |
dc offset drift | 300 µV per oC | 300 µV per oC |
Propagation delay (nominal) from an input | 10.5 ns | 10.5 ns |
of the HP MultiProbe pod to the output of | ||
the HP 16535A MultiProbe control module | ||
Rise time (calculated from bandwidth) | < 350 ps | < 467 ps |
Overshoot (for a step with a rise time > 500 ps) | < 5% | < 5% |
Coupling (for a step with a rise time> 1 ns) | < 4% | < 4% |
(internal to probe) | ||
ESD Tolerance | + 8kV | Requires ESD |
(150 pF, 330 Ohms) | precautions | |
Maximum input voltage (dc + peak ac) | +40V | +20V |
Operating temperature | 0o to +55oC | 0o to +55oC |
Notes | |
---|---|
3 | The input capacitance (3.3 pF) isthe capacitance to ground (2.4 pF) plus the mutual capacitance between each adjacent pin (0.45 pF) x 2. |
The following specifications and characteristics apply when using an HP MultiProbe pod and an HP 16535A MultiProbe control module, and an P 16534A oscilloscope.
Characteristics | HP E5320A | HP E5321A |
---|---|---|
HP E5322A | ||
dc voltage measurement accuracy | +3% of full scale | +3% of full scale] |
dc offset accuracy | +[1% of offset + 2% of full scale] /td> | +[1% of offset + 2% of full scale |
(1 mV + 300µV/ oC) x probe attenuation factor) | (1 mV + 300µV/ oC) x probe attenuation factor) | |
Bandwidth | > 500MHz | > 500MHz |
Rise time (calculated from bandwidth) | < 700 ps | < 700 ps |
Differential time interval | ||
measurement accuracy | ||
For two inputs on the same pod | < 200 ps | < 250 ps |
For two inputs on different pods | < 300 ps | < 300 ps |
The following specifications and characteristics apply when using an HP MultiProbe pod and an HP 16535A MultiProbe control module with other oscilloscopes.
Characteristics | HP E5320A | HP E5321A |
---|---|---|
HP E5322A | ||
Bandwidth (typical) | ||
HP 54121T | 1 GHz | 750 MHz |
HP 54750A | 1 GHz | 750 MHz |
HP 54710A, D | ||
HP 54720A, D | ||
with plug-in:D | ||
| 1 GHz | 750 MHz |
| 800 MHz | 600 MHz |
| 800 MHz | 600 MHz |
| 500 MHz | 500 MHz |
| 500 MHz | 500 MHz |
| 500 MHz | 500 MHz |
HP 16532A | 250 MHz | 250 MHz |
HP 16533A | 250 MHz | 250 MHz |
Differential time interval | ||
measurement accuracy4 | ||
Signals on same HP MultiProbe pod | ||
| + 200 ps | + 250 ps |
| + 300 ps | + 500 ps |
Signals on different HP MultiProbe pods | ||
| + 300 ps | + 800 ps |
| + 300 ps | + 800 ps |
Notes | |
---|---|
4 | Skew correction factors are provided on the HP MultiProbe control module user interface that can be used to correct differential time interval measurements on other oscilloscopes. |
One required.
HP16535A HP Two-output MultiProbe control module
As Needed. Up to four pods can be connected at one time.
HP E5320A 9-to-2, general-purpose MultiProbe pod Has nine miniature scope-probe flying-lead inputs Opt 001 Delete accessories Opt 002 Delete calibration fixture HP E5321A 240-to-2, high-density MultiProbe pod Opt 001 240-pin QFP personality adapter kit 240-pin personality adapter Retainer kit Locator tool Calibration adapter Calibration fixture Opt 101 240-pin QFP personality adapter Opt 201 Retainer kit for 240-pin QFP Five retainers Two knurled nuts One tube of glue Opt 202 Locator tool Opt 301 Calibration kit Calibration adapter Calibration fixture HP E5322A 208-to-2, high-density MultiProbe pod Opt 001 208-pin QFP personality adapter kit 208-pin personality adapter Retainer kit Locator tool Calibration adapter Calibration fixture Opt 101 208-pin QFP personality adapter Opt 002 160-pin QFP personality adapter kit 160-pin personality adapter Retainer kit Locator tool Calibration fixture Opt 102 160-pin personality adapter Opt 201 Retainer kit for both 160-pin and 208-pin QFPs Five retainers Two knurled nuts One tube of glue Opt 202 208-pin locator tool Opt 203 160-pin locator tool Opt 301 Calibration kit Calibration adapter Calibration fixture
One required.
HP 16534A 2-GSa/s, 500-MHz oscilloscope module for HP 16500C HP 16533A 1-GSa/s, 250-MHz oscilloscope module for HP 16500C Any other oscilloscope can be used with the HP MultiProbe system.
One required.
HP 16500C Logic Analyzer mainframe
Optional
HP 16505A Prototype AnalyzerHP MultiProbe System Technical Specifications
If you do not have the Adobe Acrobat Reader necessary for viewing this documentation, download your Free Acrobat Reader now. |
![]() |
Page Updated: Thursday June 19 16:11:05 UTC 1997