HP CaLan 2010B SLM Plus![]()
|
The HP CaLan 2010B is also a data management system. It has an RS-232 serial computer interface and supports both serial and parallel printers. The automatic testing capability provides you with pass/fail reports and analysis. The SLM Plus features unattended 24-hour testing and optional compliance reporting. It is portable, rugged, and economically priced. An optional fiber optic power meter lets you measure coax and fiber with a single instrument.
Frequency Specifications
Frequency Range: 5 MHz to 1 GHz Frequency Reference Stability: ±5 x 106 over full temperature range Marker Frequency Accuracy Accuracy: ±25 kHz Resolution: 10 kHz Frequency Span Range: 0 MHz to 999 MHz Resolution: 1 MHz Resolution Bandwidth: 230 kHz, ±30 kHz Video Bandwidth Typical: 300 kHz Switchable: 10 Hz Internal Calibrator Frequency: 113.36 MHz ±0.2 MHz
Amplitude Specifications
Amplitude Range: -45 dBmV to 70 dBmV Maximum Safe Input Level Average Continuous Power: +70 dBmV Peak Pulse Power: +70 dBmV dc: 300 Vdc Residual Responses: Input terminated and FS Ref: 0 dBmV 5 MHz to 1.0 GHz: <-40 dBmV Display Range Log Scale: 0 to -45 dB from reference level is calibrated; 1, 2, 5, 10 dB/div; 4.5 divisions displayed. Scale Units: dBmV, dBµV Marker Readout Resolution: 0.1 dB Reference Level Range: -40 to +70 dBmV Resolution: ±1 dB Accuracy: ±0.5 dB Frequency Response: 10 dB input attenuation 1 MHz to 1.0 GHz: ±0.5 dB Calibrator Output Amplitude: ±0.3 dB Display Scale Fidelity Log Maximum Cumulative: 0 to -40 dB from reference level ±0.5 dB Input Return Loss 0 dB Attenuation: >14 dB All Other Settings: >20 dB
Cable TV Measurements
(These specifications describe warranted performance of the instrument over the operatingtemperature range.) Input Configuration: 75 ohms replaceable type F, female Channel Selection: Analyzer tunes to specified channels based upon selected tune configuration. Tune Configuration: standard, off-the-air, HRC, IRC, PAL, SECAM, user-defined Frequency Range: 5 to 1000 MHz (channel mode) Visual Carrier Frequency Resolution: 10 kHz Accuracy: ±25 kHz Visual Carrier Level: The peak amplitude of the visual carrier is measured to an absolute standard traceable to the National Institute of Standards and Technology. Typical Accuracy: ±1 dB Absolute Accuracy: ±1.8 dB (ref level >5 dBmV), ±2.3 dB (ref level <=;5 dBmV) Relative Accuracy: ±0.7 dB relative to adjacent channels in frequency ±1.5 dB relative to all other channels (ref level >5 dBmV) ±2.0 dB relative to all other channels (ref level <=;5 dBmV) Visual-to-Aural Carrier Level Difference (The difference between peak amplitudes of the visual and aural carrier is measured.) Difference Range: 0 to 40 dB Resolution: 0.1 dB Accuracy: ±0.7 dB Digital Channel Power Characteristic Maximum Channel Bandwidth: 8 MHz Additional Error: ±0.2 dB Measurement Speed: 125 mS / 2 MHz of BW Hum/Low Frequency Disturbance (Power line frequency and low frequency disturbance measured on modulated and and/or unmodulated carriers; may not be valid for scrambled channels.) AM Range: 0.5 to 5% Resolution: 0.1% Accuracy: ±0.2%, ±30% of reading Visual Carrier-to-Noise Ratio (C/N)1 C/N Accuracy: ±2 dB for carriers >5 dBmV Range: 50 dBc, 55 dBc with carrier turned off (measured in band) Repeatability: ±1 dB 1 A preselector filter may be required to achieve specifications.
Fiber-Optic Power Meter Option
Wavelength: 1310 and 1550 nm Measurement Range: 1310 nm: +20 to -38 dBm 1550 nm: +18 to -38 dBm Resolution: 0.1 dB Accuracy: (characteristic) ±5% Display Units: dB, dBm, nW, µW, mW Connector Styles: ST, FC, biconic, D4, SMA, bare fiber, rotary splice, RM
General Specifications
Temperature Range Operating: -20 to 55 °C Storage: -20 to 70 °C Power Requirements: +10 to +15 Vdc, 550 mA max Battery Type: 12 V @ 1.9 Ah rechargeable sealed lead acid Usage Time: 4 hours continuous; 8 hours typical intermittent duty Display Area: 5.0 in x 1.33 in (127 mm x 34 mm) Resolution: 240 x 64 pixels Type: LCD with EL backlight Internal Memory Data and Graphic Files: 90 Channel Plans: 4 Reference Traces: 12 total, forward and return Printer Output ofScreen Display Parallel and RS-232
Physical Characteristics
Remote Interface RS-232: 9 pin subminiature D-shell, female Parallel Printer Port (PCL compatible) 25 pin subminiature D-shell, female Weight Net: 4.9 kg (10.7 lb) with battery Shipping: 8.6 kg (19.0 lb) Dimensions: 267 mm H x 318 mm W x 95 mm D 10.5 in. H x 12.5 in. W 3.75 in. D
If you have Adobe's Acrobat Reader already available on your system, please click on the icon provided below to print or view the specifications.
If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader, and then click on the icon below.
HP 85960B HP CaLan 2010 SLM Plus Opt 003 External 12 Vdc/6.5 AH Battery with Case Opt 010 Ingress Measurement Opt 020 Fiber-Optic Power Meter Opt 030 Cable TV Data Management Software Opt 031 Calbe TV Data Management Software Opt 045 Strand Hook Adapter HP 85960U HP CaLan 2010B Upgrade Options Opt R01 2010 to 2010B Upgrade Opt R02 2010 to 3010B Upgrade Opt R03 2010 to 3010R Upgrade Opt R04 2010B to 3010B Upgrade Opt R05 2010B to 3010R Upgrade Opt R10 2010 to 2010B with Ingress Measurement UpgradeDatasheet:sftwr:English:HP85921B
HP CaLan Product Family, Product Overview
If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader now. |
![]() |
Customers outside the U.S. can obtain assistance through local sales offices.
Related Information:
Related HP Cable TV Test Equipment product datasheets:
Related product families available online:
Page Updated: Thursday October 15 19:36:10 UTC 1998