The HP 33330D high-performance planar-doped barrier detector is one of the newest additions to the HP family of high performance detectors. Utilizing a gallium arsenide, planar-doped barrier detecting diode, this detector offers superior performance when compared to Schottky diodes. It features extremely flat frequency response (typically better than ±1 dB to 50 GHz) and very stable frequency response versus temperature.


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Frequency Range:              0.01 to 33 GHz
Frequency Response:           ±0.25 to 14 GHz
                              ±0.4 to 26.5 GHz
                              ±1.25 to 33 GHz
Maximum SWR:                  1.2 to 14 GHz
                              1.4 to 26.5 GHz
                              3.0 to 33 GHz
Low-Level Sensitivity:        >0.4 mV/µW
Maximum Operating Input Power
                              200 mW
Typical Short Term Maximum Input Power (<1 minute)
                              1 W
Video Impedance (nom):        1.5 kohms
RF Bypass Capacitance (nom):  30 pf
Output Polarity:              negative
Input Connector:              3.5 mm (m)
Output Connector:             SMC (m)


General Specifications

Length: 41 mm (1.62 in) Barrel Diameter: 9.7 mm (0.38 in) Input Connector Diameter: 7.9 mm (0.31 in) Net Weight: 14 g (0.5 oz) Shipping Weight: 227 g (8 oz)

Environmental Specifications

Operating Temperature: -65° C to +100° C Temperature Cycling: -65° C to +100° C; MIL-STD 883, Method 1010 (non-operating) Vibration: 0.6 inches D.A. 10 to 80 Hz; 20 g, 80 to 200 Hz MIL-STD 883, Method 2007 Shock: 500 g, 0.5 ms; MIL-STD 883, Method 2002 Accleration: 500 g; MIL-STD 883, Method 2001 Altitude: 50,000 ft (15,240 m);MIL-STD 883, Method 1001 Salt Atmosphere: 48 hr, 5% solution;MIL-STD 883, Method 1009 Moisture Resistance: 25° C to 40° C, 95% Rh;MIL-STD 883, Method 1004 RFI: MIL-STD 461B ESD: 10 discharges at 25 kV to the body, not to the center conductor


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP 33330D     High Performance Planar-Doped Barrier Detector, 0.01 to 33 GHz
   Opt 003    Positive polarity
   Opt UKS    Certificate of calibration

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

To help you make the right instrument choices

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Related product families available online:



Contact HP Test & Measurement. (c) Copyright 1994, 1995, 1996, 1997 Hewlett-Packard Company.

Page Updated: Thursday November 13 06:56:59 UTC 1997