HP 346A Noise Source, 10 MHz to 18 GHz, 6 dB ENR

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]
The ideal companion to HP's noise figure meter and systems is the HP 346A noise source. Since it is broadband (10 MHz to 18 GHz), it eliminates the necessity for several sources at different frequency bands. It has an individually calibrated ENR value at specific frequencies. The calibration is printed on its label for easy loading into the HP 8970B. The low SWR of the noise source reduces a major source of measurement uncertainty, reflections of test signals. The HP 346A noise source is designed especially for accurate characterization of input-impedance-sensitive devices (like GaAsFETs and many UHF amplifiers). Its very small change in reflection coefficient (<0.01) from ON to OFF minimizes errors when measuring noise figure and gain as a function of input impedance.


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Frequency Range:              10 MHz to 18 GHz
Excess Noise Ratio (ENR) Limits:
                              5 to 7 dB
Maximum SWR (reflection coefficient) On and Off:	
     10 to 30 MHz:            1.3 (0.13)
     30 to 5000 MHz:          1.15 (0.07)
     5 to 18 GHz:             1.25 (0.11)
Power Required:               28 ±1 Vdc
Size:                         21 mm W x 140 mm H x 30 mm D
                              (0.8 in x 5.5 in x 1.2 in)
Standard Connector:           APC to 3.5 (male)


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP 346A     Noise Source, 10 MHz to 18 GHz, 6 dB ENR
   Opt 001  Type N (male) connector
   Opt 002  APC-7 connector
   Opt 004  Type-N (female) connector
   Opt 910  Extra operating manual(s)
   Opt W30  3 Year Return Repair Service
   Opt W32  3 Year Calibration Customer Return Service
   Opt W34  3 Year Standards Compliant Calibration Service
   Opt W50  5 Year Return Repair Service
   Opt W52  5 Year Customer Return Calibration Service
   Opt W54  5 Year Standards Compliant Calibration Service


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader now. Get Acrobat

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

To help you make the right product choices

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


Related HP 346 Broadband Noise Source Family product datasheets:

Related product families available online:



Privacy Statement    Use of this site indicates that you accept the Terms of Use
Contact HP Test & Measurement. (c) Copyright 1994 - 1998 Hewlett-Packard Company.