HP 347 Noise Source Family

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]
The HP R347B and HP Q347B Noise Sources cover the frequency ranges of 26.5 to 40 GHz and 33 to 50 GHz. These waveguide noise sources complement the HP 8970S/V Noise Figure Measurement Systems and the HP 8970B Noise Figure Meter, allowing you to make accurate and convenient noise figure measurements on millimeter-wave devices. The HP R347B and Q347B provide highly precise broadband noise at the input of the system or component under test. The noise figure meter then processes the ON/OFF ratio of noise power present in the system IF, and provides an accurate reading of noise figure and gain.


This product family includes


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Frequency Range
HP R347B:                     26.5 to 40 GHz
HP Q347B:                     33 to 50 GHz

Excess Noise Ratio (ENR) Range
HP R347B (26.5 to 40 GHz:     10 to 13 dB
HP Q347B:
   33 to 42 GHz:              10 to 13 dB
   42 to 50 GHz:              6 to 12.5 dB

Maximum SWR (reflection coefficient):
HP R347B
   Source Off:                <1.42 (0.17)
   Source On:                 <1.31 (0.13)
HP Q347B
   Source Off:                <1.57 (0.22)
   Source On:                 <1.46 (0.19)			

Power Required:               28 V ±1 V (automatically supplied by the
                              HP 8970B Noise Figure Meter

Maximum Reverse Power:        100 mW

Connectors
Bias:                         BNC (f)
Noise Output
   HP R347B:                  R-Band Waveguide
                              UG-599/U Flange
   HP Q347B:                  Q-Band Waveguide
                              UG-383/U Flange

ENR Variation with Temperature:
                              <0.009 dB/°C

ENR Variation with Voltage:   <0.005 dB/V

Switching Speed:              <10 µs


General Characteristics

Operating Temperature: 0 to 55 °C Dimensions HP R347B: 52 mm x 108 mm x 60 mm (2.1 in x 4.3 in x 2.4 in) HP Q347B: 53 mm x 108 mm x 62 mm (2.1 in x 4.3 in x 2.5 in) Net Weight HP R347B: 0.21 kg (0.46 lb) HP Q347B: 0.23 kg (0.51 lb) Shipping Weight HP R347B: 0.36 kg (0.80 lb) HP Q347B: 0.38 kg (0.85 lb)

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

No literature is available online at this time.

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP R347B     R-Band Noise Source
HP Q347B     Q-Band Noise Source


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

To help you make the right product choices

Related product groups available online:



Privacy Statement    Use of this site indicates that you accept the Terms of Use
Contact HP Test & Measurement. (c) Copyright 1994 - 1998 Hewlett-Packard Company.