HP 34908A 40 Channel Single-Ended Multiplexer

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]
Use the HP 34908A for the greatest density in common-low applications, such as battery test, component characterization, and benchtop testing.

Each module switches 40 one-wire inputs. All two-wire internal measurements except current are supported. The module low connection is isolated from earth and can float up to 300V.


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


Module Specifications
The HP 34970A accuracy specifications already include the switching offset and reference junction errors shown below. These errors are listed separately for determining system error with external measurement devices. Up to three modules, in any combination, can be inserted into a single mainframe. The HP 34970A's internal DMM connections are accessible only through the HP 34901A, HP 34902A, and HP 34908A low-frequency multiplexers. On-module screw terminals accept wire sizes from 16 gage to 22 gage. 20 gage wire is recommended for high channel count applications.

HP 34908A 40-Channel
Single-Ended Multiplexer
GENERAL
Number of Channels 40
1- wire
Connects to Internal DMM Yes
Scanning Speed 60 ch/s
Open/Close Speed 70/s
INPUT
Voltage (dc, ac rms) 300V
Current (dc, ac rms) 1A
Power (W, VA) 50W
DC CHARACTERISTICS
Offset Voltage[1] <3µV
Initial Closed Channel R[1] <1 Ohm
Isolation ch-ch, ch-earth
>10 G Ohm
AC CHARACTERISTICS
Bandwidth 10 MHz
ch-ch Cross Talk -18 dB
Capacitance HI - LO
<50 pF
Capacitance LO -Earth <80 pF
Volt-Hertz limit 108
OTHER
T/C Cold Junction Accuracy[1] (typical) 0.8°C
Switch Life No Load (typical) 100M
Switch Life Rated Load (typical)[2] 100k
Temperature Operating 0°C to 55°C
Temperature Storage -20°C to 70°C
Humidity (non-condensing) 40°C / 80% RH
[1] Errors included in DMM measurement accuracy specifications
[2] Applies to resistive loads only


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP 34908A       40 Channel Single-Ended Multiplexer

Related Modules HP 34901A 20 Channel Multiplexer (2/4 -wire) HP 34902A 16 Channel Multiplexer (2/4 -wire) HP 34903A 20 Channel Actuator/GP Switch HP 34904A 4 x 8 Matrix HP 34905A 2 GHz Dual 1:4 RF Mux, 50 ohm HP 34906A 2 GHz Dual 1:4 RF Mux, 75 ohm HP 34907A Multifunction Module

Available accessories for this product include:


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

No literature is available online at this time.

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]
Additional help is a click away!

Want to talk to a knowledgeable engineer about this product at our HP Test and Measurement Call Center? Do you have a question you want to send in by e-mail? A measurement set-up you need to fax in? Follow the link above and you will be given a choice to contact an engineer by telephone, e-mail or fax.

This service is available only in the United States. For assistance outside the U.S., contact your local HP DIRECT or HP Sales Office.

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


Related HP 34970A Data Acquisition/Switch Unit product datasheets:

Related product families available online:



Privacy Statement    Use of this site indicates that you accept the Terms of Use
Contact HP Test & Measurement. (c) Copyright 1994 - 1998 Hewlett-Packard Company.

Page Updated: Friday February 05 09:13:42 UTC 1999