HP 35670A 2- or 4-channel DC -102.4 kHz FFT Dynamic Signal Analyzer![]()
|
Frequency
Maximum Range 1 Channel Mode 102.4kHz, 51.2 kHz (option AY6 - single channel maximum range extends to 102.4 kHz without anti-alias filter protection.) 2 Channel Mode 51.2 kHz 4 Channel Mode 25.6 kHz (option AY6 only) Spans 1 Channel Mode 195.3 µHz to 102.4 kHz 2 Channel Mode 97.7 µHz to 51.2 kHz 4 Channel Mode 48.8 µHz to 25.6 kHz (option AY6 only) Minimimum Resolution 1 Channel Mode 122 µHz (1600 line display) 2 Channel Mode 61 µHz (1600 line display) 4 Channel Mode 61 µHz (800 line display) (option AY6 only) Maximum Real-Time Bandwidth (FFT Span for Continuous Data Acquisition) (Preset, Fast Averaging) 1 Channel Mode 25.6 kHz 2 Channel Mode 12.8 kHz 4 Channel Mode 6.4 kHz (option AY6 only) Measurement Rate (Typical) (Preset, Fast Averaging) 1 Channel Mode > 70 Averages/Second 2 Channel Mode > 33 Averages/Second 4 Channel Mode > 15 Averages/Second (option AY6 only) Display Update Rate Typical > 5 Updates/Second (Preset, Fast Average OFF) Maximum > 9 Updates/Second (Preset, Fast Average Off, Single Channel, Single Display, Undisplayed Trace Displays set to Data Registers) Accuracy ±30 ppm (.003%)
Amplitude/Phase
Absolute Amplitude Accuracy (FFT) (A combination of Full Scale Accuracy, Full Scale Flatness, and Amplitude Linearity.) ±2.92% (0.25dB) of Reading ±0.025% of Full Scale FFT Full Scale Accuracy at 1 kHz (0 dBfs) ±0.15 dB (1.74%) FFT Full Scale Flatness (Relative to 1 kHz) ±0.2 dB (2.33%) FFT Amplitude Linearity at 1 kHz: ±0.58% (0.05dB) of reading ±0.025% of full scale (Measured on +27 dBVrms range with time avg, 0 to -80 dBfs) Amplitude Resolution (with averaging) 16 bits less 2 dB over-range 0.0019% of full scale (typical) FFT Cross-Channel Gain Accuracy: ±0.04 dB (0.46%) (Frequency Response Mode) Cross-Channel Phase Accuracy 0.5 deg
Dynamic Range
Spurious Free Dynamic Range: 90 dB typical(<-80 dBfs) (Includes Spurs, Harmonic Distortion, Intermodulation Distortion, Alias Products). 800 Line Display. Source impedence: 50 Ohm. Harmonic Distortion: <-80 dBfs Intermodulation Distortion: <-80 dBfs Spurious and Residual Responses: <-80 dBfs Frequency Alias Responses 2.5% to 97.5% of the Frequency Span <-80 dBfs Lower and Upper 2.5% of Frequency Span <-65 dBfs Input Noise Level (Flat Top Window, -51 dBVrms range, Source Impedance = 50 Ohm) Above 1280 Hz: <-140 dBVrms/ sq.root Hz 160 Hz to 1280 Hz: <-130 dBVrms/ sq.root Hz
Input
Input Ranges: +27 dBVrms (31.7 Vpk) to -51 dBVrms (3.99 mVpk) in 2 dB steps (full scale) Maximum Input Levels: 42 Vpk Input Impedance: 1 M Ohm ±10% 90 pF nominal AC Coupling Rolloff: <3 dB rolloff at 1Hz (Source Impedance = 50 Ohm) Common Mode Rejection Ratio: Single Tone at orbelow 1 kHz -51dBVrms to -11 dBVrms Ranges: >75 dB typical -9 dBVrms to +9 dBVrms Ranges: >60 dB typical +11 dBVrms to +27 dBVrms Ranges: >50 dB typical Common Mode Range: ± 4 V pk (floating mode) IEPE Signal Conditioning Current Source: 4.25 ± 1.5 mA Open Circuit Voltage: +26 to +32 Vdc Crosstalk Between Input Channels, and Source-to-Input <-135 dB below signal or <-80 dBfs of receiving channel, whichever response is greater in amplitude
Trigger/Tachometer
Trigger Modes: Internal, Source, External (analog setting) HP-IB Maximum Trigger Delay Post Trigger: 8191 seconds Pre Trigger: 8191 sample periods External Trigger Range Low Range: -2 V to +2 V High Range: -10 V to +10 V Pulses per Revolution: 0.5 to 2048 RPM Range 5 =< RPM =< 491,519 Accuracy: ±100 ppm (0.01%) (Typical) Tach Level Range Low Range: -4 Vto +4 V High Range: -20 V to +20 V Maximum Tach Pulse Rate: 400 kHz (typical)
Source Output
Source Types: Sine, Random Noise, Chirp, Pink Noise, Burst Random, Burst Chirp Amplitude Range AC: ±5 V peak* DC: ±10 V* * Vacpk + |Vdc| =/< 10 V AC Amplitude Resolution Voltage => 0.2 Vrms: 2.5 mVpeak Voltage < 0.2 Vrms: 0.25 mVpeak DC Offset Accuracy: ±15 mV ± 3% of (|DC| + Vacpk) Settings Output Impedance: < 5 Ohm Sine Amplitude Accuracy at 1 kHz: ±4% (0.34 dB) of setting (Rload >250 Ohm, 0.1Vpk to 5Vpk) Harmonic and Sub-Harmonic Distortion and Spurious Signals (In Band) (0.1 Vpk to 5 Vpk Sine Wave) Fundamental <30 kHz: <-60 dBc Fundamental >30 kHz: <-40 dBc
Options
COMPUTED ORDER TRACKING - Option 1D0 Maximum Order x Maximum RPM -------------------------------- =< 102,400 Hz 60 Online (Real Time) 1 Channel Mode 25,600 Hz 2 Channel Mode 12,800 Hz 4 Channel Mode 6,400 Hz Capture Playback 1 Channel Mode 102,400 Hz 2 Channel Mode 51,200 Hz 4 Channel Mode 25,600 Hz Number of Orders: =<200 Delta Order: 1/128 to 1/1 Resolution (Maximum Order) / (Delta Order): = < 400 Maximum RPM Ramp Rate: 1000 RPM / second (typical) 1000 - 10,000 RPM Run Up Maximum Order = 10 Delta Order = 0.1 RPM Step = 30 (1 Channel) = 60 (2 Channel) = 120 (4 Channel) Order Track Amplitude Accuracy: ±1 dB (typical) Real Time Octave Analysis - Option 1D1 Standards Conforms to ANSI Standard S1.11 - 1986, Order 3, Type 1-D, Extended and Optional Frequency Ranges Conforms to IEC 651-1979 Type 0 Impulse , and ANSI S1.4 Frequency Ranges (at centers) Online (Real Time): Single Channel 2 Channel 4 Channel 1/1 Octave 0.063 - 16 kHz 0.063-8 kHz 0.063 - 4 kHz 1/3 Octave 0.08 - 40 kHz 0.08 - 20 kHz 0.08 - 10 kHz 1/12 Octave 0.0997 - 12.338 kHz 0.0997 - 6.169 kHz 0.0997 - 3.084 kHz Capture Playback: 1/1 Octave 0.063 - 16 kHz 0.063 - 16 kHz 0.063 - 16 kHz 1/3 Octave 0.08 - 31.5 kHz 0.08 - 31.5 kHz 0.08 - 31.5 kHz 1/12 Octave 0.0997 - 49.35 kHz 0.0997 - 49.35 kHz 0.0997 - 49.35 kHz One to 12 octaves can be measured and displayed. 1/1-, 1/3-, and 1/12-octave true center frequencies related by the formula: f(i+1)/f(i) = 2^(1/n); n=1, 3, or 12 Where 1000 Hz is the reference for 1/1, 1/3 Octave, and 1000*2^(1/24) Hz is the reference for 1/12 octave. The marker returns the ANSI standard preferred frequencies. Accuracy 1 Second Stable Average Single Tone at Band Center: =/< ± 0.20 dB (Readings are taken from the Linear Total Power Spectrum Bin . It is derived from sumof each filter.) 1/3-Octave Dynamic Range: >80 dB (typical) per ANSI S1.11-1986 (2 Second Stable Average, total power limited by input noise level) Swept Sine Measurements - Option 1D2 Dynamic Range: 130 dB Arbitrary Waveform Source - Option 1D4 Amplitude Range AC: ±5 V peak* DC: ±10 V* * Vacpk + |Vdc| <=10V Record Length # of Points = 2.56 x Lines of Resolution, or # of Complex Points = 1.28 x Lines of Resolution DAC Resolution 0.2828 Vpk to 5 Vpk 2.5 mV 0 Vpk to 0.2828 Vpk 0.25 mV
Abbreviations
dBVrms = dB relative to 1 Volt rms. dBfs = dB relative to full scale amplitude range. Full scale is approx. 2 dB below ADC overload. Typical = typical, non-warranted, performance specification included to provide general product information.
HP 35670A DC-102.4 kHz FFT Dynamic Signal Analyzer Opt 0B0 Delete manual set Opt 0B1 Add manual set Opt 0B3 Service manual Opt 0BU Additional instrument BASIC manuals Opt 1BH General export license version Opt 1BP Mil std 45662A calibration w/ test data Opt 1C2 Add HP instrument BASIC Opt 1D0 Computed order tracking Opt 1D1 Real time octave measurements Opt 1D2 Swept sine measurement Opt 1D3 Curve fit/synthesis Opt 1D4 Arbitrary source Opt 1F0 PC-style keyboard - U.S. version Opt 1F1 PC-style keyboard - German version Opt 1F2 PC-style keyboard - Spanish version Opt 1F3 PC-style keyboard - French version Opt 1F4 PC-style keyboard - U.K. version Opt 1F5 PC-style keyboard - Italian version Opt 1F6 PC-style keyboard - Swedish version Opt AB1 Korea - Korean localization Opt AB2 China - Chinese localization Opt ABD Germany - German localization Opt ABE Spain - Spanish localization Opt ABF France - French localization Opt ABJ Japan - Japanese localization Opt ABZ Italy - Italian localization Opt AN2 Add 4 MB RAM Opt AX4 Rack flange kit Opt AY6 Add 2 input channels Opt UFC Add 8 MB RAM Opt UFF Add 1 MB nonvolatile RAM Opt UK4 Microphone adapter and power supply Opt UK5 Carrying case Opt UK6 Commercial cal. certificate w/ test data Opt W30 3 Year Return Repair Service Opt W32 3 Year Cust Return Calibration Service Opt W34 3 Year Standards Compliant Cal Service Opt W50 5 Year Return Repair Service Opt W52 5 Year Cust Return Calibration Service Opt W54 5 Year Standards Compliant Cal Service
HP 35670A Product Overview
( PDF)
HP 35670A Technical Specifications
( PDF)
If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader now. |
![]() |
Related product families available online:
Page Updated: Friday February 12 00:58:01 UTC 1999