HP 3567A PC Spectrum/Network Analyzer

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]
A rich measurement set makes the HP 3567A excellent for mechanical test, signal characterization, control systems, rotating machinery analysis and production test where signals are below 102.4 kHz. This PC based dynamic signal analyzer is configurable from 2 to 48 channels, and uses modular HP 3565S hardware to achieve high system performance. Since the HP 3567A is a Microsoft Windows® application, measurement results are easily shared with other Windows applications, such as spreadsheets, word processors, and the HP 35639A Data Viewer.

The HP 3567A includes a source for stimulating circuits or systems. For fast measurement processing, a powerful hardware signal processor module converts time data to frequency domain data using the latest FFT technology. The included HP 35653C source module provides internal calibration signals and sine, noise, and pulse test stimulus.

An optional HP 35659A, SCSI interface module adds high speed transient capture, analysis, and playback capability (with HP 35656B). The optional HP 35656B, programmable DAC module also adds arbitrary waveform and chirp capability. The optional HP 35658A Tachometer/Trigger module improves order analysis.

Three optional software packages enhance the already complete measurement capability with Order Analysis, Swept-Sine, and Real-Time Octave analysis.

If portability is important, the 4 slot HP 35650B mainframe and a portable PC make very a cost effective multi-channel system.

See also the HP 3566A PC Spectrum/Network Analyzer.

Windows is a U.S. registered trademark of Microsoft Corp.


Available modules include:


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Frequency

Measurement Range                   64 µHz to 102.4 kHz

Sample Rate (per channel)           262,144 samples/sec

Spans                               195 mHz to 102.4 kHz

Frequency Resolution                25 lines to 3200 lines

Maximum Real-Time                   2 channels at 25.6 kHz 
Bandwidth                           (fast averaging, display off)

Frequency Accuracy                  ± 8 ppm

Amplitude/Phase

Absolute Amplitude 
Accuracy (FFT)                      0.15dB
(A combination of Full Scale Accuracy, 
Full Scale Flatness, and Amplitude Linearity.)

Cross-Channel Gain Accuracy:        ± 0.1 dB 
(Frequency Response)

Cross-Channel Phase Accuracy        ± 0.5 deg

Dynamic Range

FFT Spurious Free Dynamic Range:    80 dB (85 dB typical) 

Input Noise Level(Flat Top Window, most sensitive range, 
Source Impedance = 50 Ohm)
20 Hz to 200 Hz:                    <565/sq.root(f)
(frequency dependent)               nVrms/sq.root(Hz)

Above 200 Hz:                       <40nVrms/sq.root Hz
(frequency independent)

Input

BNC
Input range (full scale)            1.26 mVpk to 39.8 Vpk 
                                    in 2 dB steps

Maximum  Input Levels:              42 Vpk

IEPE Signal Conditioning
     Current Source:                4 mA 
     Open Circuit Voltage:          >20  Vdc

Connections                         Grounded or Floating

Coupling                            AC or DC

AC coupling roll off                <3dB at 1 Hz

Input impedance                     1 Mohm ± 5%

Charge
Gain (at 1 kHz)                     ± 0.3 dB

Flatness (to 51.2 kHz)              ± 0.1 dB

Phase (to 51.2 kHz)                 ± 0.5°

Trigger

From Input channel
Internal                            Positive or negative slope
Trigger level resolution            ± 0.031% of full scale range 
     
Maximum Trigger Delay
Post Trigger:                       0 to 104,575 samples 
Pre Trigger:                        0 to 8,191 samples

From HP 35658A (Tachometer/Trigger Module)
External Trigger Range     
Range:                              -23 V to +23 V
Resolution:                         <5V: 40 mV
                                    >5V: 200 mV

Input impedance                     >20 kohms (typical)

Source/DAC Output

HP35653A/C

Max output level                    10 V

Max Current                         50 mA

Max capacitive load                 0.01 uF

Min Frequency resolution            15.6 mHz

Max Frequency                       102.4 kHz (Hp 35653C)
                                    51.2 kHz (HP 35653A)

HP35658A (optional) 

Max Frequency Clock                 131,072 Hz set in 2x steps

Anti-Alias filter                   200 Hz to 51.2 kHz in 200 Hz steps

Amplitude Resolution                16 bits

Harmonics                           -68 dBc (<2 kHz)

Max DAC buffer                      7.5 Msamples (with OPT 116)

Options

Computed Order Tracking - with HP 35636A software

Amplitude Accuracy                  ± 1dB

Phase accuracy  RPM Range           1st order       2-5 order
                10 to 10,000        ± 0.5°          ± 1° 
                10,000 to 120K      ± 1°

Ramp rates                          3,000 RPM/sec
(4 input channels, 600-6000 RPM, 
delta order= .1, 30 RPM Step, 4 pulses/rev)

Swept Sine Measurements - with HP 35657A software

Dynamic Range:                      132 dB 

Maximum number of points:           32,767

Real-Time Octave Analysis - with HP 35658A software

International specification
compliance                          ANSI S1.11-1986 Order 3, Type 1D
                                    IEC 225-1986 Accuracy Class 1
                                    IEC 651 SLM Accuracy Type 1
                                    DIN 45651
                                    and others

1/3 Octave Frequency coverage
     On-line                        100 mHz - 20 kHz
     Off-line                       100 mHz - 80 kHz

Frequency resolution                1/1, 1/3, 1/12 octave

Dynamic range (ANSI S1.11-1986)     80 dB

Amlitude accuracy                   ± 0.4 dB

Band to band flatness               ± 0.4 dB

Linear Average range                31.25 mSec - 64,000 sec

Specification Abbreviations

dBVrms = dB relative to 1 Volt rms.

dBfs = dB relative to full scale amplitude range.  
Full scale is approx. 2 dB below ADC overload.

Typical = typical, non-warranted, performance 
specification included to provide general 
product information.


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP 3567A       PC Spectrum/Network Analyzer
   Opt 005     Add l HP 35650A Mainframe
               (Includes HP 35650A Opt. 715,
               Mainframe Interconnect Cable Kit)
   Opt 010     Add l HP 35652B 1-Channel Input Module
               (maximum of 48 HP 35652B modules)
   Opt 116     Convert HP 35651C RAM to 16 MB
   Opt 054     Replace HP 35651C with 35654B DSP Module
   Opt 056     Add HP 35656B Programmable DAC Module
               (required for chirp, arb, and record/playback)
   Opt 050     Replace HP 35650A Mainframe with Portable,
               4-slot, Non-expandable 35650B
   Opt 058     Add l HP 35658A Tachometer/Trigger Module
   Opt O59     Add HP 35659A SCSI Interface for External
               SCSI Disk (maximum is 1 HP 35659A)
   Opt 060     Add HP 35659A SCSI Interface with Internal 2 GB Disk
               (maximum is 1 HP 35659A)
HP 35636A      Order Tracking (optional software)
HP 35637A      Swept-Sine (optional software)
HP 35638A      Real-Time Octave (optional software)
HP 35634A      Software only for HP 3566A and 3567
               (order only if you already have the hardware)


All orders include the following:


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP 3566A and HP 3567A Product Overview ( PDF)
HP 3566A and HP 3567A Techncial Specifications ( PDF)

If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader now. Get Acrobat

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

To help you make the right product choices

Related product groups available online:



Privacy Statement    Use of this site indicates that you accept the Terms of Use
Contact HP Test & Measurement. (c) Copyright 1994 - 1998 Hewlett-Packard Company.

Page Updated: Friday February 12 00:51:18 UTC 1999