HP 3587S Signal Analysis System![]()
|
Successful signal analysis requires more than just measurement performance, however. You need signal processing tools and a wide selection of display types and presentation formats to characterize a signal fully. The HP 3587S provides all these in a commercial-off-the-shelf package.
Available modules include:
Measurement ControlProduct Functionality
Resolution Windows, Bins Averaging: RMS, Peak, Nth OverlapModes: Run, Pause/Continue, Arm Status Indicators: Overload, Triggered, External reference, GapMeasurement ResultsTypes: Frequency, Time, Phase*, Amplitude * Phase display is relative to the beginning of the data block. Data is not corrected for trigger jitter, digital filter phase error, or local oscillator phase error.MarkersMode: Off, Single, Relative (same trace, separate trace) Functions: Marker to Peak, Marker to Next Peak Right/Left, Band Power, NoiseDisplayFormat: Single, Dual, Triple, Overlay, State Active Trace: A, B, C, AB, BC, AC,ABC Standard Display Types: Spectrum, Time, Phase Advanced Display Types: Spectrogram, Rollogram, Spectral Map/ Color Map/ Time Map, azimuth, elevation, threshold, height, scroll direction, hidden line, wireframe, Histogram, PDF, CDF, Strip Chart Trace Coordinates: Y axis: log magnitude, linear magnitude, dB magnitude, dBm magnitude, real, imaginary, phase X axis: linear frequency, linear time, volts Trace Label Units: Peak/RMS Volts/Volts squared Volts/Eng Units Trace Lable Scaling: Y axis: auto scale, Y range, Y reference X axis: X reference, X magnify, X default Engineering Units Threshold Display Memory Manipulation: scroll up, scroll down, home, end Color Configuration Trace line, Trace grid, Trace background, Display BackgroundUser InterfaceInput Devices: Keyboard, Mouse, 9-knob panel, 32-button panelGeneralOn-line Help Memory and data storage: Save/Recall, Record/Playback Optional Functionality Option AGG: ProgrammingHardware Specifications
Specifications describe warranted performance over the temperature range of 0 deg. to 50 deg. C (except where noted), after a 30-minute warm-up from ambient conditions, for the system configuration listed. Supplemental characteristics identified as "typical" or "characteristic," provide useful information by giving non-warranted performance parameters. Typical perfomiance is applicable from 20 deg. to 30 deg. C. For more detailed specifications refer to the technical data sheets of the individual system components. Configuration The performance of a system depends on its hardware and software components. This technical data is based on the following configuration. Any changes in this configuration may effect system performance. Configuration data HP E1498A with 64 Mbytes RAM ADC: HP E1430A with opt AYD, 10.24 MHz time base DSP: HP E1485A with four option 1FL, 96002 DSP Down Converter: None System Disk: HP E4208A (500 MByte) Mainframe: HP E1401A (13slot, C size) Monitor: HP A4032A (17 inch, color) Operating Software: 35687B Signal analysis software
Performance
Real-Time Bandwidth: 1.0 MHz (801 lines, 0% overlap, spectrogram display, rms averaging, 16-bit word, 1024 x 768 pixel display) SIGNAL CAPTURE BUFFER (typical): 2048 spectra, gap free (4 MHz span, 801 lines, 16 bit word width) Display Update Rate (typical) Spectrogram: 60/s Waterfall: 30/s Color Waterfall: 30/s Frequency Trace: 30/s Time Trace: 30/s Map Rotate*: 2 Repaint/s * 100 Spectra, 1024 x 768 pixel display.
Baseband Specifications
Frequency Range: dc to 4 MHz Spans:0.24 Hz to 4.0 MHz, octave steps, includes frequency translation (zoom) capability for spans < 4 MHz Tuning resolution: 10 Hz, span < 4 MHz Frequency resolution: 51 to 12,801 lines Bin width: 18.6 Hz to 78.4 kHz Window factors: Uniform, Hann, Flattop, Gausstop, Blackman, Gaussian Amplitude Input impedance: 50 Ohms ADC resolution (raw): 23 bits Input ranges: + 28 dBm to - 32 dBm, 6 dB steps Measurementrange: + 28 dBm to - 142 dBm Sensitivity: - 159 dBm/Hz (- 32 dBm range) Noise figure: 15 dB Damage level: > + 31 dBm Coupling: ac/dc Autorange: Off/up/single Autozero: Single by command ADC sample rate: 10.24 MSa/s ADC clock accuracy: 10.24 MHz +/- 100 Hz, < 10 ps jitter Anti-alias filter: 4 MHz, low pass Digital filters: 0.24 Hz to 2.0 MHz, octave steps, includes frequency translation (zoom) capability Dynamic Range Spurious free dynamic range:< 110 dbfs Spurious signals (includes alias products): < - 110 dbfs, using internal clock Harmonic distortion: < - 110 dbfs or - 80 dBc, whichever is greater Input Noise: (For 1 kHz < fo < 4 MHz) Range Noise Density +28 dBm to -16 dBm < -136 dBfs/Hz -22 dBm to -32 dBm < -127 dBfs/Hz
General Specifications
Signal Capture Mode: On/Off Buffer Size:4MSamples (4 Msamples = 2048 spectra, 801 line, 4 MHz span, 16 bit word width) Buffer Length (seconds): 4 Msamples / (span x 2.5) Averaging Mode: Off, rms, Peak, Nth Number of Averages: 1 to 32,767 Overlap Percentage 99% (span < 125 kHz), 0 to 50% (125 kHz > span >500 kHz), 0% (span > 500 kHz) (Overlap processing does not require averaging) Memory Save/Recall State: 32 registers Trace: 32 registers Data: 32 registers Record/Playback Mode: Off/Playback/Record Triggering Modes: Freerun, Level, Magnitude, External Level: + 1 00% of input Slope: Positive, Negative Delay: - 4,194,303 to + 33,554,432 samples Arm: Auto/Manual External Clock Input* Frequency Range: 100 kHz to 10.24 MHz Input Levels: TTL Accuracy: Source accuracy + 10 ps rms jitter Printer Output Modes: Print Screen/Print Trace Printers: HP Paintjet, HP Laserjet * Performance specifications valid for internal sample clock only. Spurs and noise of greater than -110 dBc on external clock signal will degrade the performance of the ADC.
Physical Specifications
(Based on configuration in Table 1. VXI module power consumption, weight, and size are included with VXI chassis) Environmental VXI Chassis Monitor Disk Drive (HP E1401A) (HP A4032A) (HPE4208A) Operating 0 to 55 deg C 0 to 40 deg C 5 to 45 deg C Temperature Operating 75% 10% to 80% 10 to 80% Humidity (40 deg C) (40 deg C) (40 deg C) (non-condensing) Storage -40 to 75 deg C -40 to 60 deg C -20 to 60 deg C Temperature Storage Humidity 75% 5 to 95% 5% to 95% (non-condensing) (40 deg C) (65 deg C, 24 hrs) Altitude 4600 m 15333 m n/a (15,000 ft) (50,000 ft) Weight 22 kg (47 lbs.) 22 kg (47 lbs.) n/a Power (110/120 VAC, 12A 2.7 A n/a 50-60 Hz) (220/240 VAC, 6 A 1.5 A 50-60 Hz) Size (H x W x D) mm 323 x 426 x 602 414 x 406 x 450 1 slot inch 12.7 x 16.8 x 23.7 16.7 x 16 x 17.7 C size VXI
HP 3587B HP 3587S Signal Analysis System Opt 001 Delete 20" Monitor Opt 421 Replace 13-slot VXI Mainframe with 6-slot Opt 428 Replaces VXI 2GB system disk with 1GB disk. Opt 430 Adds HP 89430A Downconverter (2 GHz to 1.8 GHz) Opt 431 Adds HP 89431A Downconverter (2 GHz to 2.65 GHz) Opt 900 240V (UK) Dial & Button Box Opt 902 220V (Europe) Dial & Button Box Opt 903 120V (N. America) Dial & Button Box Opt 62C Adds one HP E1562C 4 GByte data disk Opt ABA Localization Kit U.S. Opt ABC Localization Kit - Canada Opt ABD Localization Kit - Germany Opt ABF Localization Kit - France Opt ABJ Localization Kit - Japan Opt ABU Localization Kit - U.K. Opt AGG Development version Opt ATR Real time signal capture software
The HP 3587B includes: 13 slot VXI mainframe 100 MHz VXI controller w/128M RAM VXI 2GB system disk w/DAT 20" display ADC (HP E1430) DSP (HP E1485) w/12 M RAM and 5-96K DSP HP-UX Real Time Signal Analysis Software and documentation
HP 3587S Technical Specification