HP OmniBER 717 Communications Performance Analyzer

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

The HP OmniBER 717 communications performance analyzer offers a one box tester, field portable tester for installation, maintenance, and commissioning of today's hybrid PDH/SDH and ATM transport networks and network equipment.

Additionally the instrument can be configured as a dual standard SONET/SDH analyzer which makes it the ideal tool for testing network equipment in a manufacturing environment. The dual SONET/SDH capability along with comprehensive remote control of the instruments rich feature set means that production lines can be rapidly reconfigured to test either SONET or SDH. Training costs and program development times are reduced as there is only one easy to use interface to learn.

Optical 1310nm and 1550nm interfaces at 51Mb/s( STM-0/OC-1), 155Mb/s (STM-1/OC-3) and 622Mb/s(STM-4/OC-12) are supported as well as electrical interfaces at all the commonly used telecom rates. PDH interfaces from 704kb/s to 140Mb/s (E4) and DSn interfaces at 1.5Mb/s (DS1) and 45Mb/s (DS3) are supported. The measurement capability is configured using a modular architecture, allowing flexible configurations and full upgradeability, thus protecting your investment in test equipment.

The instrument can be configured to include DSn, PDH, SONET, SDH, ATM and jitter; or it can be configured to have any one of these technologies with or without jitter generation and measurement.

Use the HP OmniBER 717 to perform non intrusive monitoring of live traffic or test network's protection switching mechanisms, perform out-of-service BER measurements or fully test ATM physical and cell layers and B.ISDN protocol stack layers. Functionally test PDH/DSn/SDH/SONET multiplexors, regenerative repeaters, digital cross connects, etc., as well as ATM switches, routers and rate converters.

Stress network elements by performing frequency offset, auto-jitter tolerance, auto-jitter transfer and output jitter measurements to ITU-T recommendations. Important parametric tests such as protection switch times, frequency, delay and power measurement are also supported.


Available modules include:


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

General

SONET

SDH

DSn

PDH

ATM


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Specifications depend on test module selection. Refer to technical 
specifications publication no. 5964-2255E. Available in the "Print
Formatted Documentation" section, under "Product Ordering Information." 


SONET

Frame formats: OC-12c, OC-12, OC-3c, OC-3; OC-1, STS-3c, STS-3, 
               STS-1, DS3, DS1 to Bellcore GR.253-CORE.
Payloads:     DS3, E3, E1, DS1 and STS-3c, STS-1, VT-6, VT-2, 
               bulk filled to Bellcore GR.253-CORE.
Drop/Insert:  DS1, E1 to/from SONET signal


Clock Sources

Intenal:      Accuracy ±4.5 ppm
Offset:       Range ±999 ppm
External:     64 kb/s MIS, 1.544 Mb/s BITS, 2 Mb/s MTS signal


Connectors

Optical:      Customer exvhangeable range of adaptors (FC/PC, 
               DIN47526, ST, Biconic, NEC D4, SC, HMS-10/HP)
Electrical:   155 Mb/s (CMI), 52 Mb/s (B325)

Error addition and detection: AIAL frame, B1, B2, REI-L, B3, 
               REI-P,VT BIP, REI-V, payload bit errors.

Alarm generation and detection: LOS,LOF,SEF, AIS-L, RD1-L, 
               AIS-P, RDI-P, LOP-P, Path unequipped, AIS-V,ROI-V, LOP-V, 
               UT unequipped, Loss of VT superframe.

Pointer Sequences: Generation and analysis to ITU-T G.783 and 
               GR.253 plus APS time measurement.



SDH

Frame formats:STM-4, STM-1 to ITU-T G.707, STM-0 to ITU-R F.750. Payloads: ITU-T G.707 (140 Mb/s, DS3, 34 Mb/s, 2 Mb/s, DS1 and C-4, C-3, C-2, C-12 bulk-filled) Drop/insert: 140 Mb/s, 34 Mb/s and 2 Mb/s to/from SDH signal. Clock sources Internal: Accuracy ±4.5 ppm. Offset: Range ±999 ppm. External: 2 Mb/s / 2 MHz MTS signal, received STM-4, STM-1 and STM-0 signal. Connectors Optical: Customer exchangeable range of adaptors (FC/PC, DIN47526, ST, Biconic, NECD4, SC, HMS-10/HP). Electrical: BNC or small Siemens on all transmit/receive and drop/insert ports except STM-1/STM-4 NRZ which is SMA. Interfaces Optical: 622Mb/s, 155 Mb/s (1310nm and 1550 nm). Electrical: 155 Mb/s (CMI), 52 Mb/s (B3ZS), STM-1/STM-4 NRZ (ECL). Error addition and detection: A1A2 frame, B1, B2, MS-FEBE, HP B3, HP FEBE, HP IEC, LP B3/BIP-2, LP FEBE, payload bit errors. Alarm generation and detection: LOS, LOF, OOF, MS AIS, MS FERF, HP AIS, HP FERF, HP LOP, HP unequipped, LP AIS, LP FERF, LP LOP, LP unequipped, H4 LOM. Pointer sequences: Generation and analysis to ITU-T G.783. Protection switching: MSP linear messages to ITU-T G.783 and MSP ring architecture to ITU-T G.841 plus protection switch time measurement.

DSn

FrameFormats DS3: C-Bit parity as per ANSI T1.107 - 1990 M23 as per ANSI T1.107-1995 DS1: SF (D4), ESF as per ANSI T1.403-1989, TR-TSY.000499 and ITU-T Rec. G.704, SLC-96 E3: As per ITU-T G.751 E1: As per ITU-T G.704 Payloads Framed, unframed or structured (full mux-demux) with 64 kb/s, 56 kb/s, n x 64 kb/s, nx 56 kb/s, timeslot testing. Clock sources Internal: Accuracy; ±0.5 ppm for DS3, E3 and E1; ±0.7 ppm for DS1 Offset: Range ±100 ppm External: Clock recovered by receiver (loop timed) Connectors Electrical: Unbalanced DS3 - BNC Balanced DS1 - WECO bantam Unbalanced E3 - BNC Balanced E1 - WECO bantam Unbalanced E1 - BNC DS1 add/drop - WECO bantam E1add/drop - BNC Interfaces Electrical: DS3 (B325), E3 (HDB3), DS1 (AMI or B825), E1 (AMI or HDB3). Error addition and detection Code. Frame, CRC, Parity, FEBE, REBE, Payload. Alarm generation and detection LOS, AIS, LOF, RAI

PDH

Frame formats:140 Mb/s to ITU-T G.754, 34 Mb/s to ITU-T G.751, 8 Mb/s to ITU-T G.742 and 2 Mb/s to ITU-T G.704 (Std and CRC4). Payloads: Framed, unframed or structured (full mux-demux) with 64 kb/s and n x 64kb/s timeslot testing. Drop/Insert: 2 Mb/s to/from 8 Mb/s, 34 Mb/s or 140 Mb/s signal. Clock sources Internal: Accuracy ±4.5 ppm. Offset: Range ±100 ppm. External: Received 2 Mb/s signal, external 2 Mb/s / 2 MHz MTS signal on jitter generator module, external clock input on PDH binary interface module. Connectors Electrical: BNC or small Siemens on all transmit/receive and drop/insert ports. Interfaces Electrical: 140 Mb/s (CMI), 34 Mb/s (HDB3), 8 Mb/s (HDB3/AMI), 2 Mb/s (HDB3/AMI). Error addition and detection: Code, frame, CRC4, REBE and payload bit errors. Alarm generation and detection: LOS, AIS, LOF, RAI, RMFAI, CASMFL.

Jitter

Jitter can be added to and measured on PDH, SDH and ATM signal structures. Rates: STM-4o, STM-1o, STM-1e, 140 Mb/s, 34 Mb/s, 8 Mb/s and 2 Mb/s. Jitter and wander generation: Automatc jitter tolerance to ITU-T G.958 and G.823 with jitter and wander tolerance to ITU-T G.825. Jitter frequency range: 0.1 Hz to 5 MHz Jitter amplitude range: up to 200 UI Wander frequency range: 10uHz to 0.125 Hz Wander amplitude range: up to 14 400 UI Jitter tolerance pass masks: ITU-T G.958 type A and B, ITU-T G.823 low Q and high Q. Jitter measurement: Peak-to-peak and rms output jitter and pointer jitter measurements with internal 12 kHz HP plus HP1, HP2 and LP filters to ITU-T O.171. Jitter transfer: Automatic jitter transfer with narrow bandwidth selective filtering to ITU-T G.823 and G.958. Jitter transfer pass masks: ITU-T G.958 type A and B, ITU-T G.823 low Q and High Q. Wander: Measurement 2 Mb/s wander measurement with time interval error, estmated frame slips, bit slips and implied frequency offset results.


ATM

Frame formats: 2.048 Mb/s (E1): As per ITU-T G.804/G.704: 34.368 Mb/s (E3)and 139.264 Mb/s: Asper IUT-T G.832; 155.52Mb/s (STM-1), to ITU-TG.707.

DS1: ESF to ANSI T1.403 - 1989, Bellcore TR-TSY-000499 and ITU-T G.704. DS3: C-Bit parity to ANSI T1.105 - 1991 and TR-TSY-000253.

Foregrounnd VC payload: Cross cell PRBS (2^15 - 1, 2^23 - 1 to ITU-T 0.151). Single cell PRBS (2^9 - 1). User defined byte repeated to fill cell payload (48 bytes). Test cell (to draft ITU-T O.191).

Error add and detection: E1: FAS, BPV/code, CRC-4, REBE, E3: BPV/code, BIP, STM-1/OC-3c: As per SDH Single HEC or double HEC Payload errors

Alarm generation and detection: Physical layer DS1: LOS, LOF, AIS, RAI. DS3: LOS, LOF, AIS, RAI, FEAC. DS3: PLCP:RAI. E1: LOF, AIS, RAI. E3: LOF, AIS, FERF. OC-3c/STM-1: Asper SDH section

ATM layer: VP-FERF/VP-ROI,VP-AIS, VC-FERF/VC-ROI, VC-AIS (all end-to-end) Continuity check: VP-CC, VC-CC

Additional ATM layer alarms detected: Loss of cell sync, selected cell not received, congestion experienced, test cell loss, payload pattern loss, PM-OAM loss.

Declarations of Conformity


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

As there are certain working restrictions between some modules,
please refer to the configuration guides, available as PDF files
in the "Print Formatted Documentation" section of this page.

HP 37717C communications performance analyzer - mainframe only
             (color display, VGA output, disk drive, supports ATM services)


             PDH/ATM cell (BNC connector options)

   Opt UKJ   140 MB/s, 34 Mb/s, 8 Mb/s and 2 Mb/s (structured full
             mux/demux).
   Opt UKN   ATM cell at 140 Mb/s, 34 Mb/s and 2 Mb/s plus option
             UKJ capability.
   Opt UKK   140 MB/s, 34 Mb/s, 8 Mb/s and 2 Mb/s and 704 kb/s
             (unframed) with framed in-service measurements.
   Opt UKZ   ATM cell at DS3, DS1, E1 (2 Mb/s) and E3 (34 Mb/s).
   Opt UH3   PDH binary interfaces (NRZ), external clock input plus
             Tx clock/data and Rx clock/data

             DSn interfaces

   Opt 110   DS3/DS1/E3/E1 structured test interfaces.

 SDH electrical (BNC connector options)

   Opt A3R   STM-1e, STM-0e, DCC drop/insert, error O/P SDH module.
   Opt A1T   STM-1e, DCC drop/insert SDH module for ATM testing.

             Dual standard SONET/SDH

   Opt 120   155/Mb/s, 52 Mb/s, SONET/SDH overhead and test interface module

             Optical interfaces

   Opt UH1   STM-1o (1310 nm).
   Opt UH2   STM-4o/STM-1o (1310 nm).
   Opt URU   STM-4o/STM-1o (1550 nm).
   Opt USN   STM-4 test (overhead access etc) plus STM-4o/STM-1o
             dual wavelength (1310 nm and 1550 nm)
   Opt UKT   STM-4 test (overhead access etc) plus STM-4o/STM-1o
             single wavelength (1310nm).

   Opt 130   STM-4 test (overhead access etc) plus STM-4o/STM-1o/STM-0o
             dual wavelength (1310 nm and 1550 nm)
   Opt 131   STM-4 test (overhead access etc) plus STM-4o/STM-1o/STM-0o
             single wavelength (1310nm).

             SDH  Binary Interfaces (SMA connectors)

   Opt 0YH   STM-4/STM-1 SDH binary interfaces (NRZ Tx/Rx clock
             and data).

             Jitter (BNC connector options)

   Opt 140   Jitter generation.
   Opt A3K   Jitter and wander generation
   Opt UHN   140 Mb/s, 34 Mb/s, 8 Mb/s and 2 Mb/s peak-to-peak
             jitter measurement.
   Opt A3L   STM-1e peak-to-peak & rms jitter measurement plus option
             UHN capability.
   Opt A3V   STM-1o, STM-1epeak-to-peak & rms jitter measurement plus
             Opt UHN capability
   Opt A3N   STM-4o, STM-1o, STM-1e peak-to-peak & rms jitter measurement
             plus Opt UHN capability

             ATM Services Layer

   Opt 0YK   ATM and AAL services module (Channel View, rate history,
             graphical CDV etc.)
   Opt USL   Ethernet and Token Ring LAN plus option 0YK capability.

             Printer

   Opt UKX   Integrated graphics printer (80 column) - includes screen dump
             capability.

             Optical adaptors

   Opt UH4   FC/PC.
   Opt UH5   DIN 47526.
   Opt UH6   ST.
   Opt UH7   Biconic.
   Opt UH8   NEC D4.
   Opt UKP   SC.
   Opt UKQ   HMS-10/HP.

             Remote control/external printing

   Opt A3D   RS-232-C and GP-IB remote control/external printer.
   Opt A3B   LAN, RS-232-C and GP-IB remote-control/external printer.
   Opt USS   HP distributed network analyzer firmware for HP E4540A.

             Manual and Calibration Certificate

   Opt AVA   Calibration Manual.
   Opt 0B2   Additional Operating Manual.
   Opt 0BF   Remote Operation Manual.
   Opt UK6   Calibration certificate.

             Remote interactive control software (E4540A)

             HP distributed network analyzer software for operating the
             HP analyzer remotely from your PC using MS Windows via a
             graphical user interface. (Also requires Option USS and
             RS-232-C remote-control option A3D orA3B).

 Carrying cases and rack mount kits

HP 15910B    Soft, vinyl carrying case.
HP 15772B    Hard, robust transit case.
HP 15770A    Rack mount kit.

             Accessories

HP 15744A    Optical Coupler.
HP 15772A    Telephone handset (Talk/Listen capability) for option UKJ or UKN.

             Upgrade kits

HP 15777C    Contact your local HP representative to upgrade your
             HP 37717C analyzer at a later date.





[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Ease your network's migration to ATM ( PDF)
Accurate jitter and wander testing ( PDF)
Tributary jitter testing ( PDF)
Automated verification of network equipment ( PDF)
Automated test sequences ( PDF)
Physical layer jitter testing ( PDF)
Testing PDH digital radios ( PDF)
Resolve finger pointing problems ( PDF)
Verifying protection switching mechanism ( PDF)
Global SDH testing at your fingertips ( PDF)
Configuration guide ( PDF)

If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader now. Get Acrobat

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

To help you make the right product choices

Related product groups available online:



Privacy Statement    Use of this site indicates that you accept the Terms of Use
Contact HP Test & Measurement. (c) Copyright 1994 - 1998 Hewlett-Packard Company.

Page Updated: Friday February 12 01:46:15 UTC 1999