HP 4291A RF Impedance/Material Analyzer, 1MHz to 1.8GHz |
The HP 4291A RF impedance /material analyzer provides a total solution for high-accuracy and easy measurement of surface-mount components and dielectric/magnetic materials. The HP 4291A uses a direct current-voltage measurement technique, as opposed to the reflection measurement technique, for more accurate impedance measurement over wide impedance range.
Basic impedance accuracy is +-0.8%. High Q accuracy enables low-loss component analysis. An internal synthesizer sweeps frequency from 1MHz to 1.8GHz with 1mHz resolution. A 1.8m error-less cable connects the analyzer to a test station so you can extend your test point away from the analyzer without losing accuracy. Advanced calibration and error compensation function eliminate measurement error factors in fixture and assure high accuracy and repeatability at DUT/MUT.
The HP 4291A also provides automatic level control and monitor of test signals by using I-BASIC programming function; devices can be measured under aconstant voltage or current. Measure bias-dependent impedance characteristics with optional dc bias.(up to 40V and 100mA).
At the push of a button, the built-in Equivalent Circuit Analysis Function
automatically calculates the circuit constant values of five circuit models.
Operating Frequency:1MHz to 1.8GHz* Impedance Parameters:|Z|,|Y|,phase,R,X,G,B,Cp,Cs,Lp,Ls,D,Q Converted Parameters:Reflection coefficient=|(gamma)|,phase,(gamma)x,(gamma)y Material Parameters(opt.002):Permittivity=|e|,e',e", Permeability=|u|,u',u", Dissipation factor=tan-d=D
Frequency(Hz) | Impedance Accuracy(%) | Phase Accuracy(in radians) |
---|---|---|
1M to 100M | 0.8 | 8m |
200M | 1.0 | 10m |
500M | 1.5 | 15m |
1.0G | 2.5 | 25m |
1.8G | 4.0 | 40m |
Impedance Range:0.1ohm to 50kohm DC bias(opt.001):0 to +-40V, 0 to +-100 mA No. of points per sweep:2 to 801pts. Other Features:Two independent measurement channels, built-infloppy disk drive, limit-line testing,equivalent circuit analysis, and HP IBASIC(opt. 1C2)
* 1MHz to 1.0GHz when using HP 16453A dielectric test fixture.
HP 4291A RF Impedance/Material Analyzer Opt 1D5 Add High Stability Frequence Reference Opt 1C2 Add HP IBASIC,HP-HIL,Keyboard and Cable Opt 001 Add DC bias Opt 002 Add Material Measurement Software Opt 011 Delete High Impedance Test Head Opt 012 Add Low Impedance Test Head Opt 013 Add High Temperature High Impedance Test Head Opt 014 Add High Temperature Low Impedance Test Head Opt W30 Extended Repair Service Opt W32 Calibration ServiceHP4291A 1.8GHz Impedance/Material Analyzer Data SheetAccessories Available HP 16190A Performance Test Kit HP 16191A Side Electrode SMD Fixture HP 16192A Parallel Electrode SMD Fixture HP 16193A Small Side Electrode SMD Fixture HP 16194A High Temperature Test Fixture HP 16453A Dielectric Material Test Fixture HP 16454A Magnetic Material Test Fixture
If you do not have the Adobe Acrobat Reader necessary for viewing this documentation, download your Free Acrobat Reader now. |
![]() |
Accessory product datasheets available online:
Page Updated: Thursday June 19 16:12:05 UTC 1997