HP 432 Thermistor Power Meter Family

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]
The HP 432 power meter family includes the HP 432A power meter and the HP 8477A power meter calibrator. The HP 432A power meter delivers high accuracy over a wide temperature range. By measuring the output voltage of the thermistor bridges and computing the corresponding power, even higher accuracy of 0.2 percent 0.5 microwave can be obtained. Accuracy can be maintained on even the most sensitive range because the error due to thermoelectric effect is reduced to a negligible level. Each thermistor mount is furnished with data stating the calibration factor and effective efficiency at various frequencies across the operating range. The HP 8477A power meter calibrator is specifically designed for use with the HP 432A power meter. It allows you to verify full-scale meter readings on all ranges, and meter tracking. Simply connect three cables between the power meter and calibrator; no charts or additional instruments are required.


This product family includes


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

General Characteristics

Power Range HP 432A: 7 ranges with full-scale readings of 10, 30, 100, and 300 µW; 1, 3, and 10 mW also calibrated in dBm from -20 dBm to +10 dBm full scale in 5 dB steps HP 8477A: 115 or 230 Vac + 10%, 50 to 400 Hz, 3 watts Noise: Less than 0.25% of full scale peak (typical) Response Time: At recorder output, 35 ms time constant (typical) Fine Zero: Automatic, operated by front panel switch Zero Carryover: Less than 0.50% of full scale when zeroed on most sensitive range Meter: Taut-band suspension, individually calibrated, mirror-backed scales. Milliwatt scale more than 108 mm (4.25 in) long. Calibration Factor Control: 13-position switch normalizes meter reading to account for thermistor sensor calibration factor. Range 100% to 88% in 1% steps Thermistor Sensor: Thermistor sensors are required for operation of the HP432A B. For microwave sensors HP 478B, 8478B, and 486 series, see chart below.

Outputs

Recorder Output: Proportional to indicated power with 1 volt corresponding to full scale. 1 k-ohm output impedance BCD Output: 8, 4, 2,1 code: "1" positive. TTL compatible logic. Operates with HP 5150A, Opt 002 (BCD) digital recorder. Power Consumption: 115 or 230 Vac +10%, 50 to 400 Hz, 1.5 watts Dimensions Size: 130 mm W x 155 mm H x 279 mm D (5.2 in x 6.1 in x 11.0 in) Net Weight: 2.3 kg (5.5 lb) Shipping Weight: 4.6 kg (10 lb)

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

No literature is available online at this time.

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP 432A     Thermistor Power Meter Family
  Opt 001   Rechargeable battery installed, provides up to 20 hours of
            continuous operation
  Opt 002   Input connector placed on rear panel in parallel with front
  Opt 003   Input connector on rear panel only
  Opt 009   3.1-m (10-ft) Cable for 100-ohm or 200-ohm Sensor
  Opt 010   6.1-m (20-ft) Cable for 100-ohm or 200-ohm Sensor
  Opt 011   15.2-m (50-ft) Cable for 100-ohm or 200-ohm Sensor
  Opt 012   30.5-m (100-ft) Cable for 100-ohm or 200-ohm Sensor
  Opt 013   61-m (200-ft) Cable for 100-ohm 200-ohm Sensor
  Opt 100   100 Vac Operation, 48 to 66 Hz
  Opt 910   Extra Operating and Service Manual (HP 00432-90009)
  Opt W30   Extended Repair Service
  Opt W32   Calibration Service
HP 8477A    Power Meter Calibrator
  Opt W30   Extended Repair Service
  Opt W32   Calibration Service


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

To help you make the right product choices



Privacy Statement    Use of this site indicates that you accept the Terms of Use
Contact HP Test & Measurement. (c) Copyright 1994 - 1998 Hewlett-Packard Company.

Page Updated: Thursday February 18 05:58:00 UTC 1999