HP 4395A Network/Spectrum/Impedance Analyzer, 10 Hz to 500 MHz![]()
|
When combined with a test set, the HP 4395A provides reflection measurements, such as return loss, and SWR, and S parameters. As a vector network analyzer, the HP 4395A operates from 10Hz to 500MHz with 1 mHz resolution; and its integrated synthesized source provides -50 to + 15dBm of output power with 0.1 dB resolution. The dynamic magnitude and phase accuracy are ±0.05dB and ±0.3deg. so that it can accurately measure gain and group delay flatness, which are becoming more important in modern electronics systems.
As a spectrum analyzer, the HP 4395A operates from 10Hz to 500MHz with resolution bandwidths (RBWs) spanning 1Hz to 1MHz in a 1-3-10 sequence. A fully synthesized local oscillator allows stable and accurate frequency analysis. As a impedance analyzer by the addition of Option 010 and test kit, the HP 4395A operates from 100kHz to 500MHz with Z basic accuracy: 3% (typical).
Network Analyzer Specification | Spectrum Analyzer Specification | ||
---|---|---|---|
Frequency | 10Hz to 500MHz* | Frequency Range | 10Hz to 500MHz |
Frequency Resolution | 1mHz | Noise Sidebands | <-105dBc/Hz typical at 10 kHz offset |
Output Power Range | -50 to 15dBm | Resolution Bandwidth | 1Hz to 1MHz in 1-3-10 steps |
Dynamic Range | >110dB@10Hz IFBW | Dynamic Range | >100dB third-order dynamic range |
Dynamic Accuracy | ±0.05dB/0.3deg. | Level Accuracy | <+-0.8dB at 50MHz |
Calibration | full two-port | Sensitivity | <-145dBm/Hz@freq.= 10MHz |
Standard Features: | Built-in IBASIC, HP-IB port, 3.5-inch floppy disk drive, direct color print, RAM disk, VGA Output | ||
Optional Features: | Time-Gated Spectrum Analysis (opt.1D6), High-Stability Frequency, DC Source (opt.001), Reference (opt.1D5), 50 ohm to 75 ohm Conversion Pad (opt.1D7) |
Impedance Measurement (Option 010)
Measurement Parameters | |Z|, |Y|, Phase, R, X, G, B, Cp, Cs, Lp, Ls, Rp, Rs, D, Q, |(gamma)| ,(gamma)x ,(gamma)y (gamma=reflection coefficient) |
Frequency Range | 100kHz to 500MHz |
Measurement Port | APC-7 on the HP43961A Test Kit |
Source Level at DUT | -56dBm to +9dBm |
DC Bias | ±40V (20mA max), BNC(f) Connector on HP43961A (A 2 kohm±5% internal resistor is used for dc bias current limitation. An option 001 dc source or an external dc bias source is required.) |
Calibration | OPEN (0S)/SHORT (0ohm)/LOAD (50ohm) calibration OPEN/SHORT/LOAD compensation on test fixtures port extension compensation |
Accuracy (Supplemental Performance Characteristic) | 3% basic accuracy at 23deg ±5deg C, after OPEN/SHORT/LOAD calibration |
HP 4395A Network/Specrum/Impedance Analyzer Opt 1D5 High Stability Frequency Reference Opt 1D6 Time-Gated Specrum Analysis Opt 1D7 50ohm to 75ohm Spectrum Input Impedance Conversion Opt 010 Impedance Measurement Function (Requires 43961A) Opt 001 DC Source (±40V, ±100mA)Accessories Available HP 43961A RF Impedance Test Kit (Add Test Fixtures listed below) HP 16191A Side Electrode SMD Fixture (DC to 2GHz) HP 16192A Parallel Electrode SMD Fixture (DC to 2GHz) HP 16193A Small Side Electrode SMD Fixture (DC to 2GHz) HP 16092A Spring Clip Fixture (DC to 500MHz)
Available accessories for this product include:
HP's family of Combination Analyzer HP4395A/HP4396B
( PDF)
Dramatic Speed Improvement for Narrow RBW Sweeps by Audio/Video/IF/RF/Spectrum Analyzers
( PDF)
HP 4395A Network/Spectrum/Impedance Analyzer
( PDF)
If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader now. |
![]() |
Related Information:
How to Measure Noise Accurately Using the HP Combination Analyzers
Copper Loop Measurement for ADSL
Related product families available online:
Page Updated: Thursday February 18 10:42:41 UTC 1999