HP 5315A Universal Counter |
The HP 5315A is a low-cost portable bench version of the HP 5316B high-performance universal counter. Unique counting circuitry gives the best of both reciprocal and conventional frequency counting. Full frequency range (to 100 MHz) on both input channels provides frequency, period, ratio A/B, and totalize A by B measurements.
An optional 1 GHz C channel extends the frequency range for navigations and communications applications. A high- stability oven oscillator can also be added for higher accuracy in demanding environments. The HP 5315A is suited for maintaining and calibrating all types of electronic equipment in the field. Its tough, lightweight polycarbonate case and optional battery pack make it convenient to use anywhere.
Input Characteristics (Channel A and Channel B): Range: dc coupled, 0 to 100 MHz ac coupled, 30 Hz to 100 MHz Sensitivity: 10 mV rms sine wave to 10 MHz. 25 mV rms sine wave to 100 MHz. 75 mV peak-to-peak at minimum pulse width of 5 ns. Sensitivity can be varied continuously up to 500 mV rms Nominal (Nom.) by adjusting sensitivity control. In sensitivity mode, trigger level is automatically set to 0V (Nom.). Dynamic Range: 30 mV to 5 V peak-to-peak, 0 to 10 MHz. 75 mV to 5V peak-to-peak, 10 to 100 MHz. Signal Operating Range: +2.5V dc to -2.5V dc. Coupling: ac or dc, switchable. Filter: Low pass, switchable in or out of Channel A, 3 dB point of 100 kHz (Nom.). Impedance: 1 Mohm (Nom.) shunted by less than 40 pF. Frequency (Channel A): Range: 0.1 Hz to 100 MHz. LSD Displayed: 10 Hz to 1 nHz depending upon gate time and input signal. At least 7 digits displayed in one second of gate time. Resolution: For FREQ < 10 MHz: +/- LSD +/- 1.4 x ((Trigger Error) / (Gate Time)) x FREQ. For FREQ >= 10 MHz: +/- LSD. Accuracy: +/- Resolution +/- Timebase Error. Ratio: Range: 0.1 Hz to 100 MHz, both channels. LSD: ((2.5 x Period A) / (Gate Time)) x Ratio, (rounded to nearest decade). Resolution: +/- LSD +/- ((B Trigger Error) / (Gate Time)) x Ratio. Accuracy: Same as resolution. Highest frequency input is connected to Channel A to achieve specified accuracy. Totalize: Manual Range: 0 to 100 MHz. A Gated by B: Totalizes input A between two events of B. Instrument must be reset to make new measurement. Gate opens on A slope, closes on B slope. Range: 0 to 100 MHz. Resolution: +/- 1 count. Accuracy: +/- 1 count +/- B Trigger Error x Frequency A. Period: Range: 10 ns to 10^5 s. LSD Displayed: 100 ns to 1 fs depending upon gate time and input signal. At least 7 digits displayed in one second of gate time. Resolution: For PER > 100 ns: +/- LSD +/- 1.4 x ((Trigger Error) / (Gate Time)) x PER. For PER >= 100 hs: +/- LSD. Accuracy: +/- Resolution +/- Timebase Error. Time Interval: Range: 0 ns to 10^5 s. LSD Displayed: 100 ns. Resolution: +/- LSD +/- Start Trigger Error +/- Start Trigger Error. Accuracy: +/- Resolution +/- Timebase Error. General: Check: Counts internal 10 MHz reference frequency over gate time range 500 us to 30 ms (Nom.). Error Light: LED warning light activated if logic error is found during instrument turn-on self-check. Display: 8-digit LED display, with engineering units annunciator. Overflow: Only frequency and totalize measurements will overflow. In case of overflow, eight least significant digits will be displayed and front panel overflow LED will be actuated. All other measurements which would theoretically cause a display of more than 8 digits will result in the display of the 8 most significant digits. Gate Time: Continuously variable, from 60 ms to 10s or 1 period (Nom.) except in time interval mode, where it is continuously variable from 4 readings/s to 1 reading every 10s via Gate Time control (Nom.). Sample Rate: Up to 7 readings/s (Nom.) except in time interval mode, where it is continuously variable from 4 readings/s to 1 reading every 10s via Gate Time control (Nom.). Operating Temperature: 0 degrees to 50 degrees C. Power Requirements: Selectable 100, 120, 220, 94 240V (+5%, -10%) 48-66 Hz; 15 VA maximum.
HP 5315A Portable Universal Counter Opt 001 High Stability TCXO Timebase Opt 002 Battery Pack Opt 003 1.0 GHz C Channel Opt 004 High Stability Oven Oscillator Opt W30 Extended Repair Service All orders must include one of the following no-charge power options: Opt 100 90 - 105 VAC Opt 120 108 - 126 VAC Opt 220 198 - 231 VAC Opt 240 216 - 252 VACHP 10856A Low Pass Filter Kit
Page Updated: Thursday June 19 16:21:18 UTC 1997