Product Summary
The HP 5334B counter offers high-resolution measurement power with
versatile signal conditioning, triggering, and external arming features
to handle a range of demanding measurement applications. Rise and fall
time, pulse width, and AC/DC voltage can be measured at the push of a
button or with a single HP-IB command. Automatic trigger level and
attenuator settings provide adjustment-free measurements that are
displayed in units of your choice.
Options include extended frequency to 1.3 GHz for
communication and navigation applications.
![[ Summary ]](../Graphics/tab-summary.gif)
![[ Features ]](../Graphics/tab-features.gif)
![[ Specifications ]](../Graphics/tab-specs.gif)
![[ Ordering ]](../Graphics/tab-ordering.gif)
Product Features
- 200 ps time interval average resolution, 2 ns single-shot resolution
- Selectable resolution and measurement time
- Full complement of measurements
- External arming increases measurement capability
- Two matched 100 MHz A and B input channels (optional 1.3 GHz C channel)
- Independent selection of signal conditioning for both A and B channels
- Proven reliability and excellent RFI protection
200 ps Resolution
The high resolution measurement power of the HP 5334B makes it suitable
for IC access time and propagation delay timing applications.
Selectable Resolution and Measurement Time
Digital gate time selection makes it easy to trade off resolution
and measurement time. Gate time is selectable in 1 ms
increments from 1 ms to 99.999 seconds. Resolution is at
least 9 digits per second from 0.001 Hz to 1300 MHz.
Measurements
Measurements include time-interval,
time-interval delay, and averaging.
External Arming
External arming/gating allows
synchronization with external events. This feature lets you
easily make pulsed RF communications and radar signal
measurements to 1.3 GHz.
![[ Summary ]](../Graphics/tab-summary.gif)
![[ Features ]](../Graphics/tab-features.gif)
![[ Specifications ]](../Graphics/tab-specs.gif)
![[ Ordering ]](../Graphics/tab-ordering.gif)
Product Specifications
CHANNEL A and CHANNEL B
Range:
dc coupled: 0 to 100 MHz.
ac coupled: 1 Mohm 30 Hz to 100 MHz, 50 ohm 1 MHz to 100 MHz.
Sensitivity:
15 mV rms sine wave to 20 MHz.
35 mV rms sine wave to 100 MHz.
100 mV peak-to-peak at a minimum pulse width of 5 ns.
Sensitivity can be continuously varied to 150 mV rms (NOMINAL)
using the TRIGGER LEVEL/SENS control in sensitivity mode. In
this mode, trigger levels are automatically set to 0V NOMINAL.
Trigger Level:
Range:
Manual (Auto Trigger OFF):
Continuously adjustable over ±5.1V, displayed by
Read Levels in 20 mV steps. In X10, ±51V displayed
in 200 mV steps.
Preset: 0V NOMINAL in Sensitivity Mode.
Auto Trigger: See Automatic Measurement Section.
Accuracy (Read Levels):
X1: ±30 mV ±1% of trigger level reading.
X10: ±300 mV ±1% of trigger level reading.
Damage Level:
Common Input: All specifications are the same as for separate
operation except for the following:
Sensitivity:
15 mV rms sine wave to 20 MHz.
75 mV rms sine wave to 100 MHz.
210 mV peak-to-peak at a minimum pulse width of 5 ns.
Dynamic Range (X1):
Impedance: 500 kohm NOMINAL shunted by <85 pF or
50 ohm NOMINAL.
(Resistance values are measured at dc and capacitance values at 10 MHz.)
EXTERNAL ARM
Front panel ARM Input can be used to determine Start and/or Stop point
of a measurement. External Arm can be used with all measurements except
Read Levels.
Minimum Start to Stop Time: 50 ns.
Maximum Transition Time: 1 µs.
Sensitivity: 500 mV peak-to-peak.
Signal Operating Range: -5 Vdc to +5 Vdc.
Dynamic Range: 500 mV to 10V peak-to-peak.
FREQUENCY A AND FREQUENCY B
Range: .001 Hz to 100 MHz.
LSD: ((4 ns) / (Gate Time)) × FREQ.
Resolution: ± LSD ± ((1.4 × Trigger Error + 1 ns rms) / (Gate Time)) × FREQ.
Accuracy: ± Resolution ± Timebase Error
PERIOD A
RANGE: 10 ns to 10e3 s.
LSD: ((4 ns) / (Gate Time)) × PER.
Resolution: ± LSD ± ((1.4 × Trigger Error + 1 ns rms) / (Gate Time)) × PER.
Accuracy: ± Resolution ± Timebase Error
TIME INTERVAL A TO B
Range: -1 ns to 10e3s (single-shot), 10s (100 GATE AVERAGE).
LSD: 1 ns (100 ps using 100 GATE AVERAGE).
Resolution: ± LSD ± Start Trigger Error ± Stop Trigger
Error ± 1 ns rms (100 ps rms using 100 GATE AVERAGE).
Accuracy: ±± Timebase Error ± Trigger Level
Timing Error ± Trigger Level Settling Error ± 2 ns (1pp ps rms using
100 GATE AVERAGE).
TIME INTERVAL DELAY
Used with Time Interval A to B, a selectable delay can be inserted between
START (Channel A trigger) and STOP (Channel B trigger). Electrical inputs
during delay are ignored. Specifications are the same as for Time Interval A to B.
Delay Range: 1 ms to 99.999 s (1 ms steps).
Delay Accuracy: ±100 µs ±0.05% × DELAY TIME.
AUTOMATIC MEASUREMENTS
These features are specified from 100 Hz to 20 MHz unless noted. Minimum
width at peak of signal: 5 ns. Auto Trigger and Auto Attenuation automatically
engaged for Rise/Fall Time, Pulse Width, and AC/DC Voltage measurements.
AUTO TRIGGER
RISE/FALL TIME A
PULSE WIDTH A
AC/DC VOLTAGE
Frequency Range: dc, 100 Hz to 20 MHz.
MATH
All measurements except for Totalize and Read Levels may be operated upon by
MATH functions. Math values are toggled on or off using the DISABLE key.
Offset and Normalize may be used independently or together.
![[ Summary ]](../Graphics/tab-summary.gif)
![[ Features ]](../Graphics/tab-features.gif)
![[ Specifications ]](../Graphics/tab-specs.gif)
![[ Ordering ]](../Graphics/tab-ordering.gif)
Product Ordering Information
HP 5334B Universal Counter
Opt 010 High Stability Oven Timebase
Opt 030 1.3 GHz C Channel (not retrofittable)
Opt 060 Rear Terminals (not retrofittable)
Opt 907 Front Handles Kit
Opt 908 Rack Mount Kit (for use with handles removed)
Opt 909 Front Handles and Rack Mount Kit
Opt W30 Extended Repair Service
HP 10855A Broadband Preamplifier
![[ Summary ]](../Graphics/tab-summary.gif)
![[ Features ]](../Graphics/tab-features.gif)
![[ Specifications ]](../Graphics/tab-specs.gif)
![[ Ordering ]](../Graphics/tab-ordering.gif)
To help you make the
right instrument choices
Your Field Sales Engineer is ready to answer your questions about
specifications, applications and ordering. Find your Field Sales Engineer at
your Local Sales Office.

Page Updated: Thursday June 19 16:21:15 UTC 1997