HP 5347A Microwave Counter/Power Meter![]()
|
With only five function keys, the HP 5347A is especially easy to use. Its rugged, lightweight package, along with an optional internal battery that provides up to two hours of cordless measurements, makes the HP 5347A ideal for taking portable measurements.
Counter Specifications
INPUT1 HP 5347A
Frequency Range: 500 MHz = 20.0 GHz Sensitivity: 500 MHz - 12.4 GHz -32 dBm (-35 dBm typical) 12.4 GHz - 20.0 GHz -27 dBm (-32 dBm typical) Maximum Input: +7 dBm Damage Level: +25 dBm, peak SWR: 500 MHz - 12.4 GHz <2:1 typical 12.4 GHz - 20.0 GHz <3:1 typical Accuracy: +/- 1 LSD +/- time base error x frequency Resolution: 1 Hz or 10 kHz, selectable
Optional Increased Damage Level Option 006
Protects Input 1 from damage by limiting high level signals. All specifications are the same except Input 1: Damamge Level 500 MHz to 6 GHz +39 dBm (8 watts) 6 GHz to 18 GHz +36 dBm (4 watts) 18 GHz to 20.0 GHz +34.8 dBm (3 watts) Sensitivity Sensitivity is reduced by: 3 dBm, 500 MHz to 12.4 GHz 4 dBm, 12.4 GHz to 20.0 GHz
INPUT2 HP 5347A
Frequency Range: 10 Hz - 525/MHz Sensitivity: 25 mV rms (15 mV rms typical) Maximum Input: 1 Mohm nominal shunted by <70 pF (10 Hz to 80 MHz) or 50 ohm nomimnal (10 MHz to 525 MHz) Damage Level: 50 ohm or 1 Mohm, dc - 5 kHz: 250V (dc + ac peak) >5 KHz: 5.5V rms (+28 dBm) + 1.25 x 10^6V rms/FREQ Accuracy: +/-1 LSD +/- [(1.4 x Trigger Error/Gate Time) +/- Time Base Error] x frequency Resolution: 1 Hz or 10 kHz, selectable
Automatic Amplitude Discrimination: Automatically measures the largest of all signals present, provided that signal is >6 dB (typical) above any signal within 500 MHz; >20 dB (typical) above any signal within 500 MHz to 20 GHz. Tracking Speed: Resolution = 1 Hz, Speed = 1 MHz/sec Resolution = 10 kHz, Speed = 1 GHz/sec Acquisition Time: Resolution = 1 Hz, Time = <125 ms Resolution = 10 kHz, Time = <60 ms Maximum Deviation: 20 MHz p-p, Automatic mode 60 MHz p-p, Manual mode (via HP-IB only) Maximum FM Rate: 10 MHz AM Tolerance: Any modulation index provided the minimum signal level is not less than the sensitivity specification. Power Meter Specifications Frequency Range: 10 MHz to 26.5 Ghz, sensor dependent Power Range: -70 to +20 dBm (100 pW to 100 mW), sensor dependent Dynamic Range: 50 dB in 10 dB steps Resolution: 0.01 dB in logarithmic mode, 0.1% of full scale in linear mode Auto Filter: The meter automatically selects the required number of averages for the selected range. Accuracy: Instrumentation: +/-0.02 dB or +/-0.5% Zero Set (digital settability of zero): +/-0.5% of full scale on most sensitive range. Decrease percentage by a factor of 10 for each higher range, +/-1 display count.
HP 5347A 20 GHz Counter/Power Meter Opt 002 Battery Pack Opt 006 Microwave Level Limiter Opt 011 HP-IB Interface Opt 070 Soft Carrying Case Opt 801 Include HP 8481A Power Sensor Opt 802 Include HP 8481D Power Sensor Opt 803 Include HP 8485A Power Sensor Opt 913 Rack Mount Kit Opt W30 Extended Repair Service Opt W32 Calibration Service Oven Timebase Call HP for details of this special option
HP 8481A Power Sensor HP 8481D Power Sensor HP 8485A Power Sensor HP 11730A 1.5-meter (5-ft) sensor cable (included) HP 11730B 3.0-meter (10-ft) sensor cable HP 11730C 6.1-meter (20-ft) sensor cable HP 11730D 15.2-meter (50-ft) sensor cable HP 11730E 30.5-meter (100-ft) sensor cable HP 11730F 61.0-meter (200-ft) sensor cable PN 9211-2649 Transit Case
Related product families available online: