HP 5347A Microwave Counter/Power Meter

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]
The HP 5347A counter/power meter offers the convenience of a single instrument that makes frequency and power measurements with accuracy and resolution comparable to stand-alone instruments. As a power meter, the HP 5347A offers excellent dynamic range, linearity, and accuracy. It uses the same proven power sensors as HP's standalone power meters. The high-performance frequency counter measures frequency from 10 Hz to 20 GHz.

With only five function keys, the HP 5347A is especially easy to use. Its rugged, lightweight package, along with an optional internal battery that provides up to two hours of cordless measurements, makes the HP 5347A ideal for taking portable measurements.


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

True power meter performance

Power measurements can be made from -70 dBM to +20 dBM, over a 10 MHz to 20.0 GHz frequency range, depending on the sensors used.

Negligible error

With power meter accuracy +/- 0.5% in linear mode, or +/-0.02 dB in logarithmic mode, the power meter causes virtually no increase in total measurement error.

Outstanding frequency measurements

The HP 5347A performance is comparable to HP's highest-performance CW counters, with selectable 1 Hz or 10 kHz resolution while measuring signals down to -35 dBm.

Built-in calibration

An average calibration table is permanently stored in memory. The press of a single key stores a frequency measurement for use in a power measurement. The stored frequency is then used to access the power sensor calibration factor in the calibration-factor-versus-frequency tables.


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Counter Specifications



INPUT1 HP 5347A
Frequency Range: 500 MHz = 20.0 GHz Sensitivity: 500 MHz - 12.4 GHz -32 dBm (-35 dBm typical) 12.4 GHz - 20.0 GHz -27 dBm (-32 dBm typical) Maximum Input: +7 dBm Damage Level: +25 dBm, peak SWR: 500 MHz - 12.4 GHz <2:1 typical 12.4 GHz - 20.0 GHz <3:1 typical Accuracy: +/- 1 LSD +/- time base error x frequency Resolution: 1 Hz or 10 kHz, selectable

Optional Increased Damage Level Option 006
Protects Input 1 from damage by limiting high level signals. All specifications are the same except Input 1: Damamge Level 500 MHz to 6 GHz +39 dBm (8 watts) 6 GHz to 18 GHz +36 dBm (4 watts) 18 GHz to 20.0 GHz +34.8 dBm (3 watts) Sensitivity Sensitivity is reduced by: 3 dBm, 500 MHz to 12.4 GHz 4 dBm, 12.4 GHz to 20.0 GHz

INPUT2 HP 5347A
Frequency Range: 10 Hz - 525/MHz Sensitivity: 25 mV rms (15 mV rms typical) Maximum Input: 1 Mohm nominal shunted by <70 pF (10 Hz to 80 MHz) or 50 ohm nomimnal (10 MHz to 525 MHz) Damage Level: 50 ohm or 1 Mohm, dc - 5 kHz: 250V (dc + ac peak) >5 KHz: 5.5V rms (+28 dBm) + 1.25 x 10^6V rms/FREQ Accuracy: +/-1 LSD +/- [(1.4 x Trigger Error/Gate Time) +/- Time Base Error] x frequency Resolution: 1 Hz or 10 kHz, selectable

Automatic Amplitude Discrimination: Automatically measures the largest of all signals present, provided that signal is >6 dB (typical) above any signal within 500 MHz; >20 dB (typical) above any signal within 500 MHz to 20 GHz. Tracking Speed: Resolution = 1 Hz, Speed = 1 MHz/sec Resolution = 10 kHz, Speed = 1 GHz/sec Acquisition Time: Resolution = 1 Hz, Time = <125 ms Resolution = 10 kHz, Time = <60 ms Maximum Deviation: 20 MHz p-p, Automatic mode 60 MHz p-p, Manual mode (via HP-IB only) Maximum FM Rate: 10 MHz AM Tolerance: Any modulation index provided the minimum signal level is not less than the sensitivity specification. Power Meter Specifications Frequency Range: 10 MHz to 26.5 Ghz, sensor dependent Power Range: -70 to +20 dBm (100 pW to 100 mW), sensor dependent Dynamic Range: 50 dB in 10 dB steps Resolution: 0.01 dB in logarithmic mode, 0.1% of full scale in linear mode Auto Filter: The meter automatically selects the required number of averages for the selected range. Accuracy: Instrumentation: +/-0.02 dB or +/-0.5% Zero Set (digital settability of zero): +/-0.5% of full scale on most sensitive range. Decrease percentage by a factor of 10 for each higher range, +/-1 display count.


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP 5347A	20 GHz Counter/Power Meter
   Opt 002	Battery Pack
   Opt 006	Microwave Level Limiter
   Opt 011	HP-IB Interface
   Opt 070	Soft Carrying Case
   Opt 801	Include HP 8481A Power Sensor
   Opt 802	Include HP 8481D Power Sensor
   Opt 803	Include HP 8485A Power Sensor
   Opt 913	Rack Mount Kit
   Opt W30	Extended Repair Service
   Opt W32	Calibration Service
   		Oven Timebase   Call HP for details of this special option

HP 8481A        Power Sensor
HP 8481D        Power Sensor
HP 8485A        Power Sensor
HP 11730A       1.5-meter (5-ft) sensor cable (included)
HP 11730B       3.0-meter (10-ft) sensor cable
HP 11730C       6.1-meter (20-ft) sensor cable
HP 11730D       15.2-meter (50-ft) sensor cable
HP 11730E       30.5-meter (100-ft) sensor cable
HP 11730F       61.0-meter (200-ft) sensor cable
PN 9211-2649    Transit Case
[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

To help you make the right instrument choices

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Related product families available online:



Contact HP Test & Measurement. (c) Copyright 1994, 1995, 1996, 1997 Hewlett-Packard Company.

Page Updated: Thursday November 13 07:50:35 UTC 1997