HP 5348A 26.5 GHz Microwave Counter/Power Meter
| |||
![[ Summary ]](/tmo/datasheets/Graphics/tab-summary-high.gif)
![[ Features ]](/tmo/datasheets/Graphics/tab-features-reg.gif)
![[ Specifications ]](/tmo/datasheets/Graphics/tab-specs-reg.gif)
![[ Key Literature ]](/tmo/datasheets/Graphics/tab-keylit-reg.gif)
![[ Ordering ]](/tmo/datasheets/Graphics/tab-ordering-reg.gif)
![[ Assistance ]](/tmo/datasheets/Graphics/tab-assistance-reg.gif)
The HP 5348A counter/power meter offers the convenience of a single instrument that makes frequency and power measurements with accuracy and resolution comparable to stand-alone instruments. As a true power meter, the HP 5348A offers excellent dynamic range, linearity, and accuracy. It uses the same proven power sensors as HP's standalone power meters. The high performance frequency counter measures frequency from 10 Hz to 26.5 GHz.
With only five function keys, the HP 5348A is especially easy to use. Its rugged, lightweight package, along with an optional internal battery that provides up to two hours of cordless measurements, makes the HP 5348A ideal for taking portable measurements.
Input 1
Frequency Range: 500 MHz to 26.5 GHz
Sensitivity
500 MHz - 12.4 GHz: -32 dBm (-35 dBm typical)
12.4 GHz - 20.0 GHz: -27 dBm (-32 dBm typical)
20.0 GHz - 26.5 GHz: -20 dBm (-27 dBm typical)
Maximum Input: +7 dBm
Damage Level: +25 dBm, peak
SWR
500 MHz - 12.4 GHz: <2:1 typical
12.4 GHz - 20.0 GHz: <3:1 typical
20.0 GHz - 26.5 GHz: <3:1 typical
Accuracy: ±1 LSD ±time base error x frequency
Resolution: 1 Hz or 10 kHz, selectable
Optional Increased Damage Level Option 006
Protects Input 1 from damage by limiting high level signals.
All specifications are the same except Input 1.
Damage Level
500 MHz to 6 GHz: +39 dBm (8 watts)
6 GHz to 18 GHz: +36 dBm (4 watts)
18 GHz to 26.5 GHz: +34.8 dBm (3 watts)
Sensitivity reduced by
3 dBm, 500 MHz to 12.4 GHz
4 dBm, 12.4 GHz to 20.0 GHz
5 dBm, 20.0 GHz to 26.5 GHz
Input 2
Frequency Range: 10 Hz - 525 MHz
Sensitivity: 25 mV rms (15 mV rms typical)
Impedance: 1 Mohm nominal shunted by <70 pF (10 Hz to 80 MHz)
or 50 ohm nomimnal (10 MHz to 525 MHz)
Damage Level: 50 ohm or 1 Mohm
dc - 5 kHz: 250V (dc + ac peak)
>5 KHz: 5.5V rms (+28 dBm) + 1.25 x 10^6V rms/FREQ
Accuracy: +/-1 LSD ±[(1.4 x Trigger Error/Gate Time)±Time
Base Error] x frequency
Resolution: 1 Hz or 10 kHz, selectable
Automatic Amplitude Discrimination
Automatically measures the largest of all signals present, provided
that signal is >6 dB (typical) above any signal within 500 MHz;
>20 dB (typical) above any signal within 500 MHz to 26.5 GHz.
Tracking Speed
Resolution = 1 Hz, Speed = 1 MHz/sec
Resolution = 10 kHz, Speed = 1 GHz/sec
Acquisition Time
Resolution = 1 Hz, Time = <125 ms
Resolution = 10 kHz, Time = <60 ms
Maximum Deviation
20 MHz p-p, Automatic mode
60 MHz p-p, Manual mode (via HP-IB only)
Maximum FM Rate: 10 MHz
AM Tolerance: Any modulation index provided the minimum signal level
is not less than the sensitivity specification.
Power Meter Specifications
Frequency Range: 0 MHz to 26.5 GHz, sensor dependent
Power Range: -70 to +20 dBm (100 pW to 100 mW), sensor dependent
Dynamic Range: 50 dB in 10 dB steps
Resolution: 0.01 dB in logarithmic mode, 0.1% of full scale in linear mode.
Auto Filter: The meter automatically selects the required number of
averages for the selected range.
Accuracy
Instrumentation: ±0.02 dB or ±5%;
Zero Set (digital settability of zero): ±5% of full scale
on most sensitive range. Decrease percentage by a factor of 10 for
each higher range, ±1 display count.
HP 5348A 26.5 GHz Counter/Power Meter Opt 002 Battery Pack Opt 006 Microwave Level Limiter Opt 011 HP-IB Interface Opt 070 Soft Carrying Case Opt 801 Include HP 8481A Power Sensor Opt 802 Include HP 8481D Power Sensor Opt 803 Include HP 8485A Power Sensor Opt 913 Rack Mount Kit Opt W30 Extended Repair Service Opt W32 Calibration Service
HP 11730A 1.5-meter (5-ft) sensor cable (included) HP 11730B 3.0-meter (10-ft) sensor cable HP 11730C 6.1-meter (20-ft) sensor cable HP 11730D 15.2-meter (50-ft) sensor cable HP 11730E 30.5-meter (100-ft) sensor cable HP 11730F 61.0-meter (200-ft) sensor cable
No literature is available online at this time.
To help you make the right product choices
Related Frequency Counters product datasheets:

Page Updated: Thursday February 18 09:03:10 UTC 1999