HP 5371A Frequency and Time Interval Analyzer![]()
|
The HP 5371A can help you analyze jitter effects, characterize the transient and steady state responses of a VCO, examine frequency or phase modulation, measure frequency stability, analyze computer peripherals, or measure fast hopping radio signals.
MEASUREMENT MODES
Function Range
Frequency A*, B* 125 mHz to 500 MHz Frequency A&B, A+B A-B, B-A, A/B, B/A 250 mHz to 500 MHz Period A*, B* 2 ns to 8.0 seconds Period A&B, A+B, A-B, B-A, A/B, B/A 2 ns to 4.0 seconds Totalize A, B, A&B, A+B 0 to 2^32-1 events, A-B, B-A, A/B, B/A each channel Time Interval A, B, A->B, B->A 10 ns to 8.0 seconds Continuous Time Interval A*, B* 100 ns to 8.0 seconds +/- Time Interval A->B, B->A, B -4.0 seconds to +4.0 seconds including 0 seconds Rise and Fall Time A** 1 ns to 100 us (auto-trigger) Positive and Negative Pulse Width A** 1 ns to 1 ms (auto-trigger) Duty Cycle A** 0% to 100% for pulse widths > 1 ns and periods < 1 ms (auto-trigger) Phase A rel B, B rel A > +/- 360 degrees Peak Amplitudes A, B 1 kHz to 200 MHz, 200 mV(pk-pk) to 2 V (pk-pk)
* Maximum sample rate for these measurements is 10 MHz (100 ns). For all others, maximum sample rate is 5 MHz (200 ns). ** Requres 8 ns setup time between each measurement.
FREQUENCY MEASUREMENT PRECISION For Single Measurement: Least Significant Digit Displayed: +/- (200 ps / Sample Interval) x Frequency Resolution: +/- (150 ps rms + (1.4 x Trigger Error) / Sample Interval) x Frequency Accuracy: +/- Resolution +/- (Timebase Aging x Frequency) For Continuous Frequency Measurements (Mean Estimation): rms Resolution: (square root of(13.5) x (150 ps rms + 1.4 x Trigger Error) / (No. of Blocks)^1/2 x (No. of Meas. per Block)^3/2 x Sample Interval) x Frequency Accuracy: +/- Resolution +/- (Timebase Aging x Frequency) Least Significant Digit Displayed: 200 ps / sqrt(N) Resolution: (+/- 150 ps rms +/- Start Trigger Error +/- Stop Trigger Error) / sqrt(N) Accuracy: +/- Resolution +/- (Timebase Uncertainty x Time Interval) +/- Trigger Level Timing Error +/- 1 ns Systematic Uncertainty INPUT TRIGGERING CHARACTERISTICS
Auto Trigger Manual Single or Triggering Repetitive
Voltage Range: x 1: -2 Vdc to +2 Vdc -2 Vdc to +2 Vdc x2.5: -5 Vdc to +5 Vdc -5 Vdc to +5 Vdc Frequency Range: dc to 500 MHz (HP 54001A, HP 54002A) 1 kHz to 200 MHz dc to 300 MHz (HP 54003A) Resolution x1: 2 mV NOMINAL 1% steps (2 mV minimum) x2.5: 5 mV NOMINAL 1% steps (5 mV minimum) Accuracy: 20 mV +/- 1% of setting +/- 20% of pk-pk amplitude (200 mV(pk-pk) minimum)
ARMING, GATING, AND TRIGGERING Event Holdoff and/or Gate: 0 to 4 x 10e9 events. Time Holdoff and/or Gate: 2 ns to 8.0 seconds, 2 ns resolution (settability). Actual gate time is displayed with 200 ps LSD. Edge Holdoff and/or Sampling: HP 5371A becomes armed after a delay from edge as follows: Ext Arm arms A < 15 ns B arms A, A arms B < 8 ns A arms A, B arms B < 5 ns INTERVAL Sampling: 600 ns to 8.0 seconds, 200 ns resolution (settability). Actual sample interval is displayed with 200 ps LSD. CYCLE Sampling: Cycles of input signal or 500 MHz timebase in discrete steps: 2^4, 2^8, 2^12, 2^16, 2^20, 2^24, 2^28.
HP 5371A Frequency and Time Interval Analyzer Opt 060 Rear Panel Inputs for channels A, B, and External Arm Opt W30 Extended Repair Service Opt 908 Rack Mount Kit for mounting without front handles Opt 913 Rack Mount Kit for mounting with front handles
HP 54001A 1 GHz Active Probe/Pod (10:1, 10 k ohm) HP 54002A 50 ohm Input Pod HP 54003A 1 M ohm Input Pod (with 10:1 scope probe) HP J06-59992A Time Interval Calibrator PN 1494-0059 Rack Slide Kit