HP 5372A Frequency and Time Interval Analyzer![]()
|
Unique to the HP 5372A (compared with the HP 5371A) are deeper memory, hardware histograms, and FFT capabilities for high-performance and more sophisticated jitter spectrum analysis and phase noise studies. The HP5372A lets you: analyze jitter effects, characterize the transient and steady state responses of a VCO, examine frequency or phase modulation, measure frequency stability, analyze computer peripherals, or measure fast hopping radio signals.
Single Frequency Measurement
Least Significant Digit Displayed: +/- ((200 ps) / (Sample Interval)) x Frequency Resolution: +/- ((150 ps rms + (1.4 x Trigger Error)) / (Sample Interval) x Frequency Accuracy: +/- Resolution +/- (Time Base Aging x Frequency)
Continuous Frequency Measurements (mean estimation)
rms Resolution (for number of measurements/block >3): ((square root of (13.5) x (150 ps rms + 1.4 x Trigger Error)) / ((# of Blocks)^1/2 x (# of Meas./Block)^3/2 x Sample Interval) x Freq. Accuracy: +/- Resolution +/- (Time Base Aging x Frequency)
Time Interval Measurements
Least Significant Digit Displayed: +/- ((200 ps) /( square root of (N))) N = number of measurements averaged Resolution: +/- ((150 ps rms +/- Start Trigger Error +/- Stop Trigger Error) / (square root of (N))) Accuracy: +/- Resolution +/- (Time Base Aging x Time Interval)± Trigger Level Timing Error +/- 1 ns Systematic Error (Systematic error can be reduced to <100 ps with the HP J06-59992A) Time Interval Calibrator)
Input: Channels A and B
The following refers to an HP 5372A with HP 54002A pods installed. Range: dc to 500 MHz. Sensitivity: (x1 attenuation, minimum hysteresis) 15 mV rms sine wave (45 mV pk-pk), typically 10 mV rms. 45 mV pk-pk for pulse input. Hysteresis control is available to reduce input sensitivity to trigger noise. Dynamic Range: x1:45 mV pk-pk to 2 V pk-pk. x2.5 attenuator: 115 mV pk-pk to 5 V pk-pk. Minimum Pulse Width: 1 ns for all measurement modes except holdoff arming. 1.5 ns with holdoff arming.
Input: Channel C
(option 030)Type N Connector Range: 100 MHz - 2 GHz (divide-by-4 prescaler). Sensitivity: (0 dB attenuation) 100 MHz to 1.5 GHz: -25 dBm. 1.5 GHz to 2 GHz: -20 dBm. Dynamic Range: 100 MHz to 1.5 GHz: -25 dBm to +7 dBm. 1.5 GHz to 2 GHz: -20 dBm to +7 dBm. Trigger level fixed at 0V NOMINAL. Impedance: ac coupled, 50 ohm, VSWR <2.5.
External Arm
Range: dc coupled to 100 MHz. Minimum Pulse Width: 5 ns Impedance: 1 Mohm NOMINAL, shunted by <50 pf. Triggering adjustable in 20 mV steps Range +/-5.00 V
Measurement Control
Holdoff or Sample: 0 to 4 x 1e9 events (65,000 events with fast meas. mode); 2 ns to 8.0 seconds (131 usec with fast meas. mode). Edge Holdoff or Sampling: HP 5372A becomes armed after a delay from edge as follows: Ext Arm arms A or B <15 ns B arms A, A arms B <8 ns A arms A, B arms B <5 ns Interval Sampling: 100 ns to 8 seconds (131 usec with fast meas. mode). Cycle Sampling: Cycles of input signal or 500 MHz time base in discrete steps: 2^4, 2^8, 2^12, 2^16, 2^20, 2^24, 2^28 (2^4, 2^8, 2^12 with fast mode). Random Sampling: Start of subsequent measurement delayed by a random number of events betwen 6 and 17 on channel A. Maximum input frequency 100 MHz. Inhibit Input: Rear panel input will inhibit memory acquisition when signal is above/below threshold (programmable from front panel, or HP-IB). Inhibit is independent of other arming and sampling. Pre-trigger: Measurements can be acquired before and after a pre-trigger event. These include an edge on the external arm channel for frequency, period, or time interval measurements or a detected time interval value for time interval measurements.
Rear Panel
Frequency Standard Output: 10 MHz. Short term stability not specified. Frequency Standard Input: 1, 2, 5, or 10 MHz input. Gate Outputs: Falling TTL edge indicates measurement sample. Delay Outputs: Falling TTL edge indicates completion of holdoff arming. Inhibit Input: Programmable input level suppresses measurement acquisition. TI Detect: Output is low for duration of out-of-range measurements. FastPort Outputs (option 020): Three 40-pin connectors provide unprocessed data directly from count hardware. 16 bits of data and 1 strobe for each connector.
HP 5372A Frequency and Time Interval Analyzer Opt 020 FastPort Data Output (not retrofittable) Opt 030 2 GHz Channel C (front panel input) Opt +36A Productivity Training (1/2 day) Opt 040 Jitter Spectrum Analysis Opt 060 Rear-Panel Inputs (50 ohm BNC) for Channels A and B; 1 Mohm BNC for external arm; deletes front panel inputs. Opt 090 Rear Panel Inputs, Channels A, B, and C; 1 Mohm BNC for external arm; 50 ohm BNC for channels A and B; deletes front panel inputs. Opt W30 Extended Repair Service Opt W32 Calibration ServiceHP 5364A 2 - 18 GHz Microwave Mixer/Detector HP 54001A 1 GHz Active Probe Pod (10:1, 10K ohm) HP 54002A 50 ohm Input Pod HP 54003A 1 M ohm Input Pod HP J06-59992A Time Interval Calibrator
Related HP 537X Series Family product datasheets: