HP 5373A Modulation Domain Pulse Analyzer


[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Summary

The HP 5373A provides built-in functions for pulsed signal analysis and is ideal for applications requiring radar and EW signal characterization. It minimizes design and testing efforts required for modern military, air traffic, marine, weather, and space-based radar systems. The measurements of phase, frequency, and time are optimized for complex signal modulations.

The HP 5373A simplifies the study of agile carriers, staggered PRI, chirp, phase coding, and similar modulations whether intentional or unintentional. The HP 5373A analyzer is particularly suited to working with pulsed and non- repetitive signals. It measures modulation and carrier frequency on pulsed RF signals to 500 MHz, and includes a detector channel for precise measurement of envelope parameters such as pulse width and PRI, peak envelope power, percent AM, jitter, and more.



[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Features

200 ps single-shot LSD resolution

There is usually no need to repeat a signal simply to accommodate your test equipment. If you can work with a repetitive signal, the HP 5373A provides signal averaging to improve resolution.

Broad measurement selection

Measure pulse width, duty cycle, rise and fall time; amplitude parameters; phase coded signals; tracking and capture range of a PLL; and much more.

Built-in analysis

Statistical measures of parameters such as pulse width and PRI jitter provide valuable insight on the performance of even the simplest of pulsed systems.

Graphic display

Results reveal maximum information at a glance. Markers provide read-out and analysis of your measurements. Numerical results can also be displayed.


[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Specifications

Measurements (see also HP 5372A Frequency Measurements)
	Frequency PRF:	Ch A and B: 125 mHz [8 kHz with fast meas. mode] to 500 MHz
					Ch C: 100 MHz to 2 GHz
	Period/PRI:	Ch A and B: 2 nsec to 8 sec [131 usec with fast meas. mode]
					Ch C: 0.5 nsec to 10 nsec
	Pulse Width/Off Time:	Ch A: 10 nsec to 8.0 sec
	Duty Cycle:	Ch A: 0% to 100% (provided pulse width is
					>1 nsec, and the signal period is:
						<1 msec (auto trigger)
						<2 sec [32.5 usec with fast meas. mode] 
							(manual trigger)
	Rise/Fall Time:	Ch A: 1 nsec to 100 usec transitions (auto trigger)

Input Characteristics (Channel A) 
The following refers to an HP 5373A with an HP 53702A envelope pod installed. 
All measurements are referenced to an input of 0 dBm at 275 MHz.
	Frequency Range:		30 MHz to 500 MHz
	Flatness:			+/- 1.0 dB (40 MHz to 500 MHz)
	AM Range:			20% to 85%
	AM Accuracy:			2.0%
	AM Frequency Range:		1 kHz to 9 MHz (typ)
	Minimum Pulse Width/Off Time:	<=100 nsec

Channel B Characteristics 
The following refers to an HP 5373A with an HP 54002A pod installed.
	Range:				dc coupled to 500 MHz
	Sensitivity:			-23.5 dBm (X 1 atten., minimum hysteresis)
					-27.0 dBm (typical)
	Hysteresis control is available to reduce input sensitivity to trigger noise.
	Dynamic Range:			X1: -23.5 dBm to +10 dBm
					X2.5: -14.8 dBm to +18 dBm
	Coupling:			dc
	Input Impedance (nominal):	50 ohms
	Max. Input Level:		+16 dBm (X1 attenuation)
					+24 dBm (X2.5 attenuation)

Channel C Characteristics
	Range:				100 MHz to 2000 MHz (divide
					by 4 prescale)
	Sensitivity (0 dB atten):	-25 dBm (100 MHz to 1.5 GHz)
					-20 dBm (1.5 GHz to 2 GHz)
	Dynamic Range:			-25 dBm to +7 dBm (100 MHz to 1.5 GHz)
					-20 dBm to +7 dBm (1.5 GHz to 2 GHz)
	Trigger Level:			Fixed at 0V NOMINAL
	Coupling:			ac
	Input Impedance (nominal):	50 ohms (VSWR <=2.5)
	Max. Input Level:		+20 dBm

External Arm
	Range:				dc coupled to 100 MHz
	Minimum Pulse Width:		5 nsec
	Triggering Range:		+/-5 V adjustable in 20 mV steps
	Input Impedance (nominal):	1M ohm, shunted by <50 pf
	Max. Input Level:		5 Vrms (+/-15 Vpk-pk, dc+/- peak ac)

Measurement Control
	Holdoff or Sample:		0 to 4e9 events (65,000 events with fast 
					meas. mode)
					2 nsec to 8.0 sec (131 usec with fast 
					meas. mode)
	Edge Holdoff or Sampling:	HP 5373A becomes armed after a delay from 
					edge as follows:
					     <15 nsec Ext Arm arms A or B
					     <8 nsec B arms A, A arms B
					     <5 nsec A arms A, B arms B
	Interval Sampling:		100 nsec to 8 sec (131 usec with fast
					meas. mode)
	Cycle Sampling: Cycles of input signal or 500 MHz time base in discrete
		steps: 2^4, 2^8, 2^12, 2^16, 2^20, 2^24, 2^28 (2^4, 2^8, 2^12 with 
		fast mode).
	Random Sampling: Start of subsequent measurement delayed by a
		random number of events between 6 and 17 on channel A. Maximum 
		input frequency 100 MHz.
	Inhibit Input: Rear panel input will inhibit memory acquisition
		when signal is above/below threshold (programmable from front
		panel, or HP-IB). Inhibit is independent of other arming and
		sampling.
	Pre-trigger: Measurement can be acquired before and after a
		pre-trigger event. These include an edge on the external arm channel
		for PRI/frequency, PRI/Period, or time interval measurements or a
		detected time interval value for time interval measurements.

Rear Panel
	Frequency Standard Output: 10 MHz, short term stability not specified
	Frequency Standard Input: 1, 2, 5, or 10 MHz input
	Gate Outputs: Falling TTL edge indicates measurement sample.
	Delay Outputs: Falling TTL edge indicates completion of holdoff arming.
	Inhibit Input: Programmable input level suppresses measurement acquisition.
	TI Detect: Output is low for duration of out-of-range measurements.
	FastPort Outputs (opt 020): Three 40-pin connectors provide
		unprocessed data directly from count hardware. 16 bits of data and 
		1 strobe for each connector.



[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Ordering Information

HP 5373A	Modulation Domain Pulse Analyzer
   Opt 020	FastPort Data Output (not retrofittable)
   Opt 030	2 GHz Channel C (front panel input)
   Opt +36A	Productivity Training (1/2 day)
   Opt W30	Extended Repair Service
   Opt W32	Calibration Service

HP 5364A 2 - 18 GHz Microwave Mixer/Detector HP 54001A 1 GHz Active Probe Pod (10:1, 10K ohm) HP 54002A 50 ohm Input Pod HP 54003A 1 Mohm Input Pod HP J06-59992A Time Interval Calibrator



[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

To help you make the right instrument choices



Contact HP Test & Measurement. (c) Copyright 1994, 1995, 1996, 1997 Hewlett-Packard Company.

Page Updated: Thursday June 19 16:21:12 UTC 1997