HP 5385A Frequency Counter |
The HP 5385A Frequency Counter utilizes interpolar-enhanced reciprocal counting to provide outstanding performance at a low cost. Frequency coverage allows measurements from 10 Hz to 1 GHz, with resolution of 9 digits per second. The HP 5385A incorporates a programmable alphanumeric display, accessible remotely via either HP-IB (standard) or HP-IL (optional). The liquid crystal display (LCD) reduces component count, is easy to read (indoors or out), and improves counter reliability and serviceability.
The half-rack width HP 5385A instrument is portable, rackable, and stackable for use in the field, at a workbench, or as part of an ATE system. Options include a four-hour-capacity battery pack, and a high-stability ovenized oscillator for higher accuracy.
Channel A Characteristics
Sensitivity: [MAIN LEVEL] off
15 mVrms sine wave 50 Hz to 100 MHz
25 mVrms sine wave 10 Hz to 50 Hz
45 mV pk-pk 5 ns minimum pulse width
Dynamic Range: 45 mV to 4 V pk-pk × attenuator setting.
Coupling: AC
Impedance:
X1: 1 Mohm NOMINAL | | <25 pF
X20: 500 kohm NOMINAL | | <25 pF
Attenuator: X1 or X20 NOMIONAL, X20 increases to X40 below 50 Hz
Low Pass Filter: 100 kHz NOMINAL 3 dB point
Range: 90 to 1000 MHz
Sensitivity:
10 mVrms (-27 dBm) 100-1000 MHz
15 mVrms (-33 dBm) 90-100 MHz
Dynamic Range: 10 mV to 7 Vrms (-27 to +30 dBm)
Coupling: AC
Impedance: 50 ohm NOMINAL
Attenuator Level:
Manual: variable from X1 to X18 (0 to25 dB) NOMINAL
Auto: AGC mode for improved noise suppression.
Damage Level:
AC > 1 MHz +30 dBm (7 Vrms)
AC <1 MHz 2 Vrms
DC +/- 5V
Aging Rate: <1 × 10e-7/mo.
Temperature:
<2 × 10e-6, 0-40 degrees C
(± 1 × 10e-6, 0-40 degrees C if referenced to 25 degrees C,
and set to the offset frequency.)
Line Voltage: <5 × 10e-8 for ±10% variation.
Range Channel B: 50 MHz - 225 MHz
LSD Displayed: 10 Hz to 1 nHz
LSD ((4 nsec) / (Gate Time)) × FREQ
Resolution: ±1 LSD ± ((1.4 × Trigger Error + 1 nsec rms) / (Gate Time)) × Freq
Accuracy: ± Resolution ± Time Base Error × Period
HP 5385A Frequency Counter 1.0 GHz Opt 004 High Stability Ovenized Timebase Opt 005 Battery Pack
HP 34110A Vinyl Carrying/Operating Case HP 82167A HP-IL Cable, 0.5 m PN 5061-1171 Side Handle Kit PN 5060-0173 Rack Mount Kit (single) PN 5060-0174 Rack Mount Kit (dual)
Page Updated: Thursday June 19 16:21:11 UTC 1997