HP 5386A 3 GHz Frequency Counter


[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Summary

Based on gallium arsenide (GaAs) technology, the HP 5386A Frequency Counter offers prescaled measurements to 3 GHz. The GaAs prescaler and attenuator circuitry are integrated on a hybrid substrate. Bipolar technology permits the integration of reciprocal counting and interpolation techniques on a single integrated circuit. The tightly- integrated design achieves overall high performance, excellent reliability, and low cost of ownership.

The HP 5386A incorporates a 12-character display, accessible remotely via HP-IB. The liquid crystal display (LCD) can be read in direct sunlight, and has large characters that can be seen from the end of a bench or mounted high in a test rack. The HP 5386A can be easily carried to remote locations for service and calibration applications.



[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Features

Measurements Displayed

Controlled via HP-IB, the display can be used for messages and operator prompts, or even blanked for security. Measurements are displayed in your choice of units. The number of digits displayed can be varied from the front panel or HP-IB.

Reduced Switching

Frequency measurements from 10 Hz to 3 GHz are made with two input ports, instead of three, for reduced switching.


[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Specifications

Channel B

	Range: 100 MHz to 3 GHz (3.5 GHz TYPICAL), prescaled

	Sensitivity:
		100 MHz to 3 GHz:  15 mVrms (-23.5 dBm)
		100 MHz to 3.5 GHz TYPICAL: 5 mVrms (-33.0 dBm)

	Maximum Input: 0.5 Vrms (+7 dBm)

	Coupling: ac

	Impedance:
		50 ohm NOMINAL
		VSWR <2.5:1 TYPICAL

	Note: Manual level not active for Channel B.

	Damage Level:

		ac >4V (+25 dBm)
		dc +/- 5V

Channel A

	Range: 10 Hz to 100 MHz

	Sensitivity [MAN LEVEL] off:
		15 mVrms sine wave 10 Hz to 100 MHz
		45 mV pk-pk 5 ns minimum pulse width

	Dynamic Range: 45 mV to 4V pk-pk X attenuator setting

	Impedance:
		X1:	1 Mohm NOMINAL || <25 pF
		X20:	500 kohm NOMINAL || <25 pF

	Attenuator: X1 or X20 NOMINAL, X20 increase to X40 below 50 Hz

	Trigger Level:
		[MAN LEVEL] ON: variable from -0.1V to +0.1V x attenuator
			setting about averagesignal value
		[MAN LEVEL] OFF: automatically set to average value of
			signal

	Damage Level:

		X1:	10 - 200 Hz 350V (dc + ac peak)
			0.2 - 420 kHz 170V (dc + ac peak)
			0.42 - 10 MHz (5 X 10^7 Vrms Hz) / FREQ
			>10 MHz 5 Vrms

		X20:	<1 MHz, same as X1
			>1 MHz, 50 Vrms

Frequency A and B

	Range Channel A: 10 Hz - 100 MHz

	Range Channel B: 100 MHz - 3 GHz, prescaled

	LSD Displayed: 10 Hz to 1 nHz

	LSD: ((4 ns) / (Gate Time)) X FREQ, rounded to nearest decade

	Resolution: +/- 1 LSD +/- ((1.4 X Trigger Error + 1 ns rms) / 
		(Gate Time)) X FREQ

	Accuracy: +/- Resolution +/- Time Base Error X Frequency

Period A

	Range: 10 ns to 0.1s

	LSD Displayed: .001 fs to 10 ns

	LSD: ((4 ns) / (Gate Time)) X period, rounded to nearest decade

	Resolution: +/- 1 LSD +/- ((1.4 X Trigger Error + 1 ns rms) / 
		(Gate Time)) X period

	Accuracy: +/- Resolution +/- Time Base Error X Period

Timebase (TCXO)

	Frequency: 10 MHz

	Aging Rate: <1 X 10e-7/month

	Temperature: <2 X 10e-6, 0-40 degrees C (+/-1 X 10e-6, 0-40 degrees C 
		if referenced to 25 degrees C, and set to the offset frequency)

	Line Voltage: <5 X 10e-8 for +/-10% variation

Option 004, Oven Timebase

	Frequency: 10 MHz
	
	Aging Rate: <3 X 10e-8/month, after 30 days of continuous operation

	Temperature: +/-1 X 10e-7, 0-50 degrees C referenced to 25 degrees C

	Line Voltage: <2 X 10e-0 for +-10% variation

General

	Gate Times: 0.1, 1, or 10s NOMINAL
		Accuracy: +/-15% + up to 1 period of input signal

	Timebase Output: 10 MHz 25 mV pk-pk NOMINAL into 50 ohm load

	External Timebase Input: 10 MHz, 0.5 Vrms into 500 ohm; 15V (dc + ac pk)
		maximum

	Operating Temperature: 0 degrees C to 50degrees C

	Power Requirements:

		AC Operation: Selectable, 30 VA maximum
			115V + 10%, -25%: 48-66 Hz
			230V + 10%, -15%: 48-66 Hz
			115V + 10%, -10%: 380-420 Hz
	Weight: Net, 3.4 kg (7 lb 8 oz) Shipping 5.3 kg (11 lb 9 oz)

	Dimensions: 212.3 mm W X 88.1 mm H X 421.6 mm D (8-1/3 X 3-1/2 X 16-1/2 in)



[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Ordering Information

HP 5386A	Frequency Counter 3.0 GHz
   Opt 004	High Stability Oven Timebase
   Opt 1A3	Bellcore CLEI code
   Opt 910	One additional manual
   Opt W30      3 Year Return Repair Service
   Opt W32      3 Yr Calibration Customer Return Svc.
   Opt W34      3 Year Standards Compliant Cal Service
   Opt W50      5 Year Return Repair Service
   Opt W52      5 Year Cust Return Calibration Service
   Opt W54      5 Year Standards Compliant Cal Service

PN 5061-9688    Front Handle Kit
PN 5061-9672    Rack Mount Kit (single)
PN 5062-3974	Rack Mount Kit (dual)
PN 5061-9694	Lock Link Kit for 5062-3974



[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

To help you make the right instrument choices



Contact HP Test & Measurement. (c) Copyright 1994, 1995, 1996, 1997 Hewlett-Packard Company.

Page Updated: Thursday June 19 16:16:14 UTC 1997