HP 54008A 22-ns Delay Line for the HP 54750A High-Bandwidth Digitizing Oscilloscope

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]
View the trigger event

The HP 54008A is a 22-ns delay line whose primary application is to view the trigger event on an HP 54750-series and HP 54120-series digitizing oscilloscopes. The delay line has a frequency range which extends from dc to 20 GHz and has less than 5.70 dB of attenuation at 20 GHz.

Each delay line is shipped from the factory with a VSWR versus frequency plot and an insertion loss versus frequency plot for that particular delay line.


The HP 54750 family of digitizing oscilloscopes are sequential sampling digitizing oscilloscopes and contain about 22 ns of delay between trigger point and the first sampling point. Using the HP 54008A, the trigger point can be displayed on the screen of any HP 54750-series oscilloscope.


When using an HP 54750-series oscilloscope with the HP 54118A trigger, the delay between the trigger event and the first sample point is increased by about 4 ns. Because the HP 54008A delay line's electrical timing length is greater than 22 ns long, the HP 54008A will work in this application also.


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Performance Specifications

Delay:			>= 22 ns  

Frequency Range:	dc to 20 GHz

VSWR:			<= 1.45:1  

Insertion Loss:		<= 5.70 dB at 20 GHz  

Performance Characteristics

Risetime:		<= 40 ps (20% to 70%)  

Overshoot:		<= 5 mV as displayed with the  
			HP 54754A TDR step generator  

Connectors:		SMA (f)

Environmental Characteristics


Temperature Operating:	+ 15 °C to + 35 °C  (+ 59 °F to + 95 °F)  

Temperature: 		-40 °C to + 70 °C  (-40 °F to + 158 °F)  
NonOperating:

Humidity Operating:	Up to 90 % relative humidity 
			at +35 °C (+ 95 °F)  

Humidity Nonoperating:	Up to 95 % relative humidity 
			at +65 °C (+ 149 °F)  

Altitude Operating:	4,600 m (15,000 feet)  

Altitude Nonoperating:	15,300 m (50,000 feet) 

Weight:			Net Approximately 1.53 kg (3.4 lb)  

Dimensions:  		Width: 212.60 millimeters (8.37 inches)    
			Heighth: 68.58 millimeters (2.7 inches)   
			Depth: 222.25 millimeters (8.75 inches) 


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP 54008A      22-ns Delay Line


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

No literature is available online at this time.

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

To help you make the right product choices

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Related product families available online:



Privacy Statement    Use of this site indicates that you accept the Terms of Use
Contact HP Test & Measurement. (c) Copyright 1994 - 1998 Hewlett-Packard Company.

Page Updated: Thursday February 18 09:25:35 UTC 1999