HP 54520A 2 Channel, 1 GSa/s Monochrome Digitizing Oscilloscope

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]
The Performance You Need... The Features You Want.

These state of the art, 500 MHz realtime oscilloscopes have the extra performance and features that you need to get an accurate view of the high frequency signals you encounter, whether you do analog or digital design, R&D or manufacturing test. With 32 K memory per channel, 2 channels, sample rate up to 1 GSa/s, 1 ns peak detect, and sophisticated triggering, you can be confident that you won't miss critical signal activity. These feature-rich digitizing scopes also include standard FFT, floppy disk drive, and sequential single shot acquisition for better analysis and characterization of your signals. Choose between a color flat-panel display (54520C) or monochrome display (54520A).


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

General Specifications

Maximum Sample Rate
	HP 54520A   1 GSa/s (1 CH on),  500 MSa/s (2 CH on)

Number of Channels  (all are simultaneous acquisition)
	HP 54520A:  2

Record Length
        32,768 pts (real time) maximum           
	501 pts (repetitive)

Resolution
        8 bits, 10 bits via HP-IB with averaging

Vertical Specifications

   
Repetitive Bandwidth
	500 MHz   (equivalent time) (rise time <= 700 ps)

Real Time Bandwidth 
	HP 54520A  125 MHz (2 CH) 250 MHz (1CH)

Sensitivity
        1 mV/div to 5 V/div

dc Gain Accuracy                 
	± 1.25% of full scale

Input Impedance                  
	R: 1 Mohm  ± 1% or 50 ohm  ± l%
        C: 7 pF nominal

Input Coupling
        ac, dc

Maximum Input
        1 Mohm ± 250 V (dc + ac)      [ac< l0 kHz]
        50 ohms: 5 V rms

Offset Range       
 	Vertical Sensitivity     Available Offset
        1 mV to 50 mV/div           ±2 V
  >50mV to 250 mV/div               ±l0 V
        >250 mV to 1.25 V/div       ±50 V
        >1.25 V/div to 5 V/div      ±250 V

Offset Accuracy                 
	±(l.25% of channel offset + 2% of full scale)

Voltage Measurement Accuracy
Dual Cursor    	±[(1.25%)(full scale)+(0.032)(V/div)]
Single Cursor   ±[(1.25%)(full scale)+(offset accuracy)+
		(0.016)(V/div)]

Horizontal Specifications

Time Base Range                  
		            500 ps/div to 5 s/div

Resolution                        
		            10 ps

Delta-t Accuracy
Repetitive: (> 8 averages)  [(0.005%)(delta-t)+(100 ps + 
			    0.1 % of full scale)]
Real Time 	            [(0.005%)(delta-t)+(0.2)(sample period)]
Peak Detect                 [(0.005%)(delta-t)+(l sample period)]

Time Tag 
Resolution  		    100 ps
Accuracy                    ±[0.005%(reading)+l00ps]

Delay Range (posttrigger)   10E07 x sample period
Delay Range (pretrigger)    32K x (sampleperiod)

Trigger Specifications

Sensitivity           dc to 100 MHz             100 MHz to 500 MHz
Internal              0.5 div                   1.0 div
External (54520A)     0.0225 x (signal range)   0.045 x (signal range)
Auxiliary             dc to 50 MHz: 250 mVp-p

Pulse Width (minimum) 1 ns

Level Range
Internal              ±1.5 X full scale from center screen
External (54520A)     ±25V
Auxiliary	      ±5V
Modes                 Edge, pattern, glitch, time qualified pattern, 
		      line, state, event-delayed, time-delayed, TV 
		      (NTSC, PAL, and user-definable formats)

Trigger Coupling      dc, ac, low frequency reject

Display Characteristics

Display Mode    "A"  models feature monochrome CRT (Cathode Ray 
		Tube) displays.  "C" models feature flat-panel color 
		TFT (Thin Film Transistor) liquid-crystal displays.  

Modes      	Averaging from 2 to 2048, envelope, infinite&
		variable persistence from 500 ms to 10 s, connect 
		the dots, peak detect

Graticules    	full grid, axes, frame, or no graticule

Measurements


Automatic  	23 measurements on front panel or over HP-IB:
                Rise time       V p-p          V dc rms
                Fall time       V min          Preshoot
		Frequency       V max          Overshoot
                Period          V avg          Voltage at time
		- Width		V base	       Time at min voltage
     	 	+ Width         V amptd        Time at max voltage
		Duty cycle      V top          Time at voltage
		Delta time      V ac rms

User-Definable
		Both upper and lower thresholds can be set from
                -25% to 125% for all automatic measurements

Modes         	
		Continuous, statistics, limit test, or waveform 
		compare

Math/Analysis Functions Operators	
		add, subtract, multiply, versus, integrate, 
		differentiate, invert magnify, and FFT

Additional Characteristics

Peak Detect     Captures and displays glitches or other high-speed 
		events as narrow as 1 ns in real-time mode at sample 
		rates of 250 MSa/s or less with sequential 
		single-shot turned off.
             
Sequential      Captures successive single-shot events without 
Single Shot	capturing the dead time in between, and stores up 
		to 400 Kbytes of waveform data.

Sequential Single-shot Throughput
                  Record          Waveforms Stored    Max Number of
                  Length            Per Second       Waveforms Stored
                  50              	1333             4395
                  500            	1111             739
                  8000            	294              49
                  32,000         	90               12

Waveform Store  
		4 nonvolatile, 2 pixel (volatile), and segmentable
		memory for storing measurement failure waveforms or
		sequential single-shot waveforms.

Screen Update Rate (typical at 500 ns/div)           
		record length (points)
Real Time     	  500     	8K    	16K  	32K
updates/s.        150      	110     84   	58

Repetitive        normal	8 avgs  128 avgs
updates/s:        150      	91      91


Optional Capabilities


Optional Telecommunication-Mask Application

Make telecom mask-template measurements to ANSI, CCITT, and ISDN 
standards using the HP 54520 or 54540 series oscilloscopes equipped 
with Option 001.  Option 001 features 21 standard masks stored on a 
flexible disk.  You can automatically trigger on positive isolated 
ones in live traffic for many standard telecom signals.  The scope 
will automatically best-fit the test signal to many masks and give 
automatic pass-fail comparisons of the mask to its corresponding 
input signal.

For more information on Option 001, Telecommunications Masks 
Applications ask your HP Sales representative for HP Product Overview 
5963-1859E (international version) or 5963-1859 EUS  (US  version).

Optional Active Probing

The HP 1144A active probe features 800-MHz bandwidth, 2-pF input 
capacitance, and 1-Mohm input resistance.  The HP 1145A two-channel 
active probe features 750-MHz bandwidth, 2-pF input capacitace, and 
1 Mohm input resistance and is designed for easy connection to 
surface mount devices.


I/O Characteristics

HP-IB           Fully programmable, complies with IEEE 488.2
RS-232          Bidirectional serial communication
Centronics      Talk-only parallel communication for HP PCL 
		compatible printers
                           	

Hardcopy Supported Printers	
		Color "C" Models        Monochrome "A" Models
		HP 540 Centronics       Thinkjet HP-IB, RS232
		HP 560C Centronics      HP 540 Centronics
                                        HP 560C, Centronics
                                        B/W only
Supported Plotters:
		HP Color Pro HP-IB      HP   7440A   HP-IB
		HP   7475A   HP-IB   	HP   7475A   HP-IB
		HP   7470A   HP-IB 	HP   7470A   HP-IB
		HP   7550A   HP-IB   	HP   7550A   HP-IB

Disk Drive            
          	DOS-compatible, 3-1/2" disk drive to store and recall 
		setups, waveforms, and screen images (TIFF, PCX, and 
		EPS).  The disk drive is also utilized to download 
		the oscilloscope's operating-system software and can 
		be used to upgrade the scope's software as future 
		upgrades become available.  A free software-upgrade 
		notification service is offered to registered users.

Front Panel Setups
                9 (nonvolatile memories)


FFT Characteristics

Frequency Range dc to: 	
		500MHz (l ch) 	
                250MHz (2 ch)  
                                                                     		
Frequency Resolution:
Minimum (max samp rate)/(512 pts): 	
		977 kHz (2ch)	
		1.95 MHz (1 ch on)
 
Maximum (min samp rate)/(32,768 pts):
   		0.305 MHz            

Horizontal Magnify               
		Specify the frequency that is displayed at center 
		screen, and magnify the frequency-domain display 
		about that point.

Frequency  Accuracy                
		[1/2 (sample frequency) (1/32768)]+ 
		(5 x 10E-5)(signal frequency)

Amplitude Display
		Power in dBm

Windowing          
	        Hanning, Flattop, Rectangular

Environmental Characteristics


Power           Voltage 115/230 Vac, -25% to +15%, 48 to 440Hz,
		350 VA maximum
                           
Weight       	Net: approximately 11.8 kg (261b)
          	Shipping: Approximately 21.3 kg (47 lb)

Dimensions   	Height: 218 mm (8.6 in)
                Width: 440 mm (17.3 in)
                Depth: 367 mm (14.5 in)


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP 54520A	Digitizing Oscilloscope
   Opt 001	Telecom Mask Test Software (Downloaded)
   Opt 002	HP 1145A 2-ch, 750 MHz, SMT Active Probe
   Opt 003	HP 1144A 800 MHz Active Probe
   Opt 090	Delete probes (HP 54520A)
   Opt 908	Rackmount kit (5062-7379)
   Opt 910	Additional manuals (User's, Programming, Service)

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

All orders include the following:

Available accessories for this product include:


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

To help you make the right instrument choices

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Related Information:


Related HP 54520- Series and HP 54540- Series Digitizing Oscilloscopes product datasheets:

Related product families available online:



Contact HP Test & Measurement. (c) Copyright 1994, 1995, 1996, 1997 Hewlett-Packard Company.

Page Updated: Wednesday January 14 23:59:04 UTC 1998