HP 54616B 2 GSa/s 500 MHz 2 Channel Oscilloscope![]()
|
The Feel of Analog: When you're troubleshooting, you want to stay focused on two things: your circuit and the scope's display. You don't want to waste time pressing buttons or waiting for the scope to update. That's why the HP 54616B has front-panel controls that look and work like the controls on your old analog scope.
The Power of Digital: Digitized waveforms open up entirely new possibilities, such as pretriggering, autoscale to automatically set voltage, time and trigger parameters, the ease and accuracy of automatic measurements, saving and recall of traces, hardcopy output to printers, and remote control with your computer.
Primary Vertical Specifications
Channels 2 Bandwidth (-3dB) dc to 500 MHz ac coupled 0 Hz to 500 MHz Sensitivity 2 mV/div to 5 V/div Accuracy ±2% Vernier Accuracy (2) ±2% Rise Time 700 ps (calculated) Coupling dc, ac, & ground Input R&C 1 Mohm, approx. 9 pF or 50 ohm selectable
Secondary Vertical Specifications
Bandwidth Limit approximately 30 MHz Inversion CH 1 & CH 2 CMRR 20 dB at 50 MHz Dynamic Range ±12 div from center screen Maximum Input 250 V (dc + peak ac) or 5 Vrms in 50 ohm mode 50-ohm Protection Protects 50 ohm load from excessive voltage Probe Sense Automatic readout of 1X, 10X, 20X, and 100X probes
Voltage Measurement Accuracy Specifications
Single Cursor Vertical Accuracy ±1.2% of full scale ± 0.5% of position value Dual Cursor Vertical Accuracy ±0.4% of full scale Math Functions CH 1 + or - CH 2
Horizontal System Specifications
Main & Delayed Main Sweep Range 5 s/div to 1 ns/div Delayed Sweep Range Up to 200X main sweep, as fast as 1 ns/div Accuracy ±0.005% Resolution 20 ps Delay Jitter 1 ppm Pretrigger Delay The greater of 30 µs or (Negative time) 60 div, not to exceed 100 s Posttrigger Delay The greater of 10 ms or (Trigger to start of sweep) 20,000 div, not to exceed 100 s Time Skew Adjustable over a range of ±25 ns to remove effects of cabling and probe delays Roll Mode At sweep speeds of 200 ms/div and slower, waveform data moves across the display from right to left with no dead time. Display can be free running (nontrig- gered) or triggered to stop on a trigger event.
Time Measurement Accuracy
Cursor accuracy ± 0.005% of reading, (delta-t & 1/delta-t) ± 0.2% of full scale, ± 100 ps
Internal Triggering Specifications
Sensitivity (Ch 1 & 2) dc to 100 MHz: 0.5 div or 5 mV 100 MHz to 500 MHz: 1 div or 10 mV Sources Channels 1 & 2, line and external Coupling ac, dc, HF reject, LF reject, & noise reject (LF & HF reject -3db at approx. 50 kHz) Modes Auto, Autolevel, Normal, Single, & TV
External Triggering Specifications
Range ± 2 V Sensitivity dc to 100 MHz: < 75 mV 100 MHz to 500 MHz: < 150 mV Coupling dc, ac Input R&C 1 Mohm, ~12 pf or 50 ohm selectable Maximum Input 250 V (dc + peak ac) or 5 Vrms in 50-ohm mode 50-ohm Protection Protects 50-ohm load from excessive voltage Probe Sense Automatic readout of 1X, 10X, 20X, and 100X probes
TV Triggering Specifications
TV line and field. 0.5 div of composite sync for stable display (Ch 1 & Ch 2) Line Counting Delay time calibrated in NTSC and PAL line numbers All Field Trigger Oscilloscope triggers on the vertical sync pulse in both fields, allowing use with noninterlaced video.
Trigger Holdoff
Adjustable from 300 ns to approximately 13 s
X-Y Operation
Bandwidth X & Y same as vertical system Phase Difference ± 3° at 10 MHz
Display System
Display 7-inch Raster CRT Resolution 255 vertical by 500 horizontal points Controls Front-panel intensity control Graticule 8 X 10 grid or frame Autostore Saves previous sweeps in half bright display and the most recent sweep in full bright.
Acquisition System Specifications
Max Sample Rate 2 GSa/s Resolution 8 bits Simultaneous Channels 2 Record Length 5,000 Usable Single-Shot Bandwidth 500 MHz Peak Detect 1-ns glitch capture Average Number of averages selectable at 8, 64, 256
Advanced Functions
Automatic Measurements Measurements are continously updated Voltage Vavg, Vrms, Vpp, Vtop, Vbase, Vmin, & Vmax Time Frequency, Period, +Width, -Width, Duty Cycle, Rise Time, & Fall Time Cursors Manually or automatically placed Setup Functions Autoscale Sets the vertical and horizontal deflection and the trigger level. Requires a signal with >0.5% duty cycle, >49 Hz frequency, >20 mV p-p Save/Recall 16 front-panel setups Trace Memory Two volatile pixel memories
Environmental Specifications
Line Voltage Range 100 Vac to 240 Vac Line Voltage Selection Automatic Line Frequency 45 Hz to 440 Hz Max Power Consumption 300 VA Temperature Operating: -10 °C to +55 °C Nonoperating: -51 °C to +71 °C Humidity Operating:95% RH at 40°C for 24 Hrs Nonoperating:90% RH at 65°C for 24 Hrs Altitude Operating: to 4,500 m (15,000 ft) Nonoperating: to 15,000 m (50,000 ft) Size (excluding handle) Heighth: 172 mm (6.8 in) Width: 322 mm (12.7 in) Depth: 317 mm (12.5 in) Weight 6.6 kg (14.5 lbs) Safety CSA Certification, IEC 1010 Warranty 3 years with option W50
HP 54616B Two-channel 2 GSa/s Oscilloscope Opt 001 RS-03 Magnetic Interface Shielding Added to CRT Opt 002 RE-02 Display Shield added to CRT to reduce radiated interface Opt 0B0 Delete Manual Opt W50 Additional 2 year warranty Opt 1BP Mil Std 45662A cal with test data Opt 005 Enhanced TV/Video Measurements and Triggering Opt 090 Delete probes Opt 101 HP 10098A Accessory Pouch & Front Panel Cover^M Opt 103 HP 54654A Operator's Training Kit Opt 104 5041-9409 Carrying Case Opt 1CM 5062-7345 Rack Mount Kit Opt 106 HP 34810A BenchLink/Scope (V1.4 or Greater) for WindowsAll orders include the following:®
two HP 10073A 10:1 passive probes two user's guides
Trademarks:
Windows is a U.S. registered trademark of Microsoft Corporation.
Available accessories for this product include:
No literature is available online at this time.
Additional help is a click away!
Want to talk to a knowledgeable engineer about this product at our HP Test and Measurement Call Center? Do you have a question you want to send in by e-mail? A measurement set-up you need to fax in? Follow the link above and you will be given a choice to contact an engineer by telephone, e-mail or fax.
This service is available only in the United States. For assistance outside the U.S., contact your local HP DIRECT or HP Sales Office.
Related Information:
HP 54602B (2+2) Channel 150 MHz Oscilloscope For a high quality lab scope when your needs go past 100 MHz, take a closer look at the HP 54602B.
HP 54603B 2 Channel 60 MHz Oscilloscope For colleges and universities, this scope is a great way to introduct students to the world of professional test equipment.
HP 54610B 2 Channel 500 MHz Oscilloscope With its horizontal accuracy of 0.01% and 1 ns/div time base, you know you'll catch the critical details.
HP 54615B 1 GSa/s 500 MHz 2 Channel Oscilloscope The HP 54615B oscilloscope delivers the comfortable feel of an analog oscilloscope with the power of a digital architecture.
HP 54616C 2 GSa/s 500 MHz 2 Channel Color Oscilloscope This oscilloscope offers an incredibly high level of digitizing performance to give you confidence in your critical measurements at a fraction of the price that you might expect.
HP 54645A 2-Channel 100 MHz Deep Memory and MegaZoom Oscilloscope With the addition of the new HP 54645A Oscilloscope to your lab, you will be able to view easily the full range of signals in your circuit.
HP 54645D Dual Channel 100 MHz 200 MSa/s Mixed Signal Oscilloscope This MSO seamlessly integrates logic channels into an oscilloscope to offer new measurement capability for designers of combined analog and digital systems.
HP Infinium Oscilloscopes Web Site Main Page HP Infinium oscilloscope main page. Hewlett Packard test and measurement.
Related product families available online:
Page Updated: Thursday February 18 10:56:23 UTC 1999