HP 54711A 50-ohm Attentuator Plug-in for the HP 54700-series Real-time Oscilloscopes

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]
The HP 54711A provides 2-GHz bandwidth and is intended for repetitive applications. The HP 54720D with this plug-in samples the waveform at 500 MSa/s providing very fast acquisition and display.


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Time Base Specifications

Maximum Sampling Rate

  Real Time			2 GSa/s
  Equivalent Time		500 MSa/s

Record Length Per Plug-in	16 to 65,536 points

Max Record Duration 
at Max Sample Rate		32.768 us

Channel Specifications

Number of  Channels		1

Number of Mainframe Slots	1

Analog Bandwidth (-3dB)		> 2.0 GHz

Max Sample Rate to Analog 
Bandwidth			1:1

Hardware Bandwidth Limit 
(typical)			N/A

Digital Bandwidth Limit 
(typical)			dc to (sample rate used/20) in the 
				real-time mode only

Rise Time (typical)		190 ps

Channel Scale 
(8 Divisions Fullscreen)	2 mV/div to 1 V/div

Vertical Resolution and Accuracy Specifications

Full Resolution Channel Scales	20, 40, 50, 100, 200, 400, 
				500, 1000 mV/div

Vertical Resolution					
  Real Time			8 bits (0.4%), increasing to 8.6 bits
				(0.1%) with digital BW limit on
  Equivalent Time		8 bits (0.4%).  This is improved by 
				averaging to > 12 bits

dc Gain Accuracy (typical)	
  Normal Accuracy		+/- 3% of full screen at full 
				resolution channel scale
  Best  Accuracy Cal		+/- 1% of full screen at full 
				resolution channel scale

Offset Accuracy
  Normal Accuracy		+/-(3% of offset setting + 3% of full
				screen at full resolution ch. scale)
  Best Accuracy Cal		+/-(.5% of offset setting + 1% of full
				screen at full resolution ch. scale)

dc Voltage Measurement Accuracy
  Single Marker			+/-[(gain accuracy + offset accuracy) 
				+ (resolution)]
  Dual Marker			+/-[(gain accuracy) + (2)(resolution)]

dc Offset Range			+/-11 times the full resolution 
				channel scale

Miscellaneous Vertical Specifications

RMS Noise (Max Sensitivity)	<300 uV

Input Connector			N-type (f)

Input Resistance		50 ohms +/- 1.5%

Input Coupling 
(Cutoff Frequency)		dc

Maximum Safe Input Voltage 
(typical)			>20 mV/div: 5 Vrms (+27 dBm)
				< or = mV/div: 0.5 Vrms (+7 dBm)

Probe Power			Yes

Triggering Specifications

Type				External Trigger

Input Resistance		50 ohms

Input Connector			N-type (f)

Trigger Sensitivity			

  High Sensitivity (typical)	dc to 100 MHz: 40 mVpp; increasing 
				linearly to 200 mVpp at 2.5 GHz
  Normal Sensitivity		dc to 100 MHz: 40 mVpp; increasing 
				linearly to 200 mVpp at 2.0 GHz
  Noise Reject			N/A

Min Pulse Width at 		< 200 ps at > 200 mVpp
Normal Sensitivity			

RMS Jitter (typical)		< 6 ps +/-0.0001% of delay setting

Trigger Level Range		+/- 1 V

Max Safe Input Voltage 		+/- 2 V[dc + peak ac] or +16 dBm
(typical)				

Probe Power			Yes


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP 54720D	dual-channel, 8-GSa/s real-time modular oscilloscope mainframe

HP 54711A	50-ohm Attentuator Plug-in for the HP 54700-series Real-time Oscilloscopes

   Opt 002	Training kit including a PC board, a demo guide,
		and a power supply (HP P/N 54720-68701 + 9100-4503 USA only)

   Opt 907	Rackmount kit (handles only) (HP P/N 54710-68703)
		includes long screws

   Opt 908	Rackmount kit (ears only) (HP P/N 54710-68704)
		includes special long screws

   Opt 909	Rackmount kit with handles (HP P/N 54710-68705)
		includes long screws

   Opt 0B1	One additional User s and Programmer s Book Set (English only)

Plug-Ins
HP 54711A	50-W attenuator plug-in with external trigger
HP 54712A	50-W amplifier plug-in with internal trigger
HP 54713B	50-W/1-MW, amplifier plug-in with internal trigger
HP 54714A	2-channel plug-in
HP 54715A	Single channel differential probe
HP 54717A	cal plug-in (for yearly mainframe calibration)
HP 54718A	trigger plug-in
HP 54721A	2-slot amplifierplug-in (4 GSa/s sample rate)
HP 54722A	4-slot attenuator plug-in (8 GSa/s sample rate)
HP 54700A	filler panel
Options available for each plug-in:
   Opt 090	Delete probe (HP 54713B plug-in only)
   Opt 0B1	One additional User s and Programmer s Book Set
		(English only)
   Opt 1AK	HP 54714A  & 54722A delete media option
Support Options These support options are available at time of purchase. Yearly contracts can be purchased through your local service center. Each Opt applies to one specific model and serial number.
   Opt +W30	three years of return-to-HP service.
   Opt +W32	three years of HP Calibration service at HP service center.
   Opt 1BN	Mil-std calibration. Product calibration certificate in
compliance with MIL-STD-45662A.
   Opt 1BP	Mil-std cal with data report. Product calibration certificate
and performance test data report in compliance with MIL-STD-45662A.

Probes and Probing Accessories

HP 54701A 2.5-GHz active probe 

	Option 001 	Add APC 3.5(f) bulkhead to N-type(f) adapter 
			to use with the HP 54120 	
	Option 0B1	One additional User and Service guide 
			(English only)	

HP 1141A Differential Probe 	
HP 1142A Probe Control and Power Module
HP 1143A Probe Power Supply	

	Option 001 	One 1.5 meter probe power cable extension cord 
			(HP P/N 01143-61602)	
	Option 0B1 	One additional User and Service Guide 
			(English only)	

HP 54006A 6-GHz passive probe	
HP 10430A 500-MHz, 6.5-pF passive probe (1 m) 	
HP 10441A 500-MHz, 9-pF passive probe (2 m) 	
HP 1141A 200-MHz differential probe (Order HP 1142A also)
HP 1142A probe power supply	
HP 1137A 1000:1 high-voltage divider probe (500 MW, 3 pF, 1-MHz bandwidth) 	
HP 1133A TV/video sync pod	
HP 54300A 16-to-2 probe multiplexer	
HP 54002A 50 W input pod for HP 54300A	
	
HP-IB Cables

HP 10833A 1 m cable	
HP 10833B 2 m cable	
HP 10833C 4 m cable	
HP 10833D 0.5 m cable	

Centronics Cables

HP 13242G 25-pin (m)  to 25-pin (m) Centronics cable (5 m)	
HP 92284A 25-pin (m) to 36-pin (m) Centronics cable (2 m)	
HP 92219Z 25-pin (m) to 15-pin (m) Centronics cable (1 m)	

Rack Mounting Hardware

Support rail kit for all HP racks (HP 12679B)	
Standard slide kit, nontilt (HP P/N 1494-0059)	
Standard slide kit, tilting (HP P/N 1494-0063)	
Slide end brackets for non-HP racks (HP P/N 1494-0061)	
New slide cover required for slides (1 each) (HP P/N 5060-9948)	

Testmobile

HP 1181A system testmobile	
HP 35181A top-mounted printer stand for HP 1181A	
HP 35181C 3.5-inch drawer for HP 1181A	
HP 35181D Work Surface
[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

To help you make the right instrument choices

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


Related HP 54720D Four-channel, 8 GSa/s Real-time Modular Oscilloscope Mainframe product datasheets:

Related product families available online:



Contact HP Test & Measurement. (c) Copyright 1994, 1995, 1996, 1997 Hewlett-Packard Company.