HP 54845A Infinium Oscilloscope up to 4 channels, 1.5 GHz, 8 GSa/s(2 channel), 4 GSa/s(4 channel)

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]
Ultimate Performance

HP Infinium Oscilloscope model 54845A provides superior bandwidth and sample rate to capture high-speed signals accurately. 1.5-GHz bandwidth and 8-GSa/s sample rate assure your signal will be recorded accurately.

New A/D Technology

Based on new analog-to-digital converter technology developed by HP Laboratories, HP Infinium Oscilloscope model 54845A interleaves four 4-GSa/s channels to achieve two 8-GSa/s channels of simultaneous, real-time acquisition, with 64-K memory depth on each channel. The HP 54845A offers a new level of performance in a highly usable and affordable benchtop package.

Turn your HP Infinium oscilloscope into the best "Jitter" measurement tool on the planet.


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]



PERFORMANCE CHARACTERISTICS

ACQUISITION HP 54845A
Maximum Sample Rate  
     Real Time: 2 channel mode:  8 GSa/s
4 channel mode:  4 GSa/s
Maximum Effective Sample Rate Equivalent Time:  500 GSa/s
Memory Depth 2 channel mode: 65,536 points
4 channel mode: 32,768 points
Memory Depth Modes  
     Auto: Optimized for best combination of update rate and display quality
     Manual: Selectable: 2 channel mode: from 16 to 65,536 points
                   4 channel mode: from 16 to 32,768 points
Sampling Modes  
     Real Time: Successive single shot acquisitions
     Equivalent Time: Random Repetitive sampling (higher time resolution at faster sweep speeds)
Filters  
     9-bit Bandwidth Limit
      Filter:
BW = Sample Rate/20
     (Sinx)/x Interpolation: On/Off selectable FIR digital filter. Digital signal processing adds points between acquired data points to enhance measurement accuracy and waveform display quality. BW= Sample Rate/4
Averaging Selectable from 2 to 4096


*Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period, and ±5°C from firmware calibration temperature.

VERTICAL HP 54845A
Number of Channels 4 (simultaneous acquisition)
Analog Bandwidth (-3dB) 50 ohm:    1.5 GHz
1M ohm:  500 MHz (with HP 1161A probe)
System Bandwidth with  
     1161A 10:1 passive
     probe
500 MHz
     1162A 1:1 passive probe 25 MHz
     1163A 10:1, 500 ohm
     passive probe
1.5 GHz
     1152A 2.5 GHz, .6 pF
     active probe
1.3 GHz
     1153A 200 MHz
     differential probe
200 MHz
Real Time Bandwidth* 50 ohm:  1.5 GHz (2 channel mode)
               1.0 GHz (4 channel mode)
1M ohm: 500 MHz
Rise Time1 50 ohm:   233 ps
1M ohm:   700 ps
Sensitivity2  
     1M ohm: 2 mV/div to 2 V/div
     50 ohm: 2 mV/div to 1 V/div
Input Impedance* 1M ohm; ± 1% (12 pF nominal), or 50 ohm ± 1.5%
VSWR (50 ohm) dc to 500 MHz:         1.3
500 MHz to 1 GHz:  1.5
1 GHz to 1.5 GHz:     1.75
Input Coupling dc, ac (7 Hz, available in 1M ohm only)
Maximum Input Voltage*  
     1M ohm: ± 100 V (dc + ac) [ac < 10 kHz], CAT I
     50 ohm: 5 Vrms, CAT I
Channel-to-channel isolation (with channels at equal sensitivity) dc to 100 MHz:         40 dB
100 MHz to 1 GHz:   30 dB
1 GHz to 1.5 GHz:     25 db
Offset Range                     Vertical Sensitivity                 Available Offset
1 M ohm:     2 mV/div to 100 mV/div            ± 4V
                     > 100mV/div to 2 V/div             ± 40V
50 ohm:                 all                                       ± 12 div
Dynamic Range ± 8 div from center screen
Full resolution channel scales 10, 20, 50, 100, 200, 500, 1000 mV/div (plus 2000 mV/div in 1 M ohm)
dc Gain accuracy*2,3 ± 1% of full scale at full resolution channel scale
Resolution2  
     Real Time 8 bits (0.4% of full scale), 12 bits with averaging
     Equivalent Time 8 bits (0.4% of full scale), 12 bits with averaging
Offset Accuracy*2 ± (1% of channel offset + 1% of full scale) at full resolution channel scale
dc Voltage Measurement Accuracy*2,3  
     Dual Cursor ± [(dc gain accuracy)+(resolution)]
     Single Cursor ± [(dc gain accuracy)+(offset accuracy)+ (resolution/2)]
     Example (Single cursor
     accuracy, 70 mV signal,
     10 mV/div, 0 offset
Accuracy = ±[1%(80mV) + (1%(0)+ 1%(80mV)) + (.4%/2)(80mV)] = ±1.76mV
AutoProbeInterface AutoProbe is an intelligent communication and power link between compatible probes and HP Infinium Oscilloscopes. AutoProbe completely configures the scope for the attached probe. For instance, it identifies the probe type and sets up the proper input impedance, attenuation ratio, probe power and offset range, as needed.

1Rise time figures are calculated from tr = 0.35/bandwidth.

2Magnification is used below 10mV/div range and between major attenuation settings. Full scale is defined as the major attenuator setting above an intermediate setting. (Major settings 50 ohm: 10 mV, 20 mV, 50 mV, 100 mV, 200 mV,500 mV, 1 V, 1 M ohm: all of the above plus 2 V).


*Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period, and ±5°C from firmware calibration temperature.

HORIZONTAL HP 54845A
Main Time Base Range 100 ps/div to 5 s/div
Horizontal Position Range (pre trigger) 0 to -1 s or one full screen width, whichever is larger
Horizontal Position Range (post trigger) 0 to 1 s or one full screen width, whichever is larger
Delayed Sweep Range 1ps/div to current main time base setting
Delayed Sweep Delay Range Within main time base acquisition record
Resolution 2ps
Timebase Accuracy 70 ppm (.007%)
Delta-t Accuracy*  
     Real Time mode4 ±[(.007%)(delta-t) + (0.2)(sample period)]
     Equivalent Time mode
     (>= 16 avgs.)
±[(.007%)(delta-t) + (full scale/(2 * memory depth)) + 30ps]
     Example (Equivalent Time mode
     (>= 16 avgs.), 9 ns signal,
     1 ns/div, 1 channel)
Accuracy = ±[(.007%)(9 ns) + ((10 ns)/(2 * 65,536)) + 30ps] = ±[(630 x 10-15) + (76 x 10-15) + (30 x 10-12] = 31 ps
Jitter, RMS 8 ps ± .005% (delay setting)

4For bandwidth limited signals, tr >=1.4 x sample interval.


*Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period, and ±5°C from firmware calibration temperature.

TRIGGER HP 54845A
Sensitivity*2  
     Internal (normal) dc to 100 MHz:          0.5 div
100 MHz to 500 MHz: 1.0 div
500 MHz to 1 GHz:      1.5 div
     Auxiliary dc to 500 MHz:         300 mVp-p
Maximum Input Voltage*  
     Auxiliary 2.5 kohm: ±15 V, CAT I
Minimum Pulse Width (internal, external) 500 ps at > 1.0 div
Level Range  
     Internal ± 8 div from center screen
     Auxiliary ± 5 V
Sweep Modes Auto, triggered, single
Trigger Coupling dc, ac (7 Hz), low frequency reject (50 kHz), high frequency reject (50 kHz)
Trigger Holdoff Range 60 ns - 320 ms
Trigger Modes Edge, Glitch, Pattern, State, Delay by Time, Delay by Events, Violation (Runt, Setup/Hold Time, Pulse Width, Transition), Video, Line
Glitch Select positive or negative polarity, width. Captures glitches as narrow as 500 ps.
Pattern Select inputs as High, Low or X (don't care) to create pattern. Trigger when pattern is entered, exited, present > t, present < t, or present over a range of time. Captures patterns as narrow as 500 ps.
State Select one channel as clock, specify other inputs as High, Low or X. Logic Type: AND or NAND. Setup time is 1 ns and hold time is 0 ns.
Delay by Time Time: 30 ns to 160 ms. The trigger is qualified by an edge. After the delay, a rising/falling edge on any one selected input will generate the trigger.
Delay by Events Events: 1 to 16,000,000 rising or falling edges. The trigger is qualified by an edge. After the delay, a rising/falling edge on any one selected input will generate the trigger.
Violation Trigger  
     Runt Select Positive, Negative or either polarity, thresholds, time qualified. Captures runts as narrow as 500 ps.
     Setup/Hold Modes: Setup, Hold or Setup and Hold. Select clock, thresholds, setup and/or hold time. Edges must be >= 1.5 ns. The difference in slew rates between clock and data must be less than a 2:1 ratio.
     Pulse Width Triggers on pulse width > t or < t. Captures pulses as narrow as 500ps.
     Transition Select Rise Time or Fall Time, present > t or present < t, thresholds. Captures edges as fast as 800 ps.
Accuracy (time) for glitch, pulse width and time-qualified pattern 1.5ns-20ns: ± (20% setting+ 500 ps)
20 ns - 160 ms: ± (3% setting + 2 ns)
Video Triggering 525 lines/60 Hz (NTSC), 625 lines/50 Hz (PAL), 875 lines/60 Hz. Trigger on Field 1 or Field 2, any line. User defined triggering: User can specify sync pulse level, width and polarity, edge number.

2Magnification is used below 10 mV/div range and between major attenuation settings. Full Scale is defined as the major attenuator setting above an intermediate setting. (Major settings 50 ohm: 10 mV, 20 mV, 50 mV, 100 mV, 200 mV, 500 mV, 1 V, 1 M ohm: all of the above plus 2 V).


DISPLAY HP 54845A
Display 8.4 inch diagonal color active matrix LCD module incorporating amorphous silicon TFTs.
Annotation Up to 12 labels, with up to 100 characters each, can be inserted in the waveform display area. Labels can be saved to and recalled from image files and setup files.
Active Display Area 6.73"x 5.04" (33.92 sq. in.) 171 mm x 128 mm (21,888 sq. mm)
Waveform Viewing Area 4.10"x 6.25" (25.6 sq. in.) 104 mm x 159 mm (16,536 sq. mm) in full screen mode
Display Resolution 640 pixels horizontally x 480 pixels vertically
Waveform Colors Select from 100 hues, 0-100% saturation and 0-100% luminosity.
Dual Intensity Infinite Persistence Previous sweeps are stored in half-bright display and recent sweeps in full-bright. This allows easy differentiation of current and historic information.
Waveform Overlap When two waveforms overlap, a third color distinguishes the overlap area.
Full screen mode On/Off selectable
Connect-the-dots On/Off selectable
Persistence Minimum, Variable (Up to 6 levels of gray scale, 100 ms to 40 s), Infinite
Graticule On/Off (Grid/Frame)
Grid Intensity 0 to 100%
Display Update Rate Waveforms/sec                    > 1,950
Vpp Measurements/sec        > 65
     Measurement
     Conditions
Real Time mode, minimum persistence display mode, triggered sweep mode, no interpolation, markers off, math off, connect the dots off, 1 channel acquisition, 50 ns/div, statistics off, 512 pt record (2GSa/s)


MEASUREMENTS HP 54845A
Automatic Parametrics 27 automatic measurements: Vpp, Vmin, Vmax, Vavg, Vamptd, Vbase, Vtop, Vrms, Preshoot, Overshoot, Vupper, Vmiddle, Vlower, Rise Time, Fall Time, Period,Frequency, Positive Width, Negative Width, Duty Cycle, Delta Time, Tmax, Tmin, FFT Frequency, FFT Magnitude, FFT Delta Frequency, FFT Delta Magnitude. Over HP-IB only: VTime, TVolts
Threshold Definition Selectable 10%, 50%, 90% or 20%, 50%, 80% or Custom (% or absolute voltage)
Top-Base Definition Standard or Custom (in absolute voltage)
Statistics On/Off selectable. Current measurement, mean and standard deviation
Measurement Toolbar 16 Drag and Drop automatic measurement icons.
QuickMeas Activates 4 preselected automatic measurements, measurements are configurable
Markers Modes Manual Markers, Track Waveform Data, Track Measurements
Waveform Math 4 functions f1-f4. Select from Add, Subtract, Multiply, Divide, Invert, Magnify, Vs, Min, Max, Integrate, Differentiate, FFT Magnitude


FFT HP 54845A
Frequency Range5 2 channel mode: dc to 4 GHz (Sample rate/2)
4 channel mode: dc to 2 GHz (Sample rate/2)
Frequency Resolution Sample rate/Memory depth
     Best Resolution 2 channel mode: 7.63 µHz (Sample ratemin/65,536 pts)
4 channel mode: 15.3 µHz (Sample ratemin/32,768 pts)
     Best resolution at Sample ratemax 2 channel mode: 122 kHz (Sample ratemax/65,536 pts)
4 channel mode: 122 kHz (Sample ratemax/32,768 pts
Frequency Accuracy (1/2 frequency resolution)+(7x10-5) (signal frequency)
Amplitude Display Power in dBm
Signal-to-noise ratio 70 dB at 32K memory depth. Noise floor varies with memory depth and with averaging.
Window Modes Hanning, Flattop, Rectangular

5FFT amplitude readings are affected by input amplifier roll-off (-3 dB at 1.5 GHz, with amplitude decreasing as frequency increases above 1.5 GHz).


COMPUTER SYSTEM/STORAGE HP 54845A
CPU AMD-K5 PR 133 Microprocessor
Disk Drives 1.4 GByte internal hard drive. Storage capacity is limited only by disk space. 3.5" MS-DOS® compatible, high-density, floppy disk drive. Store and recall setups, waveforms, and store screen images to both the hard drive and the floppy drive.
File types nbsp;
     Waveforms Internal, verbose,Y values
     Images BMP, EPS, GIF, PCX, PS (Postscript), CSV
Mouse Standard mouse supplied. Supports any Microsoft® mouse-compatible pointing device, serial or PS/2
Waveform Memories 4 nonvolatile waveform memories


I/O HP 54845A
LAN Enables data/setup file transfers and use of network printers; supports popular network operating systems including Novell NetWare, Microsoft, Banyan VINES, SCO UNIX and IBM; 10 MBPS operation that complies with IEEE 802.3 Ethernet and ISO/IEC 8802-3 Ethernet standards; TCP/IP protocol; RJ-45, BNC, and AUI connectors.
HP-IB Fully programmable, complies with IEEE 488.2
RS-232 (serial) 2 ports: COM1, COM2. Printer and pointing device support
Centronics Printer support
Printers and Plotters Supports all printers and plotters compatible with Microsoft Windows® 95. Includes but is not limited to HP DeskJet- and LaserJet-series printers. HP-IB devices not supported.
PS/2 port For PS/2 mouse
Keyboard port For optional keyboard
Video Output 15 pin VGA, full color


GENERAL HP 54845A
Meets Hewlett-Packard's Environmental Specification (section 750) for class B-1 products with exceptions as described for temperature.
Temperature  
     Operating: 10°C to + 40°C
     Nonoperating: - 40°C to 70°C
Humidity  
     Operating: Up to 95% relative humidity (noncondensing) at +40°C
     Nonoperating: Up to 90% relative humidity at +65°C
Altitude  
     Operating: Up to 4,600 meters (15,000 ft.)
     Nonoperating: Up to 15,300 meters (50,000 ft.)
Vibration  
     Operating: Random vibration 5-500 Hz, 10 minutes per axis, 0.3g(rms)
     Nonoperating: Random vibration 5-500 Hz, 10 minutes per axis, 2.41g(rms); resonant search 5-500 Hz, swept sine, 1 octave/minute sweep rate, (0.75g), 5 minute resonant dwell at 4 resonances per axis.
Power Voltage: 100-240VAC, +/- 10%,47 to 440Hz; Max power dissipation: 390 W
Weight Net:         12 kg (26.5 lb)
Shipping: 15 kg (33 lb)
Dimensions (excluding handle) Height: 216 mm (8.50 in); Width: 437 mm (17.19 in); Depth: 440 mm (17.34 in)
Safety Meets IEC1010-1 +A2, CSA certified to C22.2 No.1010.1, Self certified to UL 3111



[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP 54845A      Infinium Oscilloscope up to 4 channels, 1.5 GHz,
               8 GSa/s (2 channel), 4 GSa/s (4 channel)


The HP 54845A includes:
1  C3751-60201     PS-2 Mouse
1  54810-85901     HP Infinium Oscilloscope Mouse pad
1  E2610A          Keyboard
1  54810-97000     User's Quick Start Guide, English language
1  54845-99000     Service Guide
1  54810-99000     Programmer s Guide
1  54810-97002     Programmer s Quick Reference Guide
4  1161A           10:1 10 M ohm passive probes
1  54810-68701     Accessory Pouch
1  US power cord
   Three-year warranty

Options
   Opt 090     Delete standard probes
               4 1161A probes

   Opt 001*    Additional standard probes
               2 1161A probes

   Opt 002*    Add 1 1162A 1:1 passive probe
   Opt 003*    Add 1 1163A 10:1 500 ohm, low C passive probe
   Opt 006*    Add 1 1152A 2.5 GHz, .6 pF active probe
   Opt 008*    Add 1 1153A 200 MHz differential probe
   Opt 1BP     Mil Std 45662A and ANSI/NCSL Z-540 calibration with test data
   Opt 1CL*    Add 1 Keyboard (E2610A)
   Opt 1CM*    Add 1 Rackmount kit (E2609A)
   Opt UL3*    Add 1 Clip-on track ball pointing device (E2611A)
   Opt UL5*    Add 1 Touchpad pointing device (E2612A)
   Opt W50     Five-year, customer-return, repair coverage (additional 2 years)

* Multiple options can be ordered

Probes listed above as options can also be ordered by model numbers. HP 1144A 800 MHz Active Probe (requires HP 1142A power supply) HP 01144-61604 Probe power extender (required when using more than 2 HP 1144A active probes) HP 1145A 2-channel, 750 MHz SMT active probe (requires HP 1142A power supply) HP 1146A Oscilloscope AC/DC Current Probe HP 1137A 5kV, 1000:1 High Voltage Probe HP 1182A Testmobile HP-IB cables HP 10833A 1 m (3.3 ft.) HP 10833B 2 m (6.6 ft.) HP 10833C 4 m (13.2 ft.) HP 10833D 0.5 m (1.6 ft.) Centronics Printer cable C2951A 3 m (9.9 ft.)

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP 54810/15/20/25/45A Infiniium Oscilloscopes ( PDF)
Probes and Accessories for HP Infiniium Oscilloscopes ( PDF)

If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader now. Get Acrobat

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]
Additional help is a click away!

Want to talk to a knowledgeable engineer about this product at our HP Test and Measurement Call Center? Do you have a question you want to send in by e-mail? A measurement set-up you need to fax in? Follow the link above and you will be given a choice to contact an engineer by telephone, e-mail or fax.

This service is available only in the United States. For assistance outside the U.S., contact your local HP DIRECT or HP Sales Office.

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Related Information:


Related HP E3830 Series Wide Bandwidth Signal Analysis Systems product datasheets:


Related HP Infinium Oscilloscopes product datasheets:



Privacy Statement    Use of this site indicates that you accept the Terms of Use
Contact HP Test & Measurement. (c) Copyright 1994 - 1998 Hewlett-Packard Company.

Page Updated: Thursday February 18 10:56:13 UTC 1999