HP 5501B Transducer Laser Head![]()
|
The HP 5501B is configured with appropriate optics and system electronics to build a laser interferometer system that meets the unique physical layout and measurement requirements of individual applications. The custom positioning system improves product quality, increases yields, and facilitates manufacture of precision products.
Physical Characteristics
Weight: 3.4 Kilograms (7.5 pounds) Magnetic Field Strength (Non-Operating): Does not exceed 5.25 milligauss at a distance of 4.6 m (15 ft) from any point on the surface of the packaged Laser Head. Clearance Required for Cabling: 10.16 cm (4 in) beyond back
Power
Power Input Requirements: +15 Volts +/-0.3 Volts at 0.79 Amperes maximum =15 Volts +/- 0.3 Volts at 0.67 Amperes maximum Heat Dissipation: 23 watts (during operation) 35 watts (during warmup) Warmup Time: less than 10 minutes (4 minutes typical)
Laser Beam Characteristics
Maximum Beam Power Output: 1 milliwatt Minimum Beam Power Output: 180 microwatts Beam Diameter: 6 millimeters (0.25 inch) typical Vacuum Wavelength Stability: 0.002 ppm typical 1 hour 0.02 ppm typical lifetime Nominal Vacuum Wavelength: 632.991372 nanometers Safety Classification: Class II Laser Product conforming to U.S. National Center for Devices and Radiological Health Regulations 21 CFR 1040.10 and 1040.11.
Outputs
Reference Frequency: 1.5 - 2.0 MHz
HP 5501B Transducer Laser Head Opt 1BN Mil std 45662A calibration certification Opt 1BP Mil std 45662A calibration w/ test data Opt 910 One Additional Manual Opt AV4 User's manual Opt UK6 Commercial cal. certificate w/ test data Opt W30 3 Yr. Return Repair Service Opt W50 5 Year Return Repair Service
HP 10778A Power Cable HP 10778B Power Cable HP 10778C Power Cable HP 10794B Laser Head Cable for 5507B electronics The HP 5501B is used with the HP 5507B electronics (HP 5527B system).