HP 5517B Laser Head

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]
The HP 5517B Laser Head, used in HP interferometer systems, offers a vacuum wavelength accuracy of +/- 0.1 ppm, 500 mm per second axis velocity for linear optics, and beam diameter of 6 mm. The laser head uses a proven long-life laser tube with a demonstrated Mean Time Between Failure (MTBF) of greater than 50,000 hours of operation.

The HP 5517B is configured with appropriate optics and system electronics to build a laser interferometer system that meets the unique physical layout and measurement requirements of individual applications. The custom positioning system improves product quality, increases yields, and facilitates manufacture of precision products.


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Physical Characteristics

Weight: 3.4 kg (7.5 lb)
Warm-Up Time: Less than 10 minutes (5 minutes typical)
Magnetic Field Strength (Non-operating): Does not exceed 5.25 milli-Gauss at 
	a distance of 4.6 m (15 ft) from any point on the surface of the packaged 
	Laser Head.
Clearance required for cabling: 10.16 cm (4.0 in) beyond back of unit

Power

Power Requirements:
	+15V +/- 0.3V at 2.2A maximum
	-15V +/- 0.3V at 0.02A maximum
Heat Dissipation (nominal):
	Warm-up:	35W
	Operation:	23W

Laser Characteristics

Type: Helium-Neon, Continuous Wave, Two-Frequency
Maximum Beam Power Output: 1 mW
Std. Beam Diamter: 6 mm (0.25 in) typical
Vacuum Wavelength Accuracy (3 sigma, lifetime): +/-0.1 ppm (+/-0.02 ppm with 
	factory calibration to MIL-STD 45663) Wavelenght Accuracy includes lifetime 
	stability effect.
Nominal Vacuum Wavelength: 632.991372 nm
Vacuum Wavelength Stability: 
	+/-0.002 ppm typical:	1 hour
	+/-0.02 ppm typical:	Lifetime

Safety Classification

Class 2 Laser Product congforming to U.S. Nation Center for Devices and Radiological
Health Regulations 21 CFR 1040.10 and 1040.11.

Reference Frequency

1.9-2.4 MHz


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP 5517B	Laser Head
   Opt 030	Replace Standard Feet with 3 Electrically-Insulated Feet
   Opt 1BN	MIL-STD 45662A Calibration
   Opt 1BP	MIL-STD 45662A Calibration with Test Data
   Opt W50	5-Year Return Repair Service
   Opt UK6	ISO 9001/2 Commercial Calibration Certification with Test Data

The HP 5517B can be used in a variety of laser-interferometer positioning systems, including the HP PC-compatible and VMEbus compatible systems.

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

To help you make the right instrument choices

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Related product families available online:



Contact HP Test & Measurement. (c) Copyright 1994, 1995, 1996, 1997 Hewlett-Packard Company.

Page Updated: Thursday November 13 08:22:29 UTC 1997