The HP 59510A is a remote-controlled relay device. It can be configured to switch dc power in multiple test fixtures or provide extra protection when a fault condition requires an emergency shutdown.


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Operating Ranges: 200 V at 20 A, 120 V at 30 A, 48 V at 60 A
dc Floating Voltage: Input to output, 200 Vdc; input or output to ground, 500 Vdc; 
                     TTL control to ground, 240 Vdc      
Settling Time: Connect, 440 ms; disconnect, 160 ms; polarity reversal, 600 ms
dc Voltage Drop (at 60 A): 0.5 V max. on each relay
ac Input: Can be set for 100, 120, 220, or 240 Vac (-13%, +6%) at 48 Hz to 63 Hz
Physical Specifications
Dimensions: 185.4 mm W x 130.6 mm H x 198.6 mm D 
               (7.26 in x 5.14 in x 7.81 in)
Weight: 2.3 kg (5 lbs)


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP59510A	Relay Accessory
   Opt 850	Rack Mount Kit (side-by-side mounting of two units.  Requires two kits)
   Opt 910	Additional Operating and Service Manual


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

No literature is available online at this time.

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

To help you make the right instrument choices

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Related product families available online:



Privacy Statement    Use of this site indicates that you accept the Terms of Use
Contact HP Test & Measurement. (c) Copyright 1994 - 1998 Hewlett-Packard Company.

Page Updated: Monday October 05 19:50:16 UTC 1998