HP 66311A Mobile Communications DC Source, 15V, 3A
| |||
![[ Summary ]](/tmo/datasheets/Graphics/tab-summary-high.gif)
![[ Features ]](/tmo/datasheets/Graphics/tab-features-reg.gif)
![[ Specifications ]](/tmo/datasheets/Graphics/tab-specs-reg.gif)
![[ Key Literature ]](/tmo/datasheets/Graphics/tab-keylit-reg.gif)
![[ Ordering ]](/tmo/datasheets/Graphics/tab-ordering-reg.gif)
![[ Assistance ]](/tmo/datasheets/Graphics/tab-assistance-reg.gif)
Output Ratings:
Voltage 15V
Average current at 45 deg C 3 A
derates linearly to 2.5 A at 55 deg C
Peak Current for 7 ms, average current 5 A
cannot exceed 3A
Programming Accuracy (25 deg C +/-5 deg C):
Voltage/+ Current 0.05% + 10 mV/1.33 mA
Ripple and Noise (20 Hz to 20 MHz, with outputs ungrounded
or with either terminal grounded):
Voltage (rms/p-p) 1 mV/6 mV
Current (rms): 2 mA
DC Measurement Accuracy:
via HP-IB or front panel meters with respect to actual
output at 25 deg C +/- 5 deg C
Voltage: 5 mV + 0.03%
Low current Range (-20 mA to +rated I): 2.5 uA + 0.1%
High current range (+20 mA to +rated I) 0.5 mA + 0.2%
(-20 mA to -rated I) 1.1 mA + 0.2%
Load Regulation: 2 mV, 0.75 mA
Line Regulation: 0.5 mV, 0.75 mA
Supplemental Characteristics
Transient Response Time: Less than 35 us for the output voltage to recover
to within 20 mV of final value after 0.1 to 1.5 A load change when
operating in the high mode compensation (loads from 5-12,000 uF).
Average Programming Resolution: 3.75 mV/ 0.67 mA
Sink current: 2 A @ 7.5 V
Physical Specifications
Weight: 9.07 kg (20 lb) net
11.1 kg (24.5 lb) shipping
Size: 212.8 mm W x 88.1 mm H x 425.8 mm D
(8.4 in x 3.5 in x 16.8 in)
HP 66311A 45 Watt Mobile Communications dc Source
Opt 8ZJ Delete instrument feet
Otp 1CM Rackmount kit
Opt AXS Rackmount kit for side-by side mounting
locking kit HP p/n 5061-9694
flange kit HP p/n 5062-3974
Opt OBN Service manual and extra operating guide
Opt 052 Device Characterization Software
Opt 080 Enhanced RFI Suppression 800-1000 MHz
AC Line Voltage Options
Opt 100 87 to 106 Vac, 47 to 63 Hz, 3.5 A
Opt 220 191 to 233 Vac, 47 to 63 Hz, 1.6 A
Opt 230 207 to 253 Vac, 47 to 63 Hz, 1.5 A
Rack mount kit and slide for side-by side mounting of two
different depth instruments (HP p/n 5062-3966 and 1494-0015)
Rackmount, slide and filler panel for mounting one
instrument (HP p/n 5062-3996, 1494-0015, 5062-4022)
Available accessories for this product include:
No literature is available online at this time.
To help you make the right product choicesCustomers outside the U.S. can obtain assistance through local HP DIRECT help.
Related Information:
HP Model 66311A Mobile Communications DC Source View the HP 66311A Mobile Communications DC Source technical specifications.
Minimizing Power Supply Transient Voltage at a Digital Wireless Telecommunications Test Fixture This paper addresses the problem of maintaining a stable, transient-free voltage at a DUT (device under test).
HP VXI plug&play drivers for HP 66311A Mobile Communications DC Source
HP 66312A, HP 66332A Dynamic Measurement DC Sources - are power source and measurement solutions for testing portable device offering dynamic measurement and analysis that provides up to 20 V and 100 W.
HP 14565A Device Characterization Software designed for easy evaluation of Battery Powered Device current profiles when used with HP Dynamic Measurement DC Sources.
HP 14565A Device Characterization Software designed for easy evaluation of Battery Powered Device current profiles when used with HP Dynamic Measurement DC Sources.