HP 6842A Harmonic/Flicker Test System

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]
The HP 6840 Series Harmonic/Flicker Test Systems are specifically designed for testing products for compliance to the IEC low-frequency emissions standards and drafts for quai-stationary current harmonics, fluctuating current harmonics, and voltage fluctuations.

An interactive demonstration of HP's 6840 Series Harmonic/Flicker Test System is available for you to download.

Recommended system requirements for this demonstration are:
 
   396 or faster PC with Microsoft® Windows®
   8 MB or more of RAM
   Approximately 7 MB of hard disk space
   3.5", high-density floppy disk drive (1.44 MB)
   VGA video display (16 colors, 640 x 480 or higher resolution)
  Harmonic/Flicker Test System - Demonstration Software is available for download as a self-extracting MS Windows file. Note: Click "OK" when asked about Disk #2 during installation.


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

  • One-box solution tests products for compliance to existing and standards for: IEC 555-2, IEC 555-3, EN 60555-2, EN 60555-3, IEC 1000-3-2, IEC 10003-3, EN 61000-3-2, and EN 61000-3-3

  • Full 1 phase coverage at 230 Vrms and 16 Arms

  • Measurement and generation implementation compliant with IEC 868, IEC 1000-7-4, and IEC 725

  • Compliant-level current harmonic measurements for the fundamental through the 40th harmonic

  • Windows graphical user interface for IEC testing capability

  • Real-time and off-line test data review and analysis

  • Test report generation

  • Advanced diagnostic capabilities for failure analysis

  • Standard ac source operating mode (Normal mode)


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Output ratings: 
       Power: 1750 VA
       rms voltage range: 0-300 V
       Maximum rms current: 13 A
Measurement accuracy (25 deg C +/-5 deg C):
       Current magnitude (low range)
            Fundamental: 0.03% + 1.5 mA
            Harmonics 2-49: 0.03% + 1 mA + 0.2%/kHz
       Current Magnitude (high range)
            Fundamental: 0.05% + 5 mA
            Harmonics 2-49: 0.05% + 3 mA + 0.2%/kHz
       Flicker: Compliant with IEC 868
       Flicker perceptibility (pst): Compliant with IEC 868
       Synchronixation accuracy: <1 ppm
       Current shunt burden: 1.5 seconds
       Power (watts low range):  0.1% + 0.3 W + 1.2 mW/V (45 to 100 Hz)
                                 0.1% + 1.2 W + 1.2 mW/V (>100 to 500 Hz)
                                 0.1% + 2.5 W + 1.2 mW/V (>500 to 1000 Hz)
       Power (watts high range): 0.1% + 0.3 W + 12.7 mW/V (45 to 100 Hz)
                                 0.1% + 1.2 W + 12.7 mW/V (>100 to 500 Hz)
         0.1% + 2.5 W +12.7 mW/V (>500 to 1000 Hz)
Load regulation: 0.5% of full scale
Line regulation: 0.1% of full scale

Physical Specifications                                                                 
Weight: 28.2 kg (62 lb)
Dimensions: 425.5 mm W x 132.6 mm H x 574.7 mm D 
            (16.75 in x 5.25 in x 22.6 in)


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP 6842A	Single-phase AC Power Source/Analyzer
   Opt 1CM	Rack Mount Kit (HP 5062-3977) Support rails required
   Opt 1CP	Rack mount kit with handles (HP 5062-3983)
   Opt 200	174-220 Vac, 48-63 Hz.  If  Opt 200 is not ordered, this
                unit will be configured to operate at 191 to 254 Vac, 47
                to 63 Hz
   Opt 831	12 AWG, 200 to 240 Vac, unterminated line cord
   Opt 832	4  mm wire size, 200 to 240 Vac, unterminated line cord
   Opt 841	Line cord with NEMA 6-20P; 20 A, 250 V plug
   Opt 842	Line cord with IEC 309; 32 A, 220 V plug
   Opt 844	Line cord with NEMA L5-30P; 30 A, 250 V locking plug
   Opt 0BN	Extra Documentation

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

All orders include the following:

Trademarks:

    Microsoft is a U.S. registered trademark of Microsoft Corporation.
    Windows is a U.S. registered trademark of Microsoft Corporation.
    

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

To help you make the right instrument choices

    If you live in the U.S., call HP DIRECT at 1-800-452-4844. It's easy to open an account or make a purchase with VISA, MasterCard or American Express, Plus, we've taken the risk out of buying instruments: If you're not satisfied with your purchase for any reason, return it in original condition within 60 days for full refund or credit.

    Customers outside the U.S. can also get local HP DIRECT help

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


Related HP 6840 Series Harmonic/Flicker Test Systems product datasheets:

Related product families available online:



Contact HP Test & Measurement. (c) Copyright 1994, 1995, 1996, 1997 Hewlett-Packard Company.

Page Updated: Thursday November 13 08:17:17 UTC 1997