HP 70110A Digital Multimeter![]()
|
All functions, including switching to guard to the low and switching between front and rear inputs, are accessible using SCPI-compatible remote commands. Specialized signals (voltmeter complete and external trigger) used in ATE systems to synchronize fast measurements with external multiplexers are available on the rear panel.
An electronic Auto Cal, taking less than 30 seconds without an external source, is used to meet 24 hour specifications. When calibration to an external standard is required, a security code allows you to enable the calibration routines. This electronic key, plus a system which tracks the number of calibrations performed, protects against calibration tampering.
Electrical measurement performance closely parallels that of the HP 3457A multimeter. Selected specifications are listed below. (For complete specifications, contact your HP sales represenative) DCV Accuracy 300 mV range: 5 ppm 3 V range: 3.5 ppm DC Resistance Accuracy: 52 ppm Resolution: 10 ต ohms DC Current Accuracy, 300 mA, 2 A: 0.03% Accuracy Other Ranges: 20 ppm Resolution DCV: 10 nV ACV: 10 nV Maximum Input Range HI to LO: 450 Vdc, 450 Vac peak LO to Guard: 200 Vdc, 200 Vac
HP 70110A Digital Multimeter Opt 001 Add armature relay multiplexer Opt 002 Add reed relay multiplexer Opt 910 Extra user manual set Opt 915 Service Guide and schematic package Opt W30 3 Year Return Repair Service Opt W32 3 Year Customer Return Calibration Service Opt W34 3 Year Standards Compliant Calibration Service Opt W50 5 Year Return Repair Service Opt W52 5 Year Customer Return Calibration Service Opt W54 5 Year Standards Compliant Calibration Service
Related product families available online:
Page Updated: Wednesday January 07 23:19:56 UTC 1998