HP 70301A Microwave Tracking Generator, 2.7 GHz to 18 GHz![]()
|
Frequency Specifications
Range: 2.7 to 18 GHz Accuracy (<= 1 0 MHz synthesized span): ±[(freq x ref¹) + 1% span + 15 Hz] Tracking Drift: <3 Hz/hr Tracking Adjust Range: ±500 Hz in 1 Hz steps Offset Range: ±5 MHz w/ext 21.4 MHz ¹Frequency reference accuracy: ±1.3 x 10e-5 per year standard, ±5.0 x 10e-10 per day with HP 70310A
Power Specifications
Maximum Leveled Power: -2 dBm, (0 dBm 20 to 30°C) Amplitude Control Range: +14 to -66 dBm Vernier Range: 11 dB (0.1 dB resolution) Amplitude Accuracy (20 to 30°C) Absolute: ±0.5 dB @ 2.7 GHz Flatness: ±1.0 dB Incremental: ±0.15 dB/dB, ±0.8 dB total Total absolute accuracy:±2.3 dB Amplitude drift (characteristic): < ±0.05 dB/°C @ -2 dBm Output Attenuator Range: 55 dB in 5-dB steps Repeatability: <= ±0.2 dB for any setting
Spectral Purity
Noise Sidebands @ -2 dBm, 6 GHz (10 kHz): < -90 dBc/Hz (characteristic) Sidebands: < -70 dBc (characteristic) Spurious (max leveled power) Harmonics 2nd: < -7 dBc (< -15 dBc typical) 3rd: < -11 dBc (< -15 dBc typical) Non-harmonic: < -60 dBc Sub-harmonic: none RF Off Residuals: < -120 dBm tracking < -65 dBm LO emissions
Temperature Operating: 0 to 5°C Storage: -40 to +75°C
System Specifications
For the HP 71209A Spectrum Analyzer (preselected) with the HP 70300A and 70301A Tracking Generators Signal Frequency Range: 50 kHz to 22 GHz Scalar Frequency Range with HP 70300A: 50 kHz to 2.9 GHz with HP 70301A: 2.7 to 18 GHz Scalar Dynamic Range at 2.0 GHz: >128 dB at 10 GHz: >130 dB at 18 GHz: >130 dB Tracking Generator Feedthrough (dBm in 10 Hz BW) at 2.0 GHz: < 138 dBm at 10 GHz: < 130 dBm at 18 GHz: < 130 dBm
HP 70301A Microwave Tracking Generator, 2.7 GHz to 18 GHz Opt 001 Delete 55 dB Input Attenuator Opt 098 CPU and RAM/ROM Upgrade (HP 80680A)¹ Opt 099 RAM/ROM Upgrade (HP 80681A)¹ Opt 1BN Certificate of Calibration Opt 1BP Certificate of Calibration and Data Opt 871 Scalar Personality Opt 910 Extra User's Manual Opt 915 Service Manual Set Opt W30 3 Year Return Repair Service Opt W32 3 Year Calibration Customer Return Service Opt W34 3 Year Standards Compliant Calibration Service Opt W50 5 Year Return Repair Service Opt W52 5 Year Customer Return Calibration Service Opt W54 5 Year Standards Compliant Calibration Service ¹If this module will be installed into an existing system, a firmware upgrade may be required. Modules with firmware version B.04.02 (920325) and earlier require upgrade HP 70860A. Modules with later version firmware may use upgrade HP 70861A. For more information see service note 70900A-14L / 70900B-01G.