HP 70310A Precision Frequency Reference

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]
The HP 70310A precision frequency reference, a 1-slot wide plug-in module for the HP 70001A mainframe, provides precision reference signals for use by HP 70000 systems and other instruments. System frequency accuracy is enhanced because the 10 MHz and 100 MHz reference signals provided by the HP 70310A are phase-locked to an internal ovenized oscillator, resulting in excellent stability over both time and temperature. With a highly accurate and stable reference in your system, you can find signals more quickly by tuning to very narrow spans. Option 002 deletes the ovenized oscillator and the accessory power pack from the HP 70310A, but retains the external reference input capability. This is a good low cost solution if you plan to connect a house standard or other external reference permanently to your HP 70000 modular spectrum analyzer.


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

General Characteristics

Frequency Reference Accuracy Aging: <5 x 10-10/day (7-day average) <1 x 10-7/year Temperature Stability: <7 x 10-10 over 0° to 55 °C Warm-up Time: <96 hours to meet aging rate specification after <24-hour off period <30 days to meet aging rate specification after indefinite off period <30 minutes to be within 1 x 10-8 of 24-hour warm-up frequency (at 25 °C) 100 MHz Output Power: 0 dBm (characteristic) 10 MHz Output Power: 0 dBm (characteristic) Harmonics: <-20 dBc Spurious: <-80 dBc External Reference Input Input Power Range: -5 to +21 dBm Impedance: 50 ohms (characteristic) Input signal characteristics required to meet HP 71100C, 71200C, and 71210C system specifications (10 MHz input) Offset Phase noise Spurious -------------------------------------------------------- 10 Hz <-95 dBc/Hz <-135 dBc 100 Hz <-125 dBc/Hz <-135 dBc 1 kHz <-145 dBc/Hz <-125 dBc HP 70000 spectrum analyzer frequency accuracy (0° to 55 °C) With HP 70310A Without HP 70310A ----------------------------------------------------------------- At 1 GHz Span No Aging ±17 Hz ±10 kHz After 1 yr Aging ±117 Hz ±13 kHz The accessory power pack included with the HP 70310A supplies dc power to the module to maintain oven temperature when the mainframe is turned off. The power pack mounts to the rear of the mainframe and may be operated using ac power with the following characteristics: 47 to 444 Hz and 100, 120, or 240 Vac. Weight: 3.0 kg (6.6 lb) characteristic Dimensions: 1-slot module


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP 70310A     Precision Frequency Reference
   Opt 002    Delete Ovenized Oscillator
   Opt 1BN    Certificate of Calibration
   Opt 1BP    Certificate of Calibration and Data
   Opt 910    Extra Installation and Verification Manual (HP 70310-90059)
   Opt W30    3 Year Return Repair Service
   Opt W32    3 Year Calibration Customer Return Service
   Opt W34    3 Year Standards Compliant Calibration Service
   Opt W50    5 Year Return Repair Service
   Opt W52    5 Year Customer Return Calibration Service
   Opt W54    5 Year Standards Compliant Calibration Service


The HP 70310A is standard in the spectrum analyzers, lightwave signal
analyzers, and GSM testers.  It also can be ordered separately for any
modular measurement system.

¹If this module will be installed into an existing system, a firmware
upgrade may be required. Modules with firmware version B.04.02 (920325)
and earlier require upgrade HP 70860A. Modules with later
version firmware may use upgrade HP 70861A. For more information
see service note 70900A-14L / 70900B-01G.

All orders include the following:

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

No literature is available online at this time.

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]
Additional help is a click away!

Want to talk to a knowledgeable engineer about this product at our HP Test and Measurement Call Center? Do you have a question you want to send in by e-mail? A measurement set-up you need to fax in? Follow the link above and you will be given a choice to contact an engineer by telephone, e-mail or fax.

This service is available only in the United States. For assistance outside the U.S., contact your local HP DIRECT or HP Sales Office.

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Related product families available online:



Privacy Statement    Use of this site indicates that you accept the Terms of Use
Contact HP Test & Measurement. (c) Copyright 1994 - 1998 Hewlett-Packard Company.

Page Updated: Thursday February 18 08:03:47 UTC 1999