HP 70600A Microwave Preselector, 2.7 GHz to 22 GHz

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]
The HP 70600A provides optional preselection for the HP 71200C microwave spectrum analyzer. Without a preselector, measuring signals of unknown frequencies can be time-consuming. A preselector eliminates the image and multiple responses generated by the harmonic mixing technique used in the spectrum analyzer. When more than one input signal is present or when the frequency of a signal is unknown, use a preselector for quick signal measurements. A built-in attenuator in the preselector provides overload protection and allows you to set the proper input level to the HP 70905A/B RF section for optimum dynamic range. The HP 70600A preselector can be peaked easily for best amplitude accuracy. The preselector tuning is stored as DAC values that can be recalled for consistent readings when automating your measurements. Enhance amplitude accuracy especially when tuning over large spans or changing frequency bands, by using the preselector hysteresis function. Simply activate the function to remove excess magnetic h ysteresis from the YIG-tuned preselector.


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

General Specifications

Frequency Range:		50 kHz to 22 GHz
Preselected Range:		2.7 to 22 GHz; low pass filter below 2.9 GHz
Attenuator:			0 to 70 dB in 10 dB steps
RF Input:			type N, 50 ohms (nominal)
LO Emissions (10 dB attenuation, nominal)
	Bypass Mode:		< -10 dBm (0 - 22 GHz)
	Filtered Mode:		< -50 dBm (0 - 2.9 GHz);
				< -80 dBm (2.7 - 22 GHz)
VSWR (nominal)

	Frequency (GHz) 	VSWR (>=10 dB attn)
	---------------------------------------------------------
	0 -12.7			< 1.9:1
	2.5 - 22/26.5 		< 2.2:1

RF Output:			SMA, 50 ohms (nominal), front panel
Tune + Span:			input SMB, 1.5 V/GHz,
				1 M ohms (nominal), rear panel


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP 70600A	Microwave Preselector, 2.7 GHz to 22 GHz
   Opt 098	Controller Board Upgrade Kit HP 70860A¹
   Opt 099	RAM/ROM Board Upgrade HP 70861A¹
   Opt 1BN	Certificate of Calibration
   Opt lBP	Certificate of Calibration and Data
   Opt 910	Extra User's Manual
   Opt 915	Service Manual Set
   Opt W30      3 Years Return Repair Service
   Opt W32      3 Years Customer Return Calibration Service
   Opt W34      3 Years Customer Return Standards Compliant Calibration Service
   Opt W50      5 Years Return Repair Service
   Opt W52      5 Years Customer Return Calibration Service
   Opt W54      5 Years Customer Return Standards Compliant Calibration Service


¹If this module will be installed into an existing system, a firmware
upgrade may be required. Modules with firmware version B.04.02 (920325)
and earlier require upgrade HP 70860A. Modules with later
version firmware may use upgrade HP 70861A. For more information
see service note 70900A-14L / 70900B-01G.

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

To help you make the right instrument choices

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Related product families available online:



Contact HP Test & Measurement. (c) Copyright 1994, 1995, 1996, 1997 Hewlett-Packard Company.

Page Updated: Thursday November 13 07:46:14 UTC 1997