HP 70620B Preamplifier

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]
Boost the sensitivity of any HP 70000 series spectrum analyzer by 15 to 25 dB. System noise figures can be better than 8 dB at 2.9 GHz, 11 dB at 12.8 GHz, and 14 dB up to 26.5 GHz.

Spurious testing can be reduced from days to hours from the wider bandwidth that is useful due to the better sensitivity. Low-level signals from antennas will become visible as the noise floor is lowered. Broadband signal detection such as pulsed RF or electromagnetic interference is easier with a lower noise figure.

Using the preamplifier is easy. Simply install the preamplifier into the mainframe and connect the cable. Whether in bypass mode or preamplifier mode, the system automatically calibrates the display of the analyzer.


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Frequency

Frequency Range 1.0 - 26.5 GHz 100 kHz - 26.5 GHz (option 001)

Amplitude

Gain 28 dB (nominal) Band Gain (characteristic) 1 - 12.7 GHz >=26 dB 12.7 - 26.5 GHz >=24 dB Option 001 28 dB (nominal) Band Gain (characteristic) 100 kHz - 2.9 GHz >=22 dB 2.7 - 12.7 GHz >=25 dB 12.7 - 22 GHz >=23 dB 22 - 26.5 GHz >=20 dB Noise Figure Band NF (characteristic) 1 - 12.7 GHz <=9 dB 12.7 - 22 GHz <=13 dB 22 - 26.5 GHz <=15 dB Option 001 Band NF (characteristic) 100 kHz - 2.9 GHz <=7.5 dB 2.7 - 12.7 GHz <=9.5 dB 12.7 - 22 GHz <=14 dB 22 - 26.5 GHz <=16 dB Frequency Response Peak Variation (ądB, characteristic) Band Preamp Bypass 1 - 12.7 GHz 1.0 0.8 1 - 26.5 GHz 2.2 1.3 Option 001 Peak Variation (ądB, characteristic) Band Preamp Bypass 100 kHz - 2.9 GHz 1.2 0.8 2.7 - 12.7 GHz 1.2 0.8 2.7 - 22 GHz 1.9 1.3 100 kHz - 26.5 GHz 2.5 1.3 Bypass Insertion Loss Band Loss (characteristic) 0 - 2.9 GHz <=1.0 dB 2.7 - 12.7 GHz <=1.8 dB 12.7 - 26.5 GHz <=2.5 dB

Distortion

Gain Compression < 1dB for signals <=+7 dBm at the output (option 001, for signals <=0 dBm at the output 100 kHz to 2.9 GHz) Third-Order Intercept +15 dBm at output (characteristic) Option 001 Band TOI (characteristic) 0 - 2.9 GHz 0 dBm 2.7 - 26.5 GHz +15 dBm Second Harmonic Intercept +30 dBm at output (characteristic)

Spectrum Analyzer Displayed Average Noise Level

HP 71209A and HP 70620B DANL (dBm, characteristic) Band Preamp Bypass 1 GHz - 2.0 GHz <-155 <-137 2 GHz - 2.9 GHz <-155 <-136 2.7 - 6.2 GHz <-155 <-136 6.0 - 12.8 GHz <-155 <-136 12.6 - 22 GHz <-150 <-128 22 - 26.5 GHz <-148 <-126 HP 71209A and HP 70620B option 001 DANL (dBm, characteristic) Band Preamp Bypass 1 GHz - 2.0 GHz <-155 <-137 2 GHz - 2.9 GHz <-155 <-136 2.7 - 6.2 GHz <-154 <-136 6.0 - 12.8 GHz <-154 <-136 12.6 - 22 GHz <-148 <-128 22 - 26.5 GHz <-145 <-126 HP 71210C and HP 70620B DANL (dBm, characteristic) Band Preamp Bypass 1 GHz - 2.9 GHz <-155 <-138 2.7 - 12.8 GHz <-153 <-132 12.6 - 22 GHz <-150 <-131 HP 71210C and HP 70620B option 001 DANL (dBm, characteristic) Band Preamp Bypass 10 MHz - 2.9 GHz <-156 <-137 2.7 - 12.8 GHz <-153 <-132 12.6 - 22 GHz <-150 <-130 HP 71910A and HP 70620B DANL (dBm, characteristic) Band Preamp 1 - 12.8 GHz <-155 12.6 - 22 GHz <-150 22 - 26.5 GHz <-148


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP 70620B	microwave preamplifier
   Opt 001	Extend RF frequency range to 100 kHz from 1 GHz

HP 70861A	RAM/ROM Board Upgrade Kit

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

To help you make the right instrument choices

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Related product families available online:



Contact HP Test & Measurement. (c) Copyright 1994, 1995, 1996, 1997 Hewlett-Packard Company.

Page Updated: Wednesday January 07 23:27:48 UTC 1998