HP 70621A Preamplifier, 100 kHz to 2.9 GHz![]()
|
Spurious testing can be reduced from days to hours from the wider bandwidth that is useful due to the better sensitivity. Low-level signals from antennas will become visible as the noise floor is lowered. Broadband signal detection such as pulsed RF or electromagnetic interference is easier with a lower noise figure.
Using the preamplifier is easy. Simply install the preamplifier into the mainframe and connect the cable. Whether in bypass mode or preamplifier mode, the system automatically calibrates the display of the analyzer.
Frequency
Frequency Range 100 kHz - 2.9 GHz
Amplitude
Gain 28 dB (nominal) >=24 dB (characteristic) Noise Figure <6 dB (characteristic) Frequency Response Peak Variation (ądB, characteristic) Band Preamp Bypass 0 Hz - 2.9 GHz N/A 0.8 100 kHz - 2.9 GHz 1.2 --
Distortion
Gain Compression < 1dB for signals <=0 dBm at the output Third-Order Intercept 0 dBm at output (characteristic) Second Harmonic Intercept +30 dBm at output (characteristic)
Spectrum Analyzer Displayed Average Noise Level
HP 71100C Peak Variation (ądB, characteristic) Band Preamp Bypass 100 - 100 kHz N/A 1.4 100 kHz - 2.5 GHz 1.8 1.4 100 kHz - 2.9 GHz 2.1 1.8 Referenced to a Calibrator Signal Peak Variation (ądB, characteristic) Band Preamp Bypass 100 - 100 kHz N/A 1.6 100 kHz - 2.5 GHz 2.0 1.6 100 kHz - 2.9 GHz 2.5 2.3
HP 70621A RF preamplifier, 100 kHz TO 2.9 GHz Opt 098 CPU/firmware upgrade to the HP 70900A Opt 099 Firmware upgrade to the HP 70900A/B Opt 910 Extra operating manual Opt 915 Service manual
Related product families available online:
Page Updated: Wednesday January 07 23:27:42 UTC 1998