HP 71100C RF Spectrum Analyzer, 100 Hz to 2.9 GHz![]()
|
As part of the Modular Measurement System (MMS) family, you can choose a configuration that best meets your requirements in performance, size, and price. Adding accessories such as a preamplifier or tracking generator is very easy.
For automated testing, the HP 70000 series of spectrum analyzers are compatible with the HP 8566B. Their proven reliability and repeatability allow a three-year calibration cycle, giving you a very low lifetime cost.
Frequency Specifications
Frequency Range: 100 Hz to 2.9 GHz, dc coupled 100 kHz to 2.9 GHz, ac coupled Tunable in 1 Hz increments Frequency Readout Accuracy Span <=10 MHz: ±[(freq readout x freq ref accuracy) + 1.0% of span + 10 Hz] Span >10 MHz Sweep >=20 ms: ±[(freq readout x freq ref accuracy) + 1.5% of span + 10 Hz] Sweep >=10 ms: ±[(freq readout x freq ref accuracy) + 2.5% of span + 10 Hz] Frequency Span Range: 1 Hz to 2.9 GHz in 0.5% increments and 0 Hz Accuracy Span <=10 MHz: ±[1% of span + (span x freq ref accuracy)] Span >10 MHz Sweep >=50 ms: ± [1.5% of span + (span x freq ref acc)] Sweep >=20 m: ± [2.5% of span + (span x freq ref acc)] Sweep >=10 ms: ± [4.0% of span + (span x freq ref acc)] Frequency Reference Accuracy w/HP 70310A w/o HP 70310A Aging: <1x10-7/year <3x10-6/year <5x10-10/day (7-day avg.) Temperature Drift: <7x10-9 <1x10-5 Sweep Time Range (Continuous): 10 ms to 1000 s with HP 70700A: Swept freq span: 15 ms to 355 s Fixed freq (zero span) 80 µs to 355 s with 800 trace points
Amplitude Specifications
Frequency Response Band Variation Peak Variation: 100 Hz to 2.5 GHz ±1.0 dB 10 dB attn: 100 Hz to 2.9 GHz ±1.5 dB Referenced to 100 Hz to 2.5 GHz ±1.3 dB 300 MHz, -10 dBm 100 Hz to 2.9 GHz ±1.8 dB Calibrator (10 dB attn) Calibrator Uncertainty: ±0.3 dB (-10 dBm, 300 MHz) Input Attenuator Range: 0 to 70 dB range in 10 dB steps Switching Repeatability: ±0.2 dB IF Gain Accuracy: Gain 20 to 30 °C 0 to 50 °C 10 dB ±0.2 dB ±0.2 dB 20 dB ±0.2 dB ±0.2 dB 30 dB ±0.2 dB ±0.3 dB 40 dB ±0.2 dB ±0.5 dB 50 dB ±0.2 dB ±0.6 dB Scale Fidelity Bandwidth Fidelity Log, corrected HP 70902A: 10 Hz ±0.7 dB (0 to 90 dB) 30 Hz to 100 kHz ±0.5 dB 300 kHz ±0.7 dB Resolution Bandwidth (3 dB, synchronously tuned) Range: 10 Hz to 300 kHz (HP 70902A) 100 kHz to 3 MHz (HP 70903A) (1, 3, 10 sequence and 10% increments except 3 kHz to 10 kHz) Accuracy: ±20% Video Bandwidth Range: 3 Hz to 300 kHz (HP 70902A) (1, 3, 10 sequence) 300 Hz to 3 MHz (HP 70903A) Accuracy: ±20% (characteristic) Marker Resolution: ±0.03 dB
Noise
Displayed Average Noise Level (0 dB attn) Band DANL 10 Hz Res BW: 10 MHz to 2.0 GHz <-134 dBm 2.0 GHz to 2.9 GHz <-131 dBm with HP 70621A 0 MHz to 2.9 GHz <-156 dBm 100 kHz Res BW: For freq >1 MHz, displayed average noise level is 40 dB higher than the above.
False Signals
Spectral Purity Noise Sidebands at 10 kHz Offset: <-108 dBc/Hz Line and System Related Sidebands: <-65 dBc Spurious Responses Input Spurious For mixer level <=-40 dBm 100 Hz to 10 MHz <-60 dBc All spurious responses, 10 MHz to 2.9 GHz <-70 dBc except as listed below, are less than these values (10 dB attn) Second Harmonic Distortion For mixer level Band Second Harmonic <=-40 dBm 100 Hz to 10 MHz <-60 dBc (10 dB attn) 10 MHz to 2.9 GHz <-70 dBc Third Order Intermodulation Distortion for two signals Center Intermod Equiv each <=-30 dBm Frequency Products TOI at mixer (10 dB attn) 100 Hz to 10 MHz <-66 dBc +3 dBm 10 MHz to 2.9 GHz <-70 dBc +5 dBm Image Responses <=-90 dBc for RF input levels <=0 dBm at 6,42.8, and 642.8 MHz offset (10 dB attn) Residual Responses Range Responses (0 dB input attn 10 MHz to 2.9 GHz <-100 dBm with input terminated)
If you have Adobe's Acrobat Reader already available on your system, please click on the icon provided below to print or view the specifications.
If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader, and then click on the icon below.
HP 71100C RF Spectrum Analyzer, 100 Hz to 2.9 GHz Opt 006 Delete HP 70902A 10 Hz to 300 kHz BW IF section (Must order HP 70903A IF section) Opt 110 Delete HP 70310A precision frequency reference Opt 121 Add distribution amplifier to HP 70310A Opt 122 Delete ovenized oscillator from HP 70310A Rack Mounting Options Opt 810 Slide kit for both HP 70001A mainframe and HP 70004A display/mainframe Opt 908 Flange kit to mount without handles Opt 913 Flange kit to mount with handlesAll orders include the following:HP 70621A preamplifier for better sensitivity HP 70300A RF tracking generator, 20 Hz to 2.9 GHz
HP 70001A, HP 70004A, HP 70310A, HP 70900B, HP 70902A, HP 70904A, and user documentation.
HP 70875A Noise Figure Measurement Personality, Product Overview
( PDF)
HP 41800A Active Probe (5 Hz TO 500 MHz), Data Sheet
( PDF)
If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader now. |
![]() |
Related MMS Spectrum Analyzer Family product datasheets:
Related product families available online:
Page Updated: Thursday February 18 07:51:07 UTC 1999