HP 71200C Microwave Spectrum Analyzer, 50 kHz to 22 GHz![]()
|
The standard configuration gives unpreselected microwave capability at just over the price of an RF analyzer with similar performance in terms of speed, phase noise, and feature set. If you make RF measurements and occasionally view microwave signals, this analyzer may be ideal. Or, if you measure known signals in a controlled environment such as manufacturing, this analyzer offers high performance at an economical cost.
To measure several unknown signals or spurious signals, preselection is recommended. It will eliminate the multiple and image responses that occur in microwave harmonic mixing, saving the time it takes to identify unknown signals. And for coverage to 325 GHz, you can add an HP 70907B external mixer interface module.
The HP 70620B preamplifier modules can be used with the HP 71200C spectrum analyzer for improved sensitivity. Option 001 on the HP 70620B provides coverage down to 100 kHz. Use the HP 70700A digitizer for 80 microsec sweeptimes in zero-span (fixed-tuned).
Partial Specifications for standard and Option 002 modelsFrequency Specifications
Frequency Range HP 70905A/B: 50 kHz to 22 GHz Tunable in 1 Hz increments N 1st IF Frequency H=high IF (3.6214 GHz) L=low IF (321.4 MHz) 1 H- 50 kHz to 2.9 GHz N=harmonic number 1 L- 2.7 GHz to 6.2 GHz 2 L- 6.0 GHz to 12.7 GHz 3 L+ 12.0 GHz to 19.9 GHz 4 L+ 19.7 GHz to 22 GHz Frequency Readout Accuracy Span <=10 MHz x N1: ±[(freq readout x freq ref accuracy) ±1.0% of Span +10 Hz] Span >10 MHz x N1 Sweep >=20 ms ±[(freq readout x freq ref accuracy) +1.5% of span +10 Hz] Sweep >=10 ms ±[(freq readout x freq ref accuracy) +2.5% of span + 10 Hz] Frequency Span Range: 0 to 22 GHz in 0.5% increments Accuracy Span <=10 MHz: ±[1% of span + (span x freq ref accuracy)] Span >10 MHz Sweep >=50ms: ±[1.5% of span + (span x freq ref accuracy)] Sweep >=20ms: ±[2.5% of span + (span x freq ref accuracy)] Sweep >=10ms: ±[4.0% of span + (span x freq ref accuracy)] Frequency Reference Accuracy with (HP 70310A) (without HP 70310A) Aging: >1 x 10(-7)/year >3 x 10(-6)/year 7-Day Average: >5 x 10(-10)/day Temperature Drift: >7 x 10(-9) >1 x 10(-5) Spectral Purity² Frequency Range Noise Sideband Offset 50 kHz to 2.9 GHz <-108 dBc/Hz 10 kHz 2.7 to 6.2 GHz <-108 dBc/Hz 30 kHz 6.0 to 12.7 GHz <-102 dBc/Hz 30 kHz 12.5 to 19.9 GHz <-98 dBc/Hz 30 kHz 19.7 to 22/26.5 GHz <-96 dBc/Hz 30 kHz Line and System Related Sidebands: <-65 dBc + 20 log N1 Residual FM Span >10 MHz x N1: <25 kHz p-p in 0.1 s (measurement BW=100 kHz) Span <=10 MHz x N1: Determined from phase-noise sidebands Frequency Drift For spans >10 MHz x N1: freq drift is ±1 kHz/s and +150 kHz/°C (Errors due to drift are not cumulative sweep to sweep.) Sweep Time Range (Continuous): 10 ms to 1000 s Accuracy: ±2% With HP 70700A Swept freq span: 15 ms to 355 s Fixed freq (zero span): 801 µs to 355 s with 800 point trace Trigger: free run, line, video, external Resolution Bandwidth (3 dB, synchronously tuned) Range (1, 3, 10, and 10% increments, except 3 kHz to 10 kHz) HP 70902A: 10 Hz to 300 kHz HP 70903A: 100 kHz to 3 MHz Accuracy: ±20% Selectivity Bandwidth (-60 dB/-3 dB) 10 Hz to 3 kHz: <12:1 10 kHz to 3 MHz: <16:1 Video Bandwidth Range (1, 3, 10 sequence) HP 70902A: 3 Hz to 300 kHz HP 70903A: 300 Hz to 3 MHz (When set to maximum (300 kHz or 3 MHz), effective bandwidth is greater than specified.) Accuracy: ±20% (characteristic) 1N=Harmonic mixing band constant. 2Refer to Figure 1 in the Spectrum Analyzer Overview for typical phase noise.Amplitude Specifications
Maximum Safe Input Power AC Average Continuous 0 dB Attenuation: +15 dBm 10 dB Attenuation: +25 dBm >10 dB Attenuation: +30 dBm Pulse Power: 100 watts, 10 ms pulse (>=40 dB attenuation) dc: 0 volts Display Range (10 divisions) Calibration Log: 0.01 to 20 dB/div in 0.5% increments Linear: 0 to 10% of reference level per division Reference Level Range Log: +30 to -1400 dBm Linear: 7.07 V to 22 nV Calibrator Uncertainty: ±0.3 dB (-10 dBm, 300 MHz) Input Attenuator Switching Repeatability: ±0.2 dB IF Gain Accuracy Gain 20 to 30 °C 0 to 50 °C 10 dB ±0.2 dB ±0.2 dB 20 dB ±0.2 dB ±0.2 dB 30 dB ±0.2 dB ±0.3 dB 40 dB ±0.2 dB ±0.5 dB 50 dB ±0.2 dB ±0.6 dB Scale Fidelity Bandwidth Fidelity Log, Corrected (1-3-10) HP 70902A: (0 to 90 dB) 10 Hz ±0.7 dB 30 Hz to 100 kHz ±0.5 dB 300 kHz ±0.7 dB Log, Uncorrected: All ±3.0 dB Incremental, Corrected: All ±0.1 dB/1 dB Linear ±7.5% of reference level Amplitude Temperature Drift (nominal) 10 dB Input Attentuation ±0.05 dB/°C 100 Hz Res BW (HP 70902A IF) 300 kHz Res BW (HP 70903A IF) (Accumulated error is eliminated by running internal correction routine.) Resolution Bandwidth Switching Repeatability In 1, 3, 10 Sequence: ±0.1 dB All Bandwidths: ±3 dB (uncorrected) Marker Resolution: ±0.03 dBInput/Output Characteristics
Front panel only for standard configuration and Option 002. See individual module characteristics for complete information. HP 70900B LO Section 300 MHz Calibrator Output: BNC (f), 50 ohms (nominal) Output Power: -10 dBm ±0.3 dB Frequency Accuracy: 300 MHz x freq reference accuracy HP 70905A/B RF Section (50 kHz to 22 GHz) RF Input: HP 70905A/B: Type N (f); 50 ohms (nominal) LO Emissions: <-10 dBm with 10 dB attenuation (nominal) VSWR (>=10 dB Attenuation) Frequency VSWR (nominal) 0 to 12.7 GHz <1.7:1 12.5 to 18.0 GHz <2.0:1 18.0 to 22 GHz <2.5:1 HP 70600A Preselector (50 kHz to 22 GHz, used with HP 70905B) RF Input HP 70600A: Type N (f) LO Emissions (10 dB attenuation, nominal) Bypass Mode 0 to 22 GHz: <-10 dBm Filtered Mode 0 to 2.9 GHz <-50 dBm 2.7 to 22 GHz: <-80 dBm VSWR (nominal, >=10 dB attenuation) Frequency VSWR 0 to 12.7 GHz <1.9:1 12.5 to 22 GHz <2.2:1 RF Output: SMA, 50 ohms (nominal), front panel HP 70902A IF Section Auxiliary Video Output: BNC (f), 0 to 1 V, 1k ohms (nominal), 3 MHz IF Output (linear): BNC (f), 50 ohms, 1.5:1 VSWR (nominal) Output Power: -15 dBm (nominal) with -10 dBm RF input, 0 dB attenuation and -10 dBm reference level HP-IB Codes: SHI, AH1, T6, L4, SR1, RL1, DC1, PP0, DT1, E2, C1General Specifications
HP 71200C System Components HP 70001A HP 70004A HP 70900B HP 70310A HP 70902A Standard: HP 70905A Opt 002: HP 70905B, HP 70600A Environmental Temperature Operational: 0 to +55 °C Storage: -40 to +75 °C Humidity Operational: 0 to 95% relative humidity at 45 °C EMC: Conducted and radiated interference is in compliance with CISPR pub 11, FTZ 526/1979, and MIL-STD 461B, RE02/part 7. Vibration and Shock: in compliance with MIL-T-28800E Type III Class 3 Power Requirements: see requirements for HP 70001A and HP 70004A all power requirements supplied by the mainframe (HP 70001A or 70004A) Weight (nominal) HP 71200C Standard: 47.7 kg (105.6 lb) Dimensions HP 70001A Mainframe: 177.0 mm H x 425.4 mm W x 526.0 mm L (6.97 in x 16.75 in x 20.7 in) HP 70004A Display: 222.0 mm H x 425.4 mm W x 526.0 mm L (8.74 in x 16.75 in x 20.7 in) Warranty and Calibration Warranty: 1 year (extendible with options) Calibration Cycle: 3 years recommended Features and Compatibility HP 70004A Display Features: memory card, direct-to-disk, keyboard (for title mode and writing small DLPs), direct plot (buffered), direct print, full color display Mass Storage Memory card: 32 KB or 128 KB RAM per card External: SS80-compatible hard or flexible disk User Memory: 128 KB minimum, 32 k bytes minimum with firmware before 901008, about 2.5 KB to store an 800 point trace with its state. System memory is reduced when slave modules are added. Optional 1 MB memory. Compatible Accessory Modules (slave modules to the HP 70900B master module) HP 70903A IF Section HP 70621A and HP 70620B Preamplifiers HP 70300A and HP 70301A Tracking Generators HP 70810B Lightwave Section HP 70907B External Mixer Interface Module HP 70700A Digitizer HP 70205A Monochrome Display Software Available HP 11990A Performance Verification Software HP 70871A Scalar Measurement Personality
If you have Adobe's Acrobat Reader already available on your system, please click on the icon provided below to print or view the specifications.
If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader, and then click on the icon below.
(Note: When adding or exchanging modules, be sure the final count will fit into the 8-slot HP 70001A mainframe and the 4-slot HP 70004A display/mainframe.) HP 71200C Spectrum Analyzer, 50 kHz to 22 GHz Opt 002 22 GHz preselected operation; delete HP 70905A, add HP 70905B and HP 70600A Opt 006 Delete HP 70902A 10 Hz to 300 kHz IF section (note: must order an HP 70903A 100 kHz to 3 MHz IF section) Opt 1BN Certificate of Calibration Opt lBP Certificate of Calibration and Data Opt 110 Delete HP 70310A Precision Frequency Reference Opt 121 Add Distribution Amplifier on the HP 70310A Precision Frequency reference (HP 70310A/001) Opt 122 Provide External Reference Capability and Delete Ovenized Reference in the HP 70310A Opt 200 Delete Display (operation and control over HP-IB only) Opt 201 Delete Mainframe (note: count modules to ensure fit in mainframe) Opt 205 Substitute HP 70205A Display for HP 70004A Display/Mainframe Opt 400 Add 400 Hz Power Line Frequency Operation to the HP 70001A mainframe; adds isolation transformer at 2.3kg (5lb) Opt 512 Additional Memory, 1 MB total, 90 KB user memory Opt 810 Rack Mount Slide Kit for both the HP 70001A mainframe and the HP 70004A display/mainframe Opt 908 Rack Flange Kit to Mount HP 70001A/HP 70004A without Handles Opt 910 Extra User Manual Set containing: Installation/Verification, Operation Program Language Reference Opt 913 Rack Flange Kit to Mount HP 70001A/HP 70004A with Handles Attached Opt 915 Service Manual Set and Software for Troubleshooting and Repair
Available accessories for this product include:
No literature is available online at this time.
Additional help is a click away!
Want to talk to a knowledgeable engineer about this product at our HP Test and Measurement Call Center? Do you have a question you want to send in by e-mail? A measurement set-up you need to fax in? Follow the link above and you will be given a choice to contact an engineer by telephone, e-mail or fax.
This service is available only in the United States. For assistance outside the U.S., contact your local HP DIRECT or HP Sales Office.
Related MMS Spectrum Analyzer Family product datasheets:
Related product families available online:
Page Updated: Thursday February 18 08:07:05 UTC 1999